A Vibration Free Closed-Cycle 1 K Cryostat with a 4 K Pulse Tube Cryocooler

被引:3
|
作者
Wang, Chao [1 ]
Lichtenwalter, Ben [1 ]
机构
[1] Cryomech, Syracuse, NY 13211 USA
来源
关键词
1 K cryostat; vibration free; pulse tube; cryocooler;
D O I
10.1063/1.4860868
中图分类号
O414.1 [热力学];
学科分类号
摘要
A 1 K closed-cycle cryostat, pre-cooled by a 4 K pulse tube cryocooler, has been developed. The Cryomech PT410 pulse tube cryocooler liquefies helium in a vacuum insulated sleeve at a pressure of similar to 1 atm. Liquid helium flows through a JT valve and into a 1 K pot that is evacuated by a vacuum pump. The discharged gas from the vacuum is routed to the top of the sleeve to be liquefied. This design accomplishes closed-cycle 1 K refrigeration and provides continuous cooling below 2 K. Using two XDS10 vacuum pumps and with the JT valve optimized for maximum cooling capacity, the 1 K cooling station can reach a no-load temperature of 1.51 K and provide a capacity of 225 mW at 1.76 K. The temperature oscillations on the 4 K and 1 K cooling stations are +/- 3 mK. The cryostat is designed so that there is no direct mechanical contact between the pulse tube cryocooler heat exchangers and the 1 K cooling station. This design feature enables exceptionally low vibration operation at the 1 K cooling station.
引用
收藏
页码:1387 / 1392
页数:6
相关论文
共 50 条
  • [1] A closed-cycle 1 K refrigeration cryostat
    Wang, Chao
    Lichtenwalter, Ben
    Friebel, Aaron
    Tang, Hong X.
    Cryogenics, 2014, 64 : 5 - 9
  • [2] A closed-cycle 1 K refrigeration cryostat
    Wang, Chao
    Ben Lichtenwalter
    Friebel, Aaron
    Tang, Hong X.
    CRYOGENICS, 2014, 64 : 5 - 9
  • [3] A closed-cycle 1 K refrigeration cryostat
    Wang, Chao
    Lichtenwalter, Ben
    Friebel, Aaron
    Tang, Hong X.
    Cryogenics, 2014, 64 : 5 - 9
  • [4] High capacity closed-cycle 1 K cryocooler
    Wang, C.
    Hanrahan, T.
    Johnson, M.
    CRYOGENICS, 2018, 95 : 64 - 68
  • [5] A vibration free cryostat using pulse tube cryocooler
    Wang, Chao
    Hartnett, John G.
    CRYOGENICS, 2010, 50 (05) : 336 - 341
  • [6] Closed-cycle, low-vibration 4 K cryostat for ion traps and other applications
    Micke, P.
    Stark, J.
    King, S. A.
    Leopold, T.
    Pfeifer, T.
    Schmoeger, L.
    Schwarz, M.
    Spiess, L. J.
    Schmidt, P. O.
    Lopez-Urrutia, J. R. Crespo
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (06):
  • [7] Ultrastable Silicon Cavity in a Continuously Operating Closed-Cycle Cryostat at 4 K
    Zhang, W.
    Robinson, J. M.
    Sonderhouse, L.
    Oelker, E.
    Benko, C.
    Hall, J. L.
    Legero, T.
    Matei, D. G.
    Riehle, F.
    Sterr, U.
    Ye, J.
    PHYSICAL REVIEW LETTERS, 2017, 119 (24)
  • [8] 1 K cryostat with sub-millikelvin stability based on a pulse-tube cryocooler
    DeMann, A.
    Mueller, Sara
    Field, S. B.
    CRYOGENICS, 2016, 73 : 60 - 67
  • [9] Vibration-free cryostat for low-noise applications of a pulse tube cryocooler
    Caparrelli, S.
    Majorana, E.
    Moscatelli, V.
    Pascucci, E.
    Perciballi, M.
    Puppo, P.
    Rapagnani, P.
    Ricci, F.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (09):
  • [10] Experimental study of a 4 K pulse tube cryocooler
    Fujimoto, S
    Kurihara, T
    Oodo, T
    Kang, YM
    Numazawa, T
    Matsubara, Y
    CRYOCOOLERS 11, 2001, : 213 - 219