Urinalysis: The Automated Versus Manual Techniques; Is It Time To Change?

被引:7
|
作者
Ahmed, Asmaa Ismail [1 ]
Baz, Heba [1 ]
Lotfy, Sarah [1 ]
机构
[1] Cairo Univ, Fac Med, Clin & Chem Pathol Dept, Kasr Al Aini St, Cairo, Egypt
关键词
urine; automation; H800-FUS100; evaluation; URINE SEDIMENT EXAMINATION; MICROSCOPY; ANALYZERS;
D O I
10.7754/Clin.Lab.2015.150520
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Urinalysis is the third major test in clinical laboratory. Manual technique imprecision urges the need for a rapid reliable automated test. We evaluated the H800-FUS100 automatic urine sediment analyzer and compared it to the manual urinalysis technique to determine if it may be a competitive substitute in laboratories of central hospitals. Methods: 1000 urine samples were examined by the two methods in parallel. Agreement, precision, carry-over, drift, sensitivity, specificity, and practicability criteria were tested. Results: Agreement ranged from excellent to good for all urine semi-quantitative components (K > 0.4, p = 0.000), except for granular casts (K = 0.317, p = 0.000). Specific gravity results correlated well between the two methods (r = 0.884, p = 0.000). RBCS and WBCs showed moderate correlation (r = 0.42, p = 0.000) and (r = 0.44, p = 0.000), respectively. The auto-analyzer's within-run precision was >= 75% for all semi-quantitative components except for proteins (50% precision). This finding in addition to the granular casts poor agreement indicate the necessity of operator interference at the critical cutoff values. As regards quantitative contents, RBCs showed a mean of 69.8 +/- 3.95, C.V. = 5.7, WBCs showed a mean of 38.9 +/- 1.9, C.V. = 4.9). Specific gravity, pH, microalbumin, and creatinine also showed good precision results with C.Vs of 0.000, 2.6, 9.1, and 0.00 respectively. In the between run precision, positive control showed good precision (C.V. = 2.9), while negative control's C.V. was strikingly high (C.V. = 127). Carryover and drift studies were satisfactory. Manual examination of inter observer results showed major discrepancies (< 60% similar readings), while intra-observer's results correlated well with each other (r = 0.99, p = 0.000). Conclusions: Automation of urinalysis decreases observer-associated variation and offers prompt competitive results when standardized for screening away from the borderline cutoffs.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 50 条
  • [1] COMPARISON OF TURNAROUND TIMES OF MANUAL VERSUS AUTOMATED URINALYSIS
    BAIN, OF
    TRENT, Y
    SACHER, RA
    MCPHERSON, RA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1988, 89 (03) : 438 - 438
  • [2] COMPARISON OF AUTOMATED AND MANUAL METHODS FOR URINALYSIS
    ELIN, RJ
    HOSSEINI, JM
    KESTNER, J
    RAWE, M
    RUDDEL, M
    NISHI, HH
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1190 - 1190
  • [3] COMPARISON OF AUTOMATED AND MANUAL METHODS FOR URINALYSIS
    ELIN, RJ
    HOSSEINI, JM
    KESTNER, J
    RAWE, M
    RUDDEL, M
    NISHI, HH
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1986, 86 (06) : 731 - 737
  • [4] AUTOMATED URINALYSIS RESULTS COMPARED WITH MANUAL PROCEDURES
    HAGER, CB
    FREE, AH
    BOTERO, JM
    BROWN, JR
    CLINICAL CHEMISTRY, 1972, 18 (07) : 691 - &
  • [5] AUTOMATED VERSUS MANUAL TECHNIQUES FOR PROTHROMBIN TIME TEST - RESULTS OF PROFICIENCY ASSESSMENT STUDIES
    POLLER, L
    THOMSON, JM
    YEE, KF
    BRITISH JOURNAL OF HAEMATOLOGY, 1978, 38 (03) : 391 - 399
  • [6] Evaluation of automated urinalysis vs manual microscopy.
    Holleman, CM
    Sedor, FA
    CLINICAL CHEMISTRY, 2000, 46 (06) : A30 - A30
  • [7] Comparison of manual urinalysis with automated yellow IRIS urinalysis in evaluation of hematuria in children.
    Saberi, MS
    Khan, SM
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2005, 45 (04) : A46 - A46
  • [8] Assessing Concordance of Results: A Comparative Study of the Manual and Automated Urinalysis Methods
    Gyamfi, Nicholas Kwame Afriyie
    Osei, George Nkrumah
    Brenyah, Ruth C.
    Agyemang, Lawrence Duah
    Ampomah, Paulina
    Darkwah, Kwame Osei
    Toboh, Emmanuel
    Ephraim, Richard K. D.
    BIOMED RESEARCH INTERNATIONAL, 2024, 2024
  • [9] Urine sediment examination: A comparison of automated urinalysis systems and manual microscopy
    Chien, Tzu-I
    Kao, Jau-Tsuen
    Liu, Hui-Lan
    Lin, Po-Chang
    Hong, Jhih-Sian
    Hsieh, Han-Peng
    Chien, Miao-Ju
    CLINICA CHIMICA ACTA, 2007, 384 (1-2) : 28 - 34
  • [10] Screening dipstick urinalysis: A time to change
    Kaplan, RE
    Springate, JE
    Feld, LG
    PEDIATRICS, 1997, 100 (06) : 919 - 921