TOPAS Monte Carlo simulation for a scanning proton therapy system in SPHIC

被引:2
|
作者
Chen, Zhi [1 ,2 ,3 ]
Liu, Hongdong [4 ]
Zhao, Jingfang [2 ,3 ,5 ]
Kaess, Stefanie [6 ]
机构
[1] Shanghai Proton & Heavy Ion Ctr, Dept Med Phys, Shanghai 201315, Peoples R China
[2] Shanghai Key Lab Radiat Oncol, 20dz2261000, Shanghai, Peoples R China
[3] Shanghai Engn Res Ctr Proton & Heavy Ion Radiat T, Shanghai, Peoples R China
[4] Sun Yat Sen Univ, Dept Radiat Oncol, Ctr Canc, Guangzhou 510060, Peoples R China
[5] Fudan Univ Canc Hosp, Shanghai Proton & Heavy Ion Ctr, Dept Med Phys, Shanghai 201315, Peoples R China
[6] RaySearch Labs, Shanghai, Peoples R China
关键词
TOPAS; Monte Carlo; Proton therapy;
D O I
10.1016/j.jrras.2022.01.016
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: This study aims to use TOPAS Monte Carlo (MC) code to model the beamline of a scanned proton therapy system in Shanghai Proton and Heavy Ion Center (SPHIC). This tool can be useful for treatment plan quality assurance. Methods and materials: In-air lateral profiles of single proton spots at five positions along the Z-axis were measured, which can then be used to calculate the angular variance, spatial angular covariance, and spatial variance at the beam exit position. The mean energy and energy spread of the primary incident beam for 15 energies were tuned by comparing with measured relative integrated depth dose at the central axis. Three spread-out Bragg peak plans in homogeneous water cubes and one plan with a half anthropomorphic head phantom were created to verify the accuracy of the model. In addition, the comparison was also performed to RayStation TPS with an MC dose calculation engine for the plans. Results: The average difference of spot sizes was 1.3% for all five energies. For all 15 energies, the average range (D90) difference was 0.04 mm, the mean distal-fall-off (DDF) difference was 0.04 mm, and the mean full width at half maximum (FWHM) difference was 0.10 mm. For the three homogeneous cube plans, a maximum difference of 0.2 mm for the D90 was observed, and the absolute differences in the lateral penumbras for each plan all agreed within 1 mm. For the plan on the anthropomorphic half head phantom, the mean dose deviation is less than 1.2% compared to measurements. Conclusion: We successfully developed an MC model for the proton scanning system used in SPHIC, which can be helpful for dose verification of the Treatment Planning System (TPS) and quality assurance in the future.
引用
收藏
页码:122 / 129
页数:8
相关论文
共 50 条
  • [21] Monte Carlo (MC) Simulation of Total Skin Electron Therapy (TSET) with TOPAS and EGSnrc
    Zhong, W.
    Ding, G.
    Ong, Y.
    Zhu, T.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [22] MONTE CARLO SIMULATION OF MULTIFOCAL STOCHASTIC SCANNING SYSTEM
    Liu, Lixin
    Qian, Jia
    Li, Yahui
    Peng, Xiao
    Yin, Jun
    [J]. JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2014, 7 (01)
  • [23] Pencil-Beam Versus Monte Carlo Based Dose Calculation for Proton Therapy Patients with Complex Geometries. Clinical Use of the TOPAS Monte Carlo System
    Schuemann, J.
    Shin, J.
    Perl, J.
    Grassberger, C.
    Verburg, J.
    Faddegon, B.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3820 - 3820
  • [24] TOPAS: An innovative proton Monte Carlo platform for research and clinical applications
    Perl, J.
    Shin, J.
    Schuemann, J.
    Faddegon, B.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2012, 39 (11) : 6818 - 6837
  • [25] A GPU Monte Carlo to support clinical routine in a compact spot scanning proton therapy system
    Gajewski, J.
    Schiavi, A.
    Krah, N.
    Patera, V.
    Vilches-Freixas, G.
    Martens, J.
    Unipan, M.
    Rucinski, A.
    Nijsten, B.
    Bosmans, G.
    Rinaldi, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S752 - S753
  • [26] Development and validation of the Dynamic Collimation Monte Carlo simulation package for pencil beam scanning proton therapy
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Hyer, Daniel E.
    Geoghegan, Theodore J.
    Patwardhan, Kaustubh A.
    Smith, Blake R.
    Flynn, Ryan T.
    Yu, Jen
    Rana, Suresh
    Gutierrez, Alonso N.
    Hill, Patrick M.
    [J]. MEDICAL PHYSICS, 2021, 48 (06) : 3172 - 3185
  • [27] Monte Carlo Computation of the Effective Dose in Scanning Beam Proton Therapy
    Mirkovic, D.
    Titt, U.
    Sawakuchi, G.
    Zhang, X.
    Li, Y.
    Mohan, R.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [28] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [29] Automation of Monte Carlo simulations for a proton therapy system
    Zheng, Y.
    Newhauser, W.
    Fontenot, J.
    Koch, N.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2073 - 2074
  • [30] Automated Monte Carlo Simulation of Proton Therapy Treatment Plans
    Verburg, Joost Mathijs
    Grassberger, Clemens
    Dowdell, Stephen
    Schuemann, Jan
    Seco, Joao
    Paganetti, Harald
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2016, 15 (06) : NP35 - NP46