Reproducibility and reliability of volumetric measurements of olfactory eloquent structures
被引:46
|作者:
Yousem, DM
Geckle, RJ
Doty, RL
Bilker, WB
机构:
[1] UNIV PENN,MED CTR,CTR SMELL & TASTE,PHILADELPHIA,PA 19104
[2] UNIV PENN,MED CTR,DEPT BIOSTAT & EPIDEMIOL,PHILADELPHIA,PA 19104
[3] UNIV PENN,MED CTR,CTR CLIN EPIDEMIOL & BIOSTAT,PHILADELPHIA,PA 19104
关键词:
brain;
volume;
MR;
magnetic resonance (MR);
volume measurement;
phantoms;
D O I:
10.1016/S1076-6332(97)80027-X
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Rationale and Objectives. The authors assessed the reliability and reproducibility of volumetric measurements of olfactory bulbs and tracts (OBTs) and temporal lobes (TLs) with magnetic resonance (MR) imaging. Methods. Repetitive MR imaging of two phantoms and OBTs and TLs of patients was performed. Regions of interest were manually drawn around the relevant structures, and their volumes were measured at workstations. Intra- and interobserver variability for the two readers were measured. Intraclass and Pearson correlation coefficients and mean percentage differences were calculated. Results. The measured phantom volumes were within 1.9%-12% of the true volumes with a variability of 15%. Intraclass and Pearson correlation coefficients were greater than or equal to 0.919 for measurements by a single reader and greater than or equal to 0.924 for measurements by different readers. The inter- and intraobserver variabilities were less than or equal to 4.2% for TL and 11.3%-14.6% for OBTs. Conclusion. Volumetric measurements of the olfactory apparatus can be reliably and accurately reproduced from MR images.
引用
收藏
页码:264 / 269
页数:6
相关论文