Scatter radiation reduction with a radiation-absorbing pad in interventional radiology examinations

被引:21
|
作者
Koenig, A. M. [1 ]
Maas, J. [1 ]
Viniol, S. [1 ]
Etzel, R. [2 ]
Fiebich, M. [1 ,2 ]
Thomas, R. P. [1 ]
Mahnken, A. H. [1 ]
机构
[1] Philipps Univ Marburg, Univ Hosp Marburg, Clin Diagnost & Intervent Radiol, Marburg, Germany
[2] Mittelhessen Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
关键词
Radiation protection; Interventional radiology; Radiation-absorbing pad; MEDICAL STAFF; CANCER-RISKS; PROTECTION; EXPOSURE; OPERATOR; DOSIMETRY; EFFICACY; SHIELD;
D O I
10.1016/j.ejrad.2020.109245
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Radiation-absorbing pads are an additional possibility to reduce scattered radiation at its source. The goal of this study is to investigate the efficacy of a new reusable radiation-absorbing pad at its origin in an experimental setup. Material and methods: All measurements were carried out using a clinical angiography system with a standardized fluoroscopy protocol, different C-arm angulations and an anthropomorphic torso phantom as a scattering body. An ionization chamber was used to measure the radiation exposure at five different heights of a simulated operator during a simulated transfemoral angiography intervention. Measurements were carried out with and without radiation-absorbing pads with lead equivalents of 0.25 and 0.5 mm placed onto the scattering body. For all measurements a mobile acrylic shield and an under-table lead curtain was used. Results: At all operator heights from 100 to 165 cm a significant radiation dose reduction of up to 80.6 % < 0.01) using the radiation-absorbing pad was measured, when compared to no radiation-absorbing pad. At the height of 165 cm the radiation-absorbing pad with a lead equivalence of 0.5 mm showed a significant radiation dose reduction (51.4 %, p < 0.01) in comparison to a lead equivalence of 0.25 mm. Conclusion: The addition of a radiation-absorbing pad to the standard protection means results in a significant dose reduction for the operator, particularly for upper body parts.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] Radiation exposure to medical staff in interventional radiology
    Williams, JR
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (852): : 1333 - 1334
  • [42] Efficacy of Radiation Safety Glasses in Interventional Radiology
    van Rooijen, Bart D.
    de Haan, Michiel W.
    Das, Marco
    Arnoldussen, Carsten W. K. P.
    de Graaf, R.
    van Zwam, Wim H.
    Backes, Walter H.
    Jeukens, Cecile R. L. P. N.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2014, 37 (05) : 1149 - 1155
  • [43] Training and accreditation in radiation protection for interventional radiology
    Vañó, E
    Gonzalez, L
    Faulkner, K
    Padovani, R
    Malone, JF
    RADIATION PROTECTION DOSIMETRY, 2001, 94 (1-2) : 137 - 142
  • [44] Efficacy of Radiation Safety Glasses in Interventional Radiology
    Bart D. van Rooijen
    Michiel W. de Haan
    Marco Das
    Carsten W. K. P. Arnoldussen
    R. de Graaf
    Wim H. van Zwam
    Walter H. Backes
    Cécile R. L. P. N. Jeukens
    CardioVascular and Interventional Radiology, 2014, 37 : 1149 - 1155
  • [45] Radiation protection for technologists and radiologists in interventional radiology
    Marande, J. L.
    JOURNAL DE RADIOLOGIE, 2010, 91 (11): : 1236 - 1236
  • [47] Development and Performance of New Gadolinium and Boron Containing Radiation-Absorbing Composite Systems
    N. M. Chikhradze
    F. D. S. Marquis
    G. S. Abashidze
    L. Kurdadze
    JOM, 2013, 65 : 728 - 738
  • [48] Occupational Radiation Protection in Interventional Radiology: A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology
    Miller, Donald L.
    Vano, Eliseo
    Bartal, Gabriel
    Balter, Stephen
    Dixon, Robert
    Padovani, Renato
    Schueler, Beth
    Cardella, John F.
    de Baere, Thierry
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 33 (02) : 230 - 239
  • [49] Occupational Radiation Protection in Interventional Radiology: A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology
    Miller, Donald L.
    Vano, Eliseo
    Bartal, Gabriel
    Balter, Stephen
    Dixon, Robert
    Padovani, Renato
    Schueler, Beth
    Cardella, John F.
    de Baere, Thierry
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (05) : 607 - 615
  • [50] Occupational Radiation Protection in Interventional Radiology: A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology
    Donald L. Miller
    Eliseo Vañó
    Gabriel Bartal
    Stephen Balter
    Robert Dixon
    Renato Padovani
    Beth Schueler
    John F. Cardella
    Thierry de Baère
    CardioVascular and Interventional Radiology, 2010, 33 : 230 - 239