Scatter radiation reduction with a radiation-absorbing pad in interventional radiology examinations

被引:21
|
作者
Koenig, A. M. [1 ]
Maas, J. [1 ]
Viniol, S. [1 ]
Etzel, R. [2 ]
Fiebich, M. [1 ,2 ]
Thomas, R. P. [1 ]
Mahnken, A. H. [1 ]
机构
[1] Philipps Univ Marburg, Univ Hosp Marburg, Clin Diagnost & Intervent Radiol, Marburg, Germany
[2] Mittelhessen Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
关键词
Radiation protection; Interventional radiology; Radiation-absorbing pad; MEDICAL STAFF; CANCER-RISKS; PROTECTION; EXPOSURE; OPERATOR; DOSIMETRY; EFFICACY; SHIELD;
D O I
10.1016/j.ejrad.2020.109245
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Radiation-absorbing pads are an additional possibility to reduce scattered radiation at its source. The goal of this study is to investigate the efficacy of a new reusable radiation-absorbing pad at its origin in an experimental setup. Material and methods: All measurements were carried out using a clinical angiography system with a standardized fluoroscopy protocol, different C-arm angulations and an anthropomorphic torso phantom as a scattering body. An ionization chamber was used to measure the radiation exposure at five different heights of a simulated operator during a simulated transfemoral angiography intervention. Measurements were carried out with and without radiation-absorbing pads with lead equivalents of 0.25 and 0.5 mm placed onto the scattering body. For all measurements a mobile acrylic shield and an under-table lead curtain was used. Results: At all operator heights from 100 to 165 cm a significant radiation dose reduction of up to 80.6 % < 0.01) using the radiation-absorbing pad was measured, when compared to no radiation-absorbing pad. At the height of 165 cm the radiation-absorbing pad with a lead equivalence of 0.5 mm showed a significant radiation dose reduction (51.4 %, p < 0.01) in comparison to a lead equivalence of 0.25 mm. Conclusion: The addition of a radiation-absorbing pad to the standard protection means results in a significant dose reduction for the operator, particularly for upper body parts.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] RADIATION-ABSORBING GLASSES
    BREWSTER, GF
    KREIDL, NJ
    JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1952, 35 (10) : 259 - 264
  • [2] Broadband radiation-absorbing metamaterial
    Narimanov, Evgenii E.
    Li, H.
    Barnakov, Yu
    Timkur, T. U.
    Noginov, M. A.
    2011 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2011,
  • [3] Radiation shielding materials and radiation scatter effects for interventional radiology (IR) physicians
    McCaffrey, J. P.
    Tessier, F.
    Shen, H.
    MEDICAL PHYSICS, 2012, 39 (07) : 4537 - 4546
  • [4] Radiation-absorbing materials on basis of metallopolymers
    Sokolov, E.A.
    Babenko, S.D.
    Kritskaya, D.A.
    Ponomarev, A.N.
    Plasticheskie Massy, (05): : 20 - 21
  • [5] Darker than black: radiation-absorbing metamaterial
    Narimanov, E. E.
    Li, H.
    Barnakov, Yu. A.
    Noginov, M. A.
    2010 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO) AND QUANTUM ELECTRONICS AND LASER SCIENCE CONFERENCE (QELS), 2010,
  • [6] Use of a sterile, disposable, radiation-absorbing shield reduces occupational. exposure to scatter radiation during pectoral device implantation
    Simons, GR
    Orrison, WW
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 726 - 729
  • [7] A RADIATION-ABSORBING CENTRE IN A NONSTATICAL HOMOGENEOUS UNIVERSE
    VAIDYA, PC
    NATURE, 1950, 166 (4222) : 565 - 565
  • [8] TEMPERATURE-FIELDS IN LARGE RADIATION-ABSORBING BODIES
    STELZER, JF
    WELZEL, R
    SIEVERS, A
    KERNTECHNIK, 1977, 19 (07) : 311 - 318
  • [9] Radiation risks in interventional radiology
    Dendy, P. P.
    BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (961): : 1 - 7
  • [10] Radiation protection in interventional radiology
    Johnson, DR
    Kyriou, J
    Morton, EJ
    Clifton, A
    Fitzgerald, M
    MacSweeney, E
    CLINICAL RADIOLOGY, 2001, 56 (02) : 99 - 106