Effects of Mitral Annulus Remodeling Following MitraClip Procedure on Reduction of Functional Mitral Regurgitation

被引:8
|
作者
Hidalgo, Francisco [1 ]
Mesa, Dolores [1 ]
Ruiz, Martin [1 ]
Delgado, Monica [1 ]
Rodriguez, Sara [1 ]
Pardo, Laura [1 ]
Pan, Manuel [1 ]
Lopez, Amador [1 ]
Romero, Miguel A. [1 ]
Suarez de Lezo, Jose [1 ]
机构
[1] Hosp Univ Reina Sofia, Serv Cardiol, Cordoba, Spain
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 2016年 / 69卷 / 11期
关键词
Mitral regurgitation; Mitral annulus; MitraClip; Three-dimensional echocardiography; TO-EDGE REPAIR; PERCUTANEOUS REPAIR; VALVE REPAIR; RECOMMENDATIONS; ANNULOPLASTY; GEOMETRY; OUTCOMES; SURGERY; THERAPY; SYSTEM;
D O I
10.1016/j.rec.2016.04.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction and objectives: The percutaneous mitral valve repair procedure (MitraClip) appears to reduce mitral annulus diameter in patients with functional mitral regurgitation, but the relationship between this and regurgitation severity has not been demonstrated. The aim of this study was to determine the effect of mitral annulus remodeling on the reduction of mitral regurgitation in patients with functional etiology. Methods: The study included all patients with functional mitral regurgitation treated with MitraClip at our hospital until January 2015. Echocardiogram (iE33 model, Philips) was performed in all patients immediately after device positioning. Changes in the mitral annulus correlated with mitral regurgitation severity, as assessed using the effective regurgitant orifice area. Results: The study included 23 patients (age, 65 +/- 14 years; 74% men; left ventricular ejection fraction, 31% +/- 13%; systolic pulmonary artery pressure, 47 +/- 10 mmHg). After the procedure, the regurgitant orifice area decreased by 0.30 cm(2) +/- 0.04 cm(2) (P < .0005), from a baseline of 0.49 cm(2) +/- 0.09 cm(2). Anteroposterior diameter decreased by 3.14 mm +/- 1.01 mm (P < .0005) from a baseline of 28.27 mm +/- 4.9 mm, with no changes in the intercommissural diameter (0.50 mm +/- 0.91 mm vs 40.68 mm +/- 4.7 mm; P = .26). A significant association was seen between anteroposterior diameter reduction and regurgitant orifice area reduction (r = .49; P = .020). Conclusions: In patients with functional mitral regurgitation, the MitraClip device produces an immediate reduction in the anteroposterior diameter. This remodeling may be related to the reduction in mitral regurgitation. (C) 2016 Sociedad Espanola de Cardiologia. Published by Elsevier Espana, S.L.U. All rights reserved.
引用
收藏
页码:1020 / 1025
页数:6
相关论文
共 50 条
  • [21] Functional uncoupling of the mitral annulus and left ventricle with mitral regurgitation and dopamine
    Nguyen, Tom C.
    Itoh, Akinobu
    Carlhall, Carl J.
    Oakes, Robert A.
    Liang, David
    Ingels, Neil B., Jr.
    Miller, D. Craig
    [J]. JOURNAL OF HEART VALVE DISEASE, 2008, 17 (02): : 168 - 177
  • [22] Transcatheter Mitral Valve Repair With MitraClip for Symptomatic Functional Mitral Valve Regurgitation
    Mendirichaga, Rodrigo
    Singh, Vikas
    Blumer, Vanessa
    Rivera, Manuel
    Rodriguez, Alex P.
    Cohen, Mauricio G.
    O'Neill, William W.
    Elmariah, Sammy
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2017, 120 (04): : 708 - 715
  • [23] Use of the Mitraclip system for percutaneous mitral repair in patients with functional mitral regurgitation
    Hoffmann, R.
    Reith, S.
    Altiok, E.
    Marx, N.
    Brehmer, K.
    [J]. EUROPEAN HEART JOURNAL, 2010, 31 : 663 - 663
  • [25] Reduction of mitral valve regurgitation with Mitraclip® percutaneous system
    Tamburino, C.
    Imme, S.
    Barbanti, M.
    Mule, M.
    Pistritto, A. M.
    Aruta, P.
    Cammalleri, V.
    Scarabelli, M.
    Mangiafico, S.
    Scandura, S.
    Ussia, G. P.
    [J]. MINERVA CARDIOANGIOLOGICA, 2010, 58 (05): : 589 - 598
  • [26] Immediate effect of the MitraClip® procedure on mitral ring geometry in primary and secondary mitral regurgitation
    Schmidt, Frank Patrick
    von Bardeleben, Ralph Stephan
    Nikolai, Philipp
    Jabs, Alexander
    Wunderlich, Nina
    Muenzel, Thomas
    Hink, Ulrich
    Warnholtz, Ascan
    [J]. EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2013, 14 (09) : 851 - 857
  • [27] Clinical outcomes of MitraClip for the treatment of functional mitral regurgitation
    Taramasso, Maurizio
    Maisano, Francesco
    Latib, Azeem
    Denti, Paolo
    Buzzatti, Nicola
    Cioni, Micaela
    La Canna, Giovanni
    Colombo, Antonio
    Alfieri, Ottavio
    [J]. EUROINTERVENTION, 2014, 10 (06) : 746 - 752
  • [28] Impact of MitraClip According to Proportionality of Functional Mitral Regurgitation
    Ooms, Joris
    Geleijnse, Marcel
    Spitzer, Ernest
    Van Wiechen, Maarten
    Van Mieghem, Nicolas
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (04) : S57 - S57
  • [29] Benefit of the MitraClip for mitral regurgitation
    Lim G.B.
    [J]. Nature Reviews Cardiology, 2018, 15 (12) : 728 - 728
  • [30] Reverse Remodeling Observed Following Percutaneous Mitral Annuloplasty for Functional Mitral Regurgitation in the TITAN™ Trial
    Myc, Jacek
    Siminiak, Tomasz
    Lipiecki, Janusz
    Hoppe, Uta C.
    Haude, Michael
    Sadowski, Jerzy
    Fajadet, Jean
    Shah, Amil
    Mehta, Nawzer
    Reuter, David G.
    Wu, Justina
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (13) : B116 - B116