Simultaneous determination of seven impurities in high-purity cobalt oxide by ICP-AES after matrix separation using 1-nitroso-2-naphthol as a precipitant

被引:7
|
作者
Wang, Y [1 ]
Zhou, CS
Duan, QB
Chen, P
机构
[1] Cent S Univ, Coll Chem & Chem Engn, Changsha 410083, Peoples R China
[2] Baling Petrochem Ltd Co, Qual Inspect Ctr Hexanelactam Plant, Yueyang 414003, Peoples R China
来源
关键词
trace impurity; high-purity cobalt oxide; matrix separation; inductively coupled plasma atomic emission spectrometry (ICP-AES);
D O I
10.1007/s11771-002-0051-x
中图分类号
TF [冶金工业];
学科分类号
0806 ;
摘要
A method was developed for the simultaneous determination of seven trace impurities (Cd, Mn. Pb, Zn, Cu, Fe and Ni) in high-purity cobalt oxide by ICP-AES. The matrix effect was eliminated by precipitation with 1-nitroso-2-naphthol. The matrix effect of cobalt on the absorptions of trace impurities, the effects of reaction time, pH value, dosage of precipitant on the formation of cobalt-1-nitroso-2-naphthol complex, the effects of hydrochloric acid on the stability of this complex and masking of elements were studied. Recoveries of the impurities in spiked sample are from 90% to 110% with a precision of 1.1%-5.0% RSD. The detection limits of the seven elements are in the range of 0.01-0.24 mug/g. ne method can be applied to the analysis of high-purity cobalt metal, cobalt oxide and other cobalt compounds.
引用
收藏
页码:100 / 103
页数:4
相关论文
共 50 条
  • [31] DETERMINATION OF Y, SM, EU, GD, DY, HO, ER IN HIGH-PURITY TERBIUM OXIDE BY ICP-AES
    BISWAS, SS
    SETHUMADHAVAN, A
    MURTY, PS
    [J]. MIKROCHIMICA ACTA, 1991, 1 (1-2) : 71 - 77
  • [32] Column Preconcentration and Determination of Cobalt(II) Using Silica Gel Loaded with 1-Nitroso-2-naphthol
    Shin, Eun Mi
    Choi, Hee-Seon
    [J]. BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2009, 30 (07) : 1516 - 1520
  • [33] EXTRACTION-PHOTOMETRIC DETERMINATION OF COBALT IN HIGH-TEMPERATURE ALLOYS AND URANIUM WITH 1-NITROSO-2-NAPHTHOL
    HASHITANI, H
    YOSHIDA, H
    ADACHI, T
    [J]. BUNSEKI KAGAKU, 1975, 24 (07) : 452 - 454
  • [34] DETERMINATION OF TRACE AMOUNTS OF IMPURITIES IN HIGH-PURITY SILICON AND SILICATE MATERIALS BY VAPOR-PHASE PRESSURED DECOMPOSITION ICP-AES AND ICP-MS
    FUJIMOTO, K
    OKANO, T
    [J]. BUNSEKI KAGAKU, 1993, 42 (10) : T135 - T142
  • [35] Fabrication of a New Selective Optical Sensor for Cobalt Determination Using 1-Nitroso-2-naphthol and Monitoring of Cobalt in Water Samples
    Darroudi, Abolfazl
    Kakhki, Javad Fadaee
    Mofrad, Issa Zakeri
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY RESEARCH, 2022, 9 (04): : 443 - 449
  • [36] Determination of 233U Using High Resolution ICP-AES After Chemical Separation
    Kumar, Mithlesh
    Rajeswari, B.
    Thulasidas, S. K.
    Kadam, R. M.
    [J]. ATOMIC SPECTROSCOPY, 2016, 37 (04) : 150 - 157
  • [37] FLAMELESS ATOMIC-ABSORPTION DETERMINATION OF IMPURITIES IN HIGH-PURITY SILVER AFTER EXTRACTIVE SEPARATION OF THE MATRIX
    STARTSEVA, EA
    POPOVA, NM
    YUDELEVICH, IG
    VANIFATOVA, NG
    ZOLOTOV, YA
    [J]. FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1980, 300 (01): : 28 - 32
  • [38] Separation with Ion Exchange Fiber Column and Determination of La, Nd, Eu and Gd in High Purity Ytterbium Oxide by ICP-AES
    Gong Qi
    Chen Jie
    Ji Ri-wen
    Pan Xue-zhen
    Wu Juan
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30 (02) : 523 - 527
  • [39] DETERMINATION OF TRACE-METALS IN HIGH-PURITY SILICA, SILICON-NITRIDE AND SILICON-CARBIDE BY ICP-AES AFTER SEPARATION WITH MINICOLUMN CHELATING RESIN
    YOSHIKAWA, H
    ISHIBASHI, Y
    GUNJI, N
    SASAYAMA, K
    MISUMI, T
    [J]. BUNSEKI KAGAKU, 1990, 39 (06) : T83 - T87
  • [40] AN ICP-AES METHOD FOR THE DETERMINATION OF HEAVY RARE-EARTH ELEMENTS (EU-LU) IN HIGH-PURITY YTTRIUM-OXIDE
    MARATHE, SM
    BISWAS, SS
    PATIL, PB
    MURTY, PS
    [J]. MIKROCHIMICA ACTA, 1992, 109 (5-6) : 261 - 268