TOPAS Monte Carlo simulation for a scanning proton therapy system in SPHIC

被引:2
|
作者
Chen, Zhi [1 ,2 ,3 ]
Liu, Hongdong [4 ]
Zhao, Jingfang [2 ,3 ,5 ]
Kaess, Stefanie [6 ]
机构
[1] Shanghai Proton & Heavy Ion Ctr, Dept Med Phys, Shanghai 201315, Peoples R China
[2] Shanghai Key Lab Radiat Oncol, 20dz2261000, Shanghai, Peoples R China
[3] Shanghai Engn Res Ctr Proton & Heavy Ion Radiat T, Shanghai, Peoples R China
[4] Sun Yat Sen Univ, Dept Radiat Oncol, Ctr Canc, Guangzhou 510060, Peoples R China
[5] Fudan Univ Canc Hosp, Shanghai Proton & Heavy Ion Ctr, Dept Med Phys, Shanghai 201315, Peoples R China
[6] RaySearch Labs, Shanghai, Peoples R China
关键词
TOPAS; Monte Carlo; Proton therapy;
D O I
10.1016/j.jrras.2022.01.016
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: This study aims to use TOPAS Monte Carlo (MC) code to model the beamline of a scanned proton therapy system in Shanghai Proton and Heavy Ion Center (SPHIC). This tool can be useful for treatment plan quality assurance. Methods and materials: In-air lateral profiles of single proton spots at five positions along the Z-axis were measured, which can then be used to calculate the angular variance, spatial angular covariance, and spatial variance at the beam exit position. The mean energy and energy spread of the primary incident beam for 15 energies were tuned by comparing with measured relative integrated depth dose at the central axis. Three spread-out Bragg peak plans in homogeneous water cubes and one plan with a half anthropomorphic head phantom were created to verify the accuracy of the model. In addition, the comparison was also performed to RayStation TPS with an MC dose calculation engine for the plans. Results: The average difference of spot sizes was 1.3% for all five energies. For all 15 energies, the average range (D90) difference was 0.04 mm, the mean distal-fall-off (DDF) difference was 0.04 mm, and the mean full width at half maximum (FWHM) difference was 0.10 mm. For the three homogeneous cube plans, a maximum difference of 0.2 mm for the D90 was observed, and the absolute differences in the lateral penumbras for each plan all agreed within 1 mm. For the plan on the anthropomorphic half head phantom, the mean dose deviation is less than 1.2% compared to measurements. Conclusion: We successfully developed an MC model for the proton scanning system used in SPHIC, which can be helpful for dose verification of the Treatment Planning System (TPS) and quality assurance in the future.
引用
收藏
页码:122 / 129
页数:8
相关论文
共 50 条
  • [1] TOPAS Monte Carlo model of MD anderson scanning proton beam for simulation studies in proton therapy
    Hartman, J.
    Zhang, X.
    Zhu, X. R.
    Frank, S. J.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (03):
  • [2] Monte Carlo Simulation Using TOPAS for Synchrotron Based Proton Discrete Spot Scanning System
    Moskvin, V.
    Pirlepesov, F.
    Tsiamas, P.
    Axente, M.
    Lukose, R.
    Zhao, L.
    Shin, J.
    Farr, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3497 - 3498
  • [3] TOPAS Monte Carlo simulation for double scattering proton therapy and dosimetric evaluation
    Liu, Hongdong
    Li, Zuofeng
    Slopsema, Roelf
    Hong, Liu
    Pei, Xi
    Xu, Xie George
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 62 : 53 - 62
  • [4] Development and Validation of a TOPAS Monte Carlo Model of a Dynamic Collimation System for Pencil Beam Scanning Proton Therapy
    Nelson, N.
    Culberson, W.
    Hyer, D.
    Geoghegan, T.
    Patwardhan, K.
    Smith, B.
    Flynn, R.
    Yu, J.
    Rana, S.
    Gutierrez, A.
    Hill, P.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [5] Experimental validation of the TOPAS Monte Carlo system for passive scattering proton therapy
    Testa, M.
    Schuemann, J.
    Lu, H. -M.
    Shin, J.
    Faddegon, B.
    Perl, J.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (12)
  • [6] Modeling of a Compact Proton Therapy System Using TOPAS Monte Carlo Code
    Prusator, M.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Validation and Implementation of Proton Pencil Beam Scanning Nozzle Using TOPAS Monte Carlo Simulation
    Lin, C.
    Feng, Y.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E875 - E875
  • [8] Meeting the Challenges of Quality Control in the TOPAS Monte Carlo Simulation Toolkit for Proton Therapy
    Hall, D.
    Perl, J.
    Schuemann, J.
    Faddegon, B.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3493 - 3494
  • [9] Benchmarking of the TOPAS Monte Carlo System Against Phantom Dose Measurements in Proton Therapy
    Testa, M.
    Schuemann, J.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3945 - 3945
  • [10] The proton therapy nozzles at Samsung Medical Center: A Monte Carlo simulation study using TOPAS
    Kwangzoo Chung
    Jinsung Kim
    Dae-Hyun Kim
    Sunghwan Ahn
    Youngyih Han
    [J]. Journal of the Korean Physical Society, 2015, 67 : 170 - 174