A chloride ion-selective solvent polymeric membrane electrode based on a hydrogen bond forming ionophore

被引:152
|
作者
Xiao, KP [1 ]
Buhlmann, P [1 ]
Nishizawa, S [1 ]
Amemiya, S [1 ]
Umezawa, Y [1 ]
机构
[1] UNIV TOKYO,SCH SCI,DEPT CHEM,TOKYO 113,JAPAN
关键词
D O I
10.1021/ac961035d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A chloride-selective solvent polymeric membrane electrode based on a neutral hydrogen bond forming ionophore, poly(vinyl chloride) (PVC), the plasticizer 2-nitrophenyl octyl ether (o-NPOE), and cationic sites (50 mol %, relative to the ionophore) is described. At pH 7.0 (HEPES buffer), this electrode responds to chloride in a linear range from 10(-5) to 10(-2) M with a slope of (-54.0 +/- 1.0) mV/decade and a detection limit of (6.5 +/- 3.0) x 10(-6) M. As compared to conventional anion-exchanger electrodes, the interference of SCN-, Br-, NO3-, I-, and even salicylate is considerably reduced, as shown by the selectivity coefficients determined with the matched potential method (MPM) in the chloride concentration range 10(-2.34)-10(-2.04) M (log KPCl,jpot (MPM): Sal(-), +1.8; SCN-, +1.6; NO3-, +0.7; I-, +0.5; Br-, +0.4). Because the discrimination of the more hydrophilic anions SO42-, HSO3-/SO32-, OAc-, HCO3-, and H2PO4-/HPO42- is too large for a determination of accurate selectivity coefficients in this high chloride concentration range, corresponding values for the range 10(-5.00)-10(-4.70) M Cl- have been determined (log K-Cl,j(pot) (MPM): SO42-, -1.2; HSO3-/SO32-, -2.0; OAC-, -2.3; HCO3-, -2.6; H2PO4-/HPO42-, < -3.5), showing here, too, a good selectivity for Cl-. The chloride ion concentration in certified control horse serum was determined by the standard addition method as 102.1 mM with a coefficient of variation of 0.42% (n = 6). This result is in excellent agreement with the certified value of 102.3 mM, as determined by coulometry, and demonstrates the applicability of this sensor for measurements in biological samples. The selectivity upon storage of the electrode in buffer solution for 4 weeks decreases but is stilt considerably better than that for a freshly prepared anion-exchanger electrode.
引用
收藏
页码:1038 / 1044
页数:7
相关论文
共 50 条
  • [1] A hydrogen ion-selective poly(vinyl chloride) membrane electrode based on calix[4]arene as a perchlorate ion-selective electrode
    Canel, Esin
    Erden, Sevcan
    Ozel, Ayca Demirel
    Memon, Sahahabuddin
    Yilmaz, Mustafa
    Kilic, Esma
    TURKISH JOURNAL OF CHEMISTRY, 2008, 32 (03) : 323 - 332
  • [2] COMBINATION ION-SELECTIVE ELECTRODE BASED ON SOLVENT POLYMERIC MEMBRANES
    DOHNER, RE
    SIMON, W
    ANALYTICAL LETTERS, 1979, 12 (03) : 205 - 212
  • [3] Using of hydrogen ion-selective poly(vinyl chloride) membrane electrode based on calix[4]arene as thiocyanate ion-selective electrode
    Erden, S
    Demirel, A
    Memon, S
    Yilmaz, M
    Canel, E
    Kiliç, E
    SENSORS AND ACTUATORS B-CHEMICAL, 2006, 113 (01) : 290 - 296
  • [4] SOLVENT POLYMERIC MEMBRANE ION-SELECTIVE ELECTRODES
    THOMAS, JDR
    ANALYTICA CHIMICA ACTA, 1986, 180 : 289 - 297
  • [5] Hydrogen ion-selective poly(vinyl chloride) membrane electrode based on a calix[4]arene
    Kuruoglu, D
    Canel, E
    Memon, S
    Yilmaz, M
    Kiliç, E
    ANALYTICAL SCIENCES, 2003, 19 (02) : 217 - 221
  • [6] Hydrogen Ion-Selective Poly(vinyl chloride) Membrane Electrode Based on a Calix[4]arene
    Demet Kuruoğlu
    Esin Canel
    Shahabuddin Memon
    Mustafa Yilmaz
    Esma Kiliç
    Analytical Sciences, 2003, 19 : 217 - 221
  • [7] CHLORATE ION-SELECTIVE ELECTRODE BASED ON A POLYVINYL CHLORIDE MATRIX MEMBRANE
    HIIRO, K
    MOODY, GJ
    THOMAS, JDR
    TALANTA, 1975, 22 (10-1) : 918 - 919
  • [8] Polymeric Membrane Ion-Selective Electrode for Butyrylcholinesterase Based on Controlled Release of Substrate
    Ding, Jiawang
    Qin, Wei
    ELECTROANALYSIS, 2009, 21 (17-18) : 2030 - 2035
  • [9] EVALUATION OF AN AMMONIUM IONOPHORE FOR USE IN POLY(VINYL CHLORIDE) MEMBRANE ION-SELECTIVE ELECTRODES - SOLVENT MEDIATOR EFFECTS
    GHAURI, MS
    THOMAS, JDR
    ANALYST, 1994, 119 (11) : 2323 - 2326
  • [10] Triiodide ion-selective polymeric membrane electrode based on a ketoconazole-triiodide ion pair
    Farhadi, K
    Maleki, R
    Shamsipur, M
    ELECTROANALYSIS, 2002, 14 (11) : 760 - 766