A Vibration Free Closed-Cycle 1 K Cryostat with a 4 K Pulse Tube Cryocooler

被引:3
|
作者
Wang, Chao [1 ]
Lichtenwalter, Ben [1 ]
机构
[1] Cryomech, Syracuse, NY 13211 USA
来源
关键词
1 K cryostat; vibration free; pulse tube; cryocooler;
D O I
10.1063/1.4860868
中图分类号
O414.1 [热力学];
学科分类号
摘要
A 1 K closed-cycle cryostat, pre-cooled by a 4 K pulse tube cryocooler, has been developed. The Cryomech PT410 pulse tube cryocooler liquefies helium in a vacuum insulated sleeve at a pressure of similar to 1 atm. Liquid helium flows through a JT valve and into a 1 K pot that is evacuated by a vacuum pump. The discharged gas from the vacuum is routed to the top of the sleeve to be liquefied. This design accomplishes closed-cycle 1 K refrigeration and provides continuous cooling below 2 K. Using two XDS10 vacuum pumps and with the JT valve optimized for maximum cooling capacity, the 1 K cooling station can reach a no-load temperature of 1.51 K and provide a capacity of 225 mW at 1.76 K. The temperature oscillations on the 4 K and 1 K cooling stations are +/- 3 mK. The cryostat is designed so that there is no direct mechanical contact between the pulse tube cryocooler heat exchangers and the 1 K cooling station. This design feature enables exceptionally low vibration operation at the 1 K cooling station.
引用
收藏
页码:1387 / 1392
页数:6
相关论文
共 50 条
  • [21] Efficient helium recondensing using a 4 K pulse tube cryocooler
    Wang, C
    CRYOGENICS, 2005, 45 (12) : 719 - 724
  • [22] Design of 4 K Stirling type pulse tube cryocooler with precooling
    Gan, Zhi-Hua
    Li, Zhuo-Pei
    Chen, Jie
    Qiu, Li-Min
    Dai, Li
    Kung Cheng Je Wu Li Hsueh Pao/Journal of Engineering Thermophysics, 2009, 30 (03): : 378 - 380
  • [23] Operation of SFQ circuits cooled with 4 K closed-cycle refrigerator
    Suzuki, H
    Hirano, HS
    Tanabe, K
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2004, 412 : 1580 - 1585
  • [24] A GENERAL-PURPOSE COLD FINGER USING A VIBRATION-FREE MOUNTED HE CLOSED-CYCLE CRYOSTAT
    BOOLCHAND, P
    LEMON, GH
    BRESSER, WJ
    ENZWEILER, RN
    HARRIS, R
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (04): : 3051 - 3057
  • [25] Modernized Liquid Helium-Free Closed-Cycle Cryostat for Mossbauer Research
    Starchikov, S. S.
    Funtov, K. O.
    Zayakhanov, V. A.
    Frolov, K. V.
    Klenov, M. G.
    Bondarenko, I. Yu.
    Lyubutin, I. S.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2023, 66 (03) : 497 - 507
  • [26] Thermodynamic analysis and optimization of 80K closed-cycle Joule-Thompson cryocooler with gas mixture
    Luo, EC
    Zhou, Y
    Gong, MQ
    ADVANCES IN CRYOGENIC ENGINEERING, VOL 43 PTS A AND B, 1998, 43 : 1679 - 1683
  • [27] Experimental and simulation investigation on high power 4 K pulse tube cryocooler
    Dong, Wen-Qing
    Qiu, Li-Min
    Wang, Chao
    Gan, Zhi-Hua
    Kung Cheng Je Wu Li Hsueh Pao/Journal of Engineering Thermophysics, 2011, 32 (10): : 1627 - 1630
  • [28] Conceptual design of space qualified 4K pulse tube cryocooler
    Chandratilleke, GR
    Ohtani, Y
    Nakagome, H
    Mimura, K
    Yoshimura, N
    Matsubara, Y
    Okuda, H
    Iida, T
    Shinohara, S
    CRYOCOOLERS 10, 1999, : 221 - 226
  • [29] Performance dependence of 4K pulse tube cryocooler on working pressure
    Yoshimura, N
    Zhou, SL
    Matsubara, K
    Chandratilleke, GR
    Ohtani, Y
    Nakagome, H
    Okuda, H
    Shinohara, S
    CRYOCOOLERS 10, 1999, : 227 - 232
  • [30] Progress on a novel VM-type pulse tube cryocooler for 4 K
    Pan, Changzhao
    Wang, Jue
    Luo, Kaiqi
    Wang, Junjie
    Zhou, Yuan
    CRYOGENICS, 2017, 88 : 66 - 69