Constrained sessile drop as a new configuration to measure low surface tension in lung surfactant systems

被引:45
|
作者
Yu, LMY
Lu, JJ
Chan, YW
Ng, A
Zhang, L
Hoorfar, M
Policova, Z
Grundke, K
Neumann, AW
机构
[1] Univ Toronto, Dept Mech & Ind Engn, Toronto, ON M5S 3G8, Canada
[2] Univ Toronto, Dept Chem Engn & Appl Chem, Toronto, ON M5S 3E5, Canada
[3] Inst Polymerforsch Dresden, D-01069 Dresden, Germany
关键词
constrained sessile drop; sharp knife edge; ultra-low surface tension; film leakage; spread dipalmitoyl phosphatidylcholine film;
D O I
10.1152/japplphysiol.00089.2003
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Existing methodology for surface tension measurements based on drop shapes suffers from the shortcoming that it is not capable to function at very low surface tension if the liquid dispersion is opaque, such as therapeutic lung surfactants at clinically relevant concentrations. The novel configuration proposed here removes the two big restrictions, i.e., the film leakage problem that is encountered with such methods as the pulsating bubble surfactometer as well as the pendant drop arrangement, and the problem of the opaqueness of the liquid, as in the original captive bubble arrangement. A sharp knife edge is the key design feature in the constrained sessile drop that avoids film leakage at low surface tension. The use of the constrained sessile drop configuration in conjunction with axisymmetric drop shape analysis to measure surface tension allows complete automation of the setup. Dynamic studies with lung surfactant can be performed readily by changing the volume of a sessile drop, and thus the surface area, by means of a motor-driven syringe. To illustrate the validity of using this configuration, experiments were performed using an exogenous lung surfactant preparation, bovine lipid extract surfactant (BLES) at 5.0 mg/ml. A comparison of results obtained for BLES at low concentration between the constrained sessile drop and captive bubble arrangement shows excellent agreement between the two approaches. When the surface area of the BLES film (0.5 mg/ml) was compressed by about the same amount in both systems, the minimum surface tensions attained were identical within the 95% confidence limits.
引用
收藏
页码:704 / 715
页数:12
相关论文
共 50 条
  • [31] COMPUTATION OF SURFACE TENSION AND OF CONTACT ANGLE BY SESSILE-DROP METHOD (II)
    STAICOPOLUS, DN
    JOURNAL OF COLLOID SCIENCE, 1963, 18 (08): : 793 - &
  • [32] Measurement of surface tension of molten matte phases by an improved sessile drop method
    Hamuyuni, J.
    Taskinen, P.
    Akdogan, G.
    Bradshaw, S. M.
    TRANSACTIONS OF THE INSTITUTIONS OF MINING AND METALLURGY SECTION C-MINERAL PROCESSING AND EXTRACTIVE METALLURGY, 2012, 121 (03): : 173 - 177
  • [33] MEASUREMENT OF THE SURFACE TENSION OF TITANIFEROUS MOLTEN IRON BY THE SESSILE DROP METHOD.
    Luo, Weizhong
    Mao, Yuwen
    Huang, Lu
    Zhu, Yuankai
    Kang T'ieh/Iron and Steel (Peking), 1987, 22 (01): : 1 - 4
  • [34] THE DEPENDENCE OF THE SURFACE-TENSION OF SURFACTANT SOLUTIONS ON DROP SIZE
    BIANCO, H
    MARMUR, A
    JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1992, 151 (02) : 517 - 522
  • [35] Surface tension and viscosity of surfactant from the resonance of an oscillating drop
    W. Meier
    G. Greune
    A. Meyboom
    K.P. Hofmann
    European Biophysics Journal, 2000, 29 : 113 - 124
  • [36] Surface tension and viscosity of surfactant from the resonance of an oscillating drop
    Meier, W
    Greune, G
    Meyboom, A
    Hofmann, KP
    EUROPEAN BIOPHYSICS JOURNAL WITH BIOPHYSICS LETTERS, 2000, 29 (02): : 113 - 124
  • [37] Surface tension of molten silicon measured by microgravity oscillating drop method and improved sessile drop method
    Fujii, H
    Matsumoto, T
    Izutani, S
    Kiguchi, S
    Nogi, K
    ACTA MATERIALIA, 2006, 54 (05) : 1221 - 1225
  • [38] Measurement of surface tension by sessile drop tensiometer with superoleophobic surface (vol 112, 121602, 2018)
    Kwak, Wonshik
    Park, Jun Kwon
    Yoon, Jinsung
    Lee, Sanghyun
    Hwang, Woonbong
    APPLIED PHYSICS LETTERS, 2018, 112 (16)
  • [39] MEASUREMENT OF THE SURFACE-TENSION OF GALLIUM AND INDIUM IN A HYDROGEN ATMOSPHERE BY THE SESSILE DROP METHOD
    KONIG, U
    KECK, W
    JOURNAL OF THE LESS-COMMON METALS, 1983, 90 (02): : 299 - 303
  • [40] Surface tension of molten bismuth at different oxygen partial pressure with the sessile drop method
    Yuan, ZF
    Fan, JF
    Li, J
    Ke, JJ
    Mukai, K
    SCANDINAVIAN JOURNAL OF METALLURGY, 2004, 33 (06) : 338 - 346