Finite element analysis on proof testing of ceramic ball heads for total hip replacements: Trauma loading considerations

被引:0
|
作者
Carrillo, D. G. [1 ]
Curiel, M. C.
Aum, H. S.
Lee, S. W.
机构
[1] Sun Moon Univ, Dept Mat Engn, Asan 336708, Chungnam, South Korea
[2] Sun Moon Univ, Div Mech & Control Engn, Asan 336708, Chungnam, South Korea
来源
关键词
ceramic; trauma; finite element analysis; contact; proof test;
D O I
10.4028/www.scientific.net/KEM.317-318.773
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Trauma related loading of hip prostheses has been reported as a cause of failure of implanted ceramic materials, ball and/or cup. In this study a computer simulated model is made to evaluate contact pressures and stresses in the stem/ball contact pair, as well as stress concentration regions in the ceramic cup. A 3D Non-Linear Finite Element Analysis was performed in a "proof test" scenario. A high peak trauma-like loading of 15 KN was used. 28mm ball heads (in diameter) were modeled and two sizes for taper lock were simulated to report their influence on the contact pressure over the critical areas of the ballhead. It was found that the ball head cone's depth has a major incidence in the stress and contact pressure surrounding the stem when an impact load is applied. The best stress and pressure configurations were found for the case of the stem lying completely inside the ballhead.
引用
收藏
页码:773 / 776
页数:4
相关论文
共 50 条
  • [1] Stress concentration regions in ceramic ball heads for total hip replacement considering trauma-like loading
    Aum, HS
    Curiel, MC
    Carrillo, DG
    [J]. ECO-MATERIALS PROCESSING & DESIGN VI, 2005, 486-487 : 185 - 188
  • [2] Titanium deposits on the ceramic heads of dislocated total hip replacements
    Schuh, A
    Holzwarth, U
    Kachler, W
    Göske, J
    Zeiler, G
    [J]. ORTHOPADE, 2004, 33 (10): : 1194 - 1199
  • [3] Proof testing of ceramic femoral heads for hip joint implants
    Weisse, B.
    Affolter, C.
    Koller, R. E.
    Stutz, A.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H9) : 1051 - 1059
  • [4] Wear in alumina-on-alumina ceramic total hip replacements A RETRIEVAL ANALYSIS OF EDGE LOADING
    Esposito, C. I.
    Walter, W. L.
    Roques, A.
    Tuke, M. A.
    Zicat, B. A.
    Walsh, W. R.
    Walter, K.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (07): : 901 - 907
  • [5] Mesh morphing for finite element analysis of implant positioning in cementless total hip replacements
    Bah, Mamadou T.
    Nair, Prasanth B.
    Browne, Martin
    [J]. MEDICAL ENGINEERING & PHYSICS, 2009, 31 (10) : 1235 - 1243
  • [6] Mesh morphing for finite element analysis of implant positioning in cementless total hip replacements
    School of Engineering Sciences, University of Southampton, Southampton, SO17 1BJ, United Kingdom
    [J]. Med. Eng. Phys., 1600, 10 (1235-1243):
  • [7] The occurrence of squeaking under wear testing standards for ceramic on ceramic total hip replacements
    Lancaster-Jones, Oscar O'Dwyer
    Reddiough, Rebecca
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 138
  • [8] Fracture analysis of ceramic heads for total hip replacement
    Willmann, G
    [J]. MATERIALWISSENSCHAFT UND WERKSTOFFTECHNIK, 2001, 32 (07) : 591 - 600
  • [9] A computational parametric study on edge loading in ceramic-on-ceramic total hip joint replacements
    Liu, Feng
    Feng, Li
    Wang, Junyuan
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 83 : 135 - 142
  • [10] Supplementary finite element analysis in experimental testing of total hip stems
    Kluess, Daniel
    Lange, Hans-E
    Heyer, Horst
    Sander, Manuela
    Mittelmeier, Wolfram
    Bader, Rainer
    [J]. MATERIALS TESTING, 2018, 60 (05) : 489 - 494