Dynamic comparison between Daan real-time PCR and Cobas TaqMan for quantification of HBV DNA levels in patients with CHB

被引:41
|
作者
Cai, Shao-Hang [1 ]
Lv, Fang-Fang [2 ]
Zhang, Yong-Hong [3 ]
Jiang, Ye-Gui [4 ]
Peng, Jie [1 ]
机构
[1] Southern Med Univ, NanFang Hosp, Guangzhou, Guangdong, Peoples R China
[2] Zhejiang Univ, Sir Run Run Shaw Hosp, Hangzhou 310003, Zhejiang, Peoples R China
[3] Cent S Univ, Xiangya Hosp 2, Changsha, Hunan, Peoples R China
[4] Third Mil Med Univ, Southwest Hosp, Chongqing, Peoples R China
来源
BMC INFECTIOUS DISEASES | 2014年 / 14卷
关键词
HBV DNA quantification; Daan Test; Cobas TaqMan; Hepatitis B virus; Viral load; Serum; HBeAg; CHRONIC HEPATITIS-B; VIRUS DNA; CLINICAL-EVALUATION; MONITOR TEST; E-ANTIGEN; MANAGEMENT; LAMIVUDINE; AMPLIFICATION; QUANTITATION; RESISTANCE;
D O I
10.1186/1471-2334-14-85
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: Hepatitis B virus (HBV) DNA levels are crucial for managing chronic hepatitis B (CHB). It was unclear whether Daan real-time polymerase chain reaction test (Daan test) or COBAS TaqMan HBV DNA Test (Cobas TaqMan) was superior in measuring different HBV DNA levels in clinical specimens. Methods: We enrolled 67 treatment-naive, HBV surface antigen-positive CHB patients (high baseline viral levels) who received either lamivudine/adefovir or entecavir. Serum samples were tested at baseline and treatment week 24 using the Daan test and Cobas TaqMan. Results: In the 67-baseline samples, the HBV DNA levels with the Cobas TaqMan (7.90 +/- 0.73 log(10) IU/mL) were significantly greater than those of the Daan test (7.11 +/- 0.44 log(10) IU/mL; P < 0.001). Of the 67 24-week samples (low viral levels), the Cobas TaqMan detected 59 (88.1%; 8 undetected); the Daan test detected 33 (49.3%; 34 undetected; P < 0.001). The Cobas TaqMan detected HBV DNA in 26 of 34 samples undetectable by the Daan test (range, 1.4-3.7 log(10) IU/mL) or 38% of samples (26/67). The reductions in viral load after 24 weeks of oral antiviral treatment in the 33 samples that were positive for both the Daan test and the Cobas TaqMan test were significantly different (3.59 +/- 1.11 log(10) IU/mL versus 4.87 +/- 1.58 log(10) IU/mL, respectively; P = 0.001). Spearman correlation analysis showed positive correlation between results from two tests (r(p) = 0.602, P<0.001). The HBV genotypes and the anti-viral treatment did not affect the measurements of the HBV DNA by the Daan assay and the Cobas Taqman assay. Conclusion: The Cobas Taqman was more sensitive at low viral loads than the Daan test and the change from complete to partial virological response could affect clinical decisions. The Cobas Taqman may be more appropriate for detection of HBV DNA levels.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Comparison of PNA probe-based real-time PCR and Cobas TaqMan MTB for detection of MTBC
    Jae Woo Kim
    Young Jin Choi
    Hwi Jun Kim
    Joon Soo Park
    Hae Seon Nam
    Young Hwangbo
    Dong Uk Kim
    Kwi Sung Park
    BioChip Journal, 2013, 7 : 85 - 88
  • [22] Comparison of PNA probe-based real-time PCR and Cobas TaqMan MTB for detection of MTBC
    Kim, Jae Woo
    Choi, Young Jin
    Kim, Hwi Jun
    Park, Joon Soo
    Nam, Hae Seon
    Hwangbo, Young
    Kim, Doug Uk
    Park, Kwi Sung
    BIOCHIP JOURNAL, 2013, 7 (02) : 85 - 88
  • [23] Comparison of detection and quantification of HBV DNA in chronic HBeAg negative and positive patients by Abbott RealTime HBV and Roche Cobas TaqMan HBV assays
    Morris, Carol J.
    Hill, Mary
    de Medina, Maria
    Herman, Christine
    Cloherty, Gavin A.
    Martin, Paul
    JOURNAL OF VIROLOGICAL METHODS, 2013, 193 (02) : 391 - 393
  • [24] Real-time PCR quantification of haematopoietic chimerism after transplantation: a comparison between TaqMan and hybridization probes technologies
    Martinez-Lopez, J.
    Crooke, A.
    Grande, S.
    Ayala, R.
    Jimenez-Velasco, A.
    Gamarra, S.
    Meneu, J. C.
    Gilsanz, F.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2010, 32 (01) : E17 - E25
  • [26] A TaqMan real-time PCR system for the identification and quantification of bovine DNA in meats, milks and cheeses
    Zhang, Chun-Lai
    Fowler, Mark R.
    Scott, Nigel W.
    Lawson, Graham
    Slater, Adrian
    FOOD CONTROL, 2007, 18 (09) : 1149 - 1158
  • [27] Comparison of real-time PCR assays for monitoring serum HBV DNA levels during antiviral therapy
    Zhang, Hai-Ying
    Hui, Chee-Kin
    Wong, Anita
    Lee, Nikki P.
    Leung, Nancy
    Luk, John M.
    Lau, George K. K.
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2006, 21 : A140 - A140
  • [28] Comparison of a novel real-time PCR method (RTA) and Artus RG for the quantification of HBV DNA and HCV RNA
    Aydogan, Sibel
    Acikgoz, Ziya Cibali
    Gozalan, Aysegul
    Kirca, Fisun
    Muderris, Tuba
    Ozturk, Elife
    JOURNAL OF INFECTION IN DEVELOPING COUNTRIES, 2017, 11 (07): : 543 - 548
  • [29] REAL-TIME PCR QUANTIFICATION OF SERUM AND INTRACELLULAR HBV DNA AND HDV RNA IN DIFFERENT SUBSETS OF CHRONIC HBV INFECTED PATIENTS
    Pollicino, T.
    Raffa, G.
    Iannello, G.
    Santantonio, T.
    Gaeta, G. B.
    Squadrito, G.
    Colucci, G.
    Calvi, C.
    Cacciola, I.
    Levrero, M.
    Raimondo, G.
    JOURNAL OF HEPATOLOGY, 2009, 50 : S212 - S212
  • [30] Evaluation of an automated, highly sensitive, real time PCR based assay (COBAS Ampliprep/COBAS TaqMan) for quantification of HCV RNA
    Sarrazin, Christoph
    Wohnsland, Anette
    Gartner, Barbara
    Traver, Stella
    Colucci, Guiseppe
    Zeuzem, Stefan
    Valsamakis, Alexandra