Ability of the VITEK 2 advanced expert system to identify β-lactam phenotypes in isolates of Enterobacteriaceae and Pseudomonas aeruginosa

被引:54
|
作者
Sanders, CC
Peyret, M
Moland, ES
Shubert, C
Thomson, KS
Boeufgras, JM
Sanders, WE
机构
[1] Creighton Univ, Ctr Res Anti Infect & Biotechnol, Sch Med, Dept Med Microbiol & Immunol, Omaha, NE 68178 USA
[2] BioMerieux Inc, Hazelwood, MO 63042 USA
[3] BioMerieux, LaBalme Les Grottes, France
关键词
D O I
10.1128/JCM.38.2.570-574.2000
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The Advanced Expert System (AES) was used in conjunction with the VITEK 2 automated antimicrobial susceptibility test system to ascertain the p-lactam phenotypes of 196 isolates of the family Enterobacteriaeae and the species Pseudomonas aeruginosa. These isolates represented a panel of strains that had been collected from laboratories worldwide and whose beta-lactam phenotypes had been characterized by biochemical and molecular techniques. The antimicrobial susceptibility of each isolate tvas determined with the VITEK 2 instrument, and the results were analyzed with the AES to ascertain the beta-lactam phenotype. The results were then compared to the beta-lactam resistance mechanism determined by biochemical and molecular techniques. Overall, the AES was able to ascertain a beta-lactam phenotype for 183 of the 196 (93.4%) isolates tested. For 111 of these 183 (60.7%) isolates, the correct beta-lactam phenotype was identified definitively in a single choice by the AES, while for an additional 46 isolates (25.1%), the AES identified the correct beta-lactam phenotype provisionally within two or more choices. For the remaining 26 isolates (14.2%), the beta-lactam phenotype identified by the AES was incorrect. However, for a number of these isolates, the error nas due to remediable problems. These results suggest that the AES is capable of accurate identification of the beta-lactam phenotypes of gram-negative isolates and that certain modifications can improve its performance even further.
引用
收藏
页码:570 / 574
页数:5
相关论文
共 40 条
  • [1] Evaluation of the Osiris expert system for identification of β-lactam phenotypes in isolates of Pseudomonas aeruginosa
    Bert, F
    Ould-Hocine, Z
    Juvin, M
    Dubois, W
    Loncle-Provot, V
    Lefranc, V
    Quentin, C
    Lambert, N
    Arlet, G
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (08) : 3712 - 3718
  • [2] Performance of Vitek 2 in antimicrobial susceptibility testing of Pseudomonas aeruginosa isolates with different mechanisms of β-lactam resistance
    Mazzariol, Annarita
    Aldegheri, Marco
    Ligozzi, Marco
    Lo Cascio, Giuliana
    Koncan, Raffaella
    Fontana, Roberta
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (06) : 2095 - 2098
  • [3] RESISTANCE TO BETA-LACTAM ANTIBIOTICS IN CLINICAL ISOLATES OF PSEUDOMONAS-AERUGINOSA AND ENTEROBACTERIACEAE
    STOBBERINGH, EE
    HOUBEN, AW
    PHILIPS, J
    [J]. ANTONIE VAN LEEUWENHOEK JOURNAL OF MICROBIOLOGY, 1987, 53 (03): : 206 - 207
  • [4] Performance Evaluation of VITEK 2 System in Meropenem Susceptibility Testing of Clinical Pseudomonas aeruginosa Isolates
    Acuner, Ibrahim Cagatay
    Bayramoglu, Gulcin
    Birinci, Asuman
    Cihan, Cigdem Cekic
    Bek, Yuksel
    Durupinar, Belma
    [J]. MIKROBIYOLOJI BULTENI, 2011, 45 (03): : 411 - 421
  • [5] Accuracy of β-lactam susceptibility testing results from VITEK and VITEK 2 systems when testing Pseudomonas aeruginosa
    LaBombardi, V.
    Rekasius, V.
    Schreckenberger, P.
    Jones, R.
    [J]. INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2007, 29 : S460 - S460
  • [6] Comparison of workflow and accuracy of identification and antimicrobial susceptibility testing of clinical isolates of Enterobacteriaceae, Pseudomonas aeruginosa and enterococci by Vitek 2 and routine methods
    Rantakokko-Jalava, K
    Elo-Lehtonen, E
    Meurman, O
    [J]. APMIS, 2006, 114 (01) : 43 - 49
  • [7] Evaluation of the Vitek System to accurately test the susceptibility of Pseudomonas aeruginosa clinical isolates against cefepime
    Jones, RN
    Biedenbach, DJ
    Marshall, SA
    Pfaller, MA
    Doern, GV
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1998, 32 (02) : 107 - 110
  • [8] Accuracies of β-lactam susceptibility test results for Pseudomonas aeruginosa with four automated systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2)
    Juretschko, Stefan
    LaBombardi, Vincent J.
    Lerner, Stephen A.
    Schreckenberger, Paul C.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (04) : 1339 - 1342
  • [9] Susceptibility testing of pseudomonas aeruginosa by the Vitek 2 system: A comparison with Etest results
    Saegeman, V
    Huynen, P
    Colaert, J
    Melin, P
    Verhaegen, J
    [J]. ACTA CLINICA BELGICA, 2005, 60 (01): : 3 - 9
  • [10] The Evaluation of Antibiotic Susceptibility of Pseudomonas aeruginosa via Vitek 2 Authomated System
    Ak, Sibel
    Yildiz, Fatma
    Gunduz, Ayten
    Koroglu, Mehmet
    [J]. GAZI MEDICAL JOURNAL, 2016, 27 (02): : 62 - 64