Calculation of electric fields induced near metal implants by magnetic resonance imaging switched-gradient magnetic fields

被引:13
|
作者
Buechler, DN [1 ]
Durney, CH [1 ]
Christensen, DA [1 ]
机构
[1] UNIV UTAH,DEPT ELECT ENGN,SALT LAKE CITY,UT 84112
关键词
magnetic resonance imaging; metallic implants; induced electric fields; safety;
D O I
10.1016/S0730-725X(97)00179-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electric (E) fields induced near metal implants by MRI switched-gradient magnetic fields are calculated by a new equivalent-circuit numerical technique. Induced E-field results are found for a metallic spinal-fusion implant consisting of two thin wires connected to the metallic case of a current generator as well as for its subsections: a bare U-shaped wire, an insulated U-shaped wire, a cut insulated wire, and a generator. The presence of the metallic implants perturbs the E field significantly. Near the ends of the bare U-shaped wire, the E field is 89.7 times larger than in the absence of the wire. The greatest E field concentration occurs near the ends of the cut insulated wire, where the E field is 196.7 times greater than in the absence of the wire. In all cases, the perturbation of the induced E field by the implanted wire is highly localized within a few diameters of the wire. (C) 1997 Elsevier Science Inc.
引用
收藏
页码:1157 / 1166
页数:10
相关论文
共 50 条
  • [21] Photodetachment of H− in parallel electric and magnetic fields near metal surface
    T. T. Tang
    Z. L. Zhu
    J. G. Yao
    D. B. Kong
    D. F. Yin
    [J]. Indian Journal of Physics, 2016, 90 : 741 - 748
  • [22] Comparing modified vertical gradient freezing with rotating magnetic fields or with steady magnetic and electric fields
    Wang, X
    Ma, N
    Bliss, DF
    Iseler, GW
    Becla, P
    [J]. JOURNAL OF CRYSTAL GROWTH, 2006, 287 (02) : 270 - 274
  • [23] Anisotropy of magnetic emulsions induced by magnetic and electric fields
    Dikansky, Yury I.
    Zakinyan, Arthur R.
    Tyatyushkin, Alexander N.
    [J]. PHYSICAL REVIEW E, 2011, 84 (03)
  • [24] ELECTRON RESONANCE IN CROSSED ELECTRIC AND MAGNETIC FIELDS
    LIFSHITZ, IM
    KAGANOV, MI
    [J]. SOVIET PHYSICS JETP-USSR, 1960, 10 (02): : 392 - 393
  • [25] ON COMBINED RESONANCE IN CROSSED MAGNETIC AND ELECTRIC FIELDS
    BUISHVILI, LL
    [J]. SOVIET PHYSICS-SOLID STATE, 1962, 4 (02): : 407 - 408
  • [26] Imaging of magnetic and electric fields by electron microscopy
    Zweck, Josef
    [J]. JOURNAL OF PHYSICS-CONDENSED MATTER, 2016, 28 (40)
  • [27] Reaction imaging in uniform electric and magnetic fields
    Feagin, James M.
    Briggs, John S.
    [J]. JOURNAL OF PHYSICS B-ATOMIC MOLECULAR AND OPTICAL PHYSICS, 2014, 47 (11)
  • [28] Electromagnetic-acoustic modeling of fields induced by gradient pulses in diffusion tensor magnetic resonance imaging
    Elshafiey, I
    Melapudi, VR
    Udpa, L
    [J]. REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 25A AND 25B, 2006, 820 : 728 - 735
  • [29] Magnetic Resonance Imaging in Inhomogeneous Magnetic Fields with Noisy Signal
    Arpinar, V. E.
    Eyuboglu, B. M.
    [J]. 4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 410 - 413
  • [30] Photodetachment of H- in parallel electric and magnetic fields near metal surface
    Tang, T. T.
    Zhu, Z. L.
    Yao, J. G.
    Kong, D. B.
    Yin, D. F.
    [J]. INDIAN JOURNAL OF PHYSICS, 2016, 90 (07) : 741 - 748