Irradiation with x-rays of the energy 18 MV induces radioactivity in transfusion blood: Proposal of a safe method using 6 MV

被引:7
|
作者
Frentzel, Katharina [1 ]
Badakhshi, Harun [1 ]
机构
[1] Ernst von Bergmann Med Ctr, Dept Radiat Oncol, Charlottenstr 72, D-14467 Potsdam, Germany
关键词
induced radioactivity; x-rays; linear accelerator; blood; transfusion-associated graft-versus-host disease; VERSUS-HOST-DISEASE; GAMMA-IRRADIATION; IMPLEMENTATION; COMPONENTS;
D O I
10.1118/1.4967482
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prevent a fatal transfusion-associated graft-versus-host disease, it is recommended to irradiate transfusion blood and blood components with ionizing radiation. Using x-rays from a linear accelerator of the radiotherapy department is an accepted alternative to gamma irradiation devices of the blood bank and to the orthovoltage units that are replacing the gamma irradiators today. However, the use of high energy x-rays may carry a potential risk of induced radioactivity. The objective of this study was to investigate the effect of two different energy levels, 6 and 18 MV, which are executed in routine clinical settings. The research question was if induced radioactivity occurs at one of these standard energy levels. The authors aimed to give a proposal for a blood irradiation procedure that certainly avoids induced radioactivity. Methods: For this study, the authors developed a blood bag phantom, irradiated it with x-ray energies of 6 and 18 MV, and measured the induced radioactivity in a well counter. Thereafter, the same irradiation and measuring procedure was performed with a unit of packed red blood cells. A feasible clinical procedure was developed using 6 MV and an acrylic box. With the irradiation planning system XiO, the authors generated an irradiation protocol for the linear accelerator Siemens ONCOR Anvant-Garde. Results: Both measurement setups showed that there was induced radioactivity for 18 MV but not for 6 MV. The induced radioactivity for 18 MV was up to 190 times the background. This is significant and of clinical relevance especially since there are newborn and fetal blood recipients for whom every radiation exposure has to be strictly avoided. Conclusions: The irradiation of blood with x-rays from a linear accelerator of the radiotherapy department is safe and feasible, but by the current state of scientific knowledge, the authors recommend to use an x-ray energy of 6 MV or less to avoid induced radioactivity in transfusion blood. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:6517 / 6524
页数:8
相关论文
共 50 条
  • [1] Refined gypsum effective attenuation coefficients for Co-60, 6 MV, 10 MV and 18 MV x-rays
    Glasgow, GP
    Dogan, N
    Sethi, A
    Stanton, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (01): : 63 - 71
  • [2] CYLINDRICAL ION-CHAMBER REPLACEMENT FACTORS FOR 6-MV, 18-MV, AND 25-MV X-RAYS
    MAHOOD, SH
    ALMOND, PR
    [J]. MEDICAL PHYSICS, 1983, 10 (04) : 539 - 539
  • [3] Detecting 6 MV X-rays using an organic photovoltaic device
    Kingsley, James W.
    Pearson, Andrew J.
    Harris, Lee
    Weston, Steven J.
    Lidzey, David G.
    [J]. ORGANIC ELECTRONICS, 2009, 10 (06) : 1170 - 1173
  • [4] LUNG DOSE CORRECTIONS FOR 6-MV AND 15-MV X-RAYS
    MACKIE, TR
    ELKHATIB, E
    BATTISTA, J
    SCRIMGER, J
    VANDYK, J
    CUNNINGHAM, JR
    [J]. MEDICAL PHYSICS, 1985, 12 (03) : 327 - 332
  • [5] BROAD BEAM ATTENUATION OF CO-60 GAMMA-RAYS AND 6-MV, 18-MV, AND 25-MV X-RAYS BY LEAD
    VANDYK, J
    [J]. MEDICAL PHYSICS, 1986, 13 (01) : 105 - 110
  • [6] Radiation transmission of colemanite, tincalconite and ulexite for 6 and 18 MV X-rays by using linear accelerator
    Demir, Faruk
    Un, Adem
    [J]. APPLIED RADIATION AND ISOTOPES, 2013, 72 : 1 - 5
  • [7] Radiation transmission of heavyweight and normal-weight concretes containing colemanite for 6 MV and 18 MV X-rays using linear accelerator
    Demir, F.
    Budak, G.
    Sahin, R.
    Karabulut, A.
    Oltulu, M.
    Serifoglu, K.
    Un, A.
    [J]. ANNALS OF NUCLEAR ENERGY, 2010, 37 (03) : 339 - 344
  • [8] LUNG CORRECTION FACTORS FOR RADIOTHERAPY ISODOSE PLANNING USING 6MV AND 10MV X-RAYS
    CONSTANTINOU, C
    HARRINGTON, J
    [J]. MEDICAL PHYSICS, 1987, 14 (03) : 463 - 463
  • [9] COMPENSATING MISSING TISSUE FOR 6MV AND 10MV X-RAYS USING SOLID WATER AND LEAD
    CONSTANTINOU, C
    HARRINGTON, J
    [J]. MEDICAL PHYSICS, 1987, 14 (03) : 481 - 481
  • [10] CHARACTERISTICS OF CLINAC 18 WEDGED FIELDS FOR 10 MV X-RAYS
    TOLBERT, DD
    PALIWAL, B
    FROST, S
    CHIN, KB
    DERCKS, J
    LANE, RG
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 570 - 570