Stress concentration regions in ceramic ball heads for total hip replacement considering trauma-like loading

被引:1
|
作者
Aum, HS [1 ]
Curiel, MC
Carrillo, DG
机构
[1] Sun Moon Univ, Dept Mech & Control Engn, Asan 336708, Chungnam, South Korea
[2] SunMoon Univ, Dept Mat Engn, Asan 336708, Chungnam, South Korea
来源
关键词
ceramic ball head; trauma; finite element; contact analysis;
D O I
10.4028/www.scientific.net/MSF.486-487.185
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A high stress condition in the hip system may cause fracture of the ball head. This failure may appear after a heavy accident such as sudden fall. The aim of this investigation is to make a computer simulated model of the hip system to evaluate the regions of stress concentration as well as the pressure in the stem-ball junction. 3D Non-Linear Finite Element Analyses were performed taking into consideration a high peak load to simulate trauma conditions. Ball heads from 22 to 36 mm in diameter were modeled, and also two sizes for taper lock were simulated to report their influence on the stresses over the critical areas of the ball head. Two different materials of common ball head ceramics (Alumina and Zirconia) were considered to evaluate its relation to the stresses produced. It was found that the ball head cone's depth has major incidence in the stress concentration surrounding the stem when an impact load is applied, and that a deeper cone may offer a more relieved loading configuration when considering stress related parameters such as Von Mises stress, contact pressure and friction stress.
引用
收藏
页码:185 / 188
页数:4
相关论文
共 16 条
  • [1] Finite element analysis on proof testing of ceramic ball heads for total hip replacements: Trauma loading considerations
    Carrillo, D. G.
    Curiel, M. C.
    Aum, H. S.
    Lee, S. W.
    [J]. SCIENCE OF ENGINEERING CERAMICS III, 2006, 317-318 : 773 - 776
  • [2] Fracture of ceramic heads in total hip replacement
    Habermann, B.
    Ewald, W.
    Rauschmann, M.
    Zichner, L.
    Kurth, A. A.
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2006, 126 (07) : 464 - 470
  • [3] Fracture of ceramic heads in total hip replacement
    B. Habermann
    W. Ewald
    M. Rauschmann
    L. Zichner
    A. A. Kurth
    [J]. Archives of Orthopaedic and Trauma Surgery, 2006, 126
  • [4] Fracture analysis of ceramic heads for total hip replacement
    Willmann, G
    [J]. MATERIALWISSENSCHAFT UND WERKSTOFFTECHNIK, 2001, 32 (07) : 591 - 600
  • [5] Improvement of the safety of ceramic femoral heads for total hip replacement
    Willmann, G
    Pfaff, HG
    Richter, HG
    [J]. BIOMEDIZINISCHE TECHNIK, 1995, 40 (12): : 342 - 346
  • [6] Influence of Different Damage Patterns of the Stem Taper on Fixation and Fracture Strength of Ceramic Ball Heads for Total Hip Replacement
    Vogel, Danny
    Hembus, Jessica
    Jackszis, Mario
    Bolte, Vera
    Bader, Rainer
    [J]. BIOMED RESEARCH INTERNATIONAL, 2020, 2020
  • [7] Clinical outcomes of ceramicized ball heads in total hip replacement bearings: a literature review
    Piconi, Corrado
    De Santis, Vincenzo
    Maccauro, Giulio
    [J]. JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2017, 15 (01): : E1 - E9
  • [8] Combined wear and ageing of ceramic-on-ceramic bearings in total hip replacement under edge loading conditions
    Al-Hajjar, Mazen
    Gremillard, Laurent
    Begand, Sabine
    Oberbach, Thomas
    Hans, Karen
    Delfosse, Daniel
    Chevalier, Jerome
    Jennings, Louise M.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 98 : 40 - 47
  • [9] Tribological and material analyses of retrieved alumina and zirconia ceramic heads correlated with polyethylene wear after total hip replacement
    Kim, Y. -H.
    Kim, J. -S.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (06): : 731 - 737
  • [10] Survival of ceramic bearings in total hip replacement after high-energy trauma and periprosthetic acetabular fracture
    Salih, S.
    Currall, V. A.
    Ward, A. J.
    Chesser, T. J. S.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (11): : 1533 - 1535