Iliosacral screw insertion using CT-3D-fluoroscopy matching navigation

被引:60
|
作者
Takao, Masaki [1 ]
Nishii, Takashi [2 ]
Sakai, Takashi [1 ]
Yoshikawa, Hideki [1 ]
Sugano, Nobuhiko [2 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Orthopaed Surg, Suita, Osaka 5650871, Japan
[2] Osaka Univ, Grad Sch Med, Dept Orthopaed Med Engn, Suita, Osaka 5650871, Japan
基金
日本学术振兴会;
关键词
Iliosacral screw; Navigation; 3D fluoroscopy; CT-based plan; SACROILIAC SCREWS; PLACEMENT; FLUOROSCOPY; FIXATION; SYSTEM; FRACTURES; ACCURACY; REGIONS;
D O I
10.1016/j.injury.2014.01.015
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Percutaneous iliosacral screw insertion requires substantial experience and detailed anatomical knowledge to find the proper entry point and trajectory even with the use of a navigation system. Our hypothesis was that three-dimensional (3D) fluoroscopic navigation combined with a preoperative computed tomography (CT)-based plan could enable surgeons to perform safe and reliable iliosacral screw insertion. The purpose of the current study is two-fold: (1) to demonstrate the navigation accuracy for sacral fractures and sacroiliac dislocations on widely displaced cadaveric pelves; and (2) to report the technical and clinical aspects of percutaneous iliosacral screw insertion using the CT-3D-fluoroscopy matching navigation system. Methods: We simulated three types of posterior pelvic ring disruptions with vertical displacements of 0, 1, 2 and 3 cm using cadaveric pelvic rings. A total of six fiducial markers were fixed to the anterior surface of the sacrum. Target registration error over the sacrum was assessed with the fluoroscopic imaging centre on the second sacral vertebral body. Six patients with pelvic ring fractures underwent percutaneous iliosacral screw placement using the CT-3D-fluoroscopy matching navigation. Three pelvic ring fractures were classified as type B2 and three were classified as type C1 according to the AO-OTA classification. Iliosacral screws for the S1 and S2 vertebra were inserted. Results: The mean target registration error over the sacrum was 1.2 mm (0.5-1.9 mm) in the experimental study. Fracture type and amount of vertical displacement did not affect the target registration error. All 12 screws were positioned correctly in the clinical series. There were no postoperative complications including nerve palsy. The mean deviation between the planned and the inserted screw position was 2.5 mm at the screw entry point, 1.8 mm at the area around the nerve root tunnels and 2.2 mm at the tip of the screw. Conclusion: The CT-3D-fluoroscopy matching navigation system was accurate and robust regardless of pelvic ring fracture type and fragment displacement. Percutaneous iliosacral screw insertion with the navigation system is clinically feasible. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:988 / 994
页数:7
相关论文
共 50 条
  • [1] CT-3D-Fluoroscopy Matching Navigation Can Reduce the Malposition Rate of Iliosacral Screw Insertion for Less-Experienced Surgeons
    Takao, Masaki
    Nishii, Takashi
    Sakai, Takashi
    Sugano, Nobuhiko
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (12) : 716 - 721
  • [2] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Takao, Masaki
    Hamada, Hidetoshi
    Sakai, Takashi
    Sugano, Nobuhiko
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (02) : 189 - 195
  • [3] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Masaki Takao
    Hidetoshi Hamada
    Takashi Sakai
    Nobuhiko Sugano
    Archives of Orthopaedic and Trauma Surgery, 2019, 139 : 189 - 195
  • [4] Navigation versus fluoroscopy in minimalinvasive iliosacral screw placement
    Prost, Max
    Taday, Roman
    Beyersdorf, Carl Christoph Paul
    Latz, David
    Windolf, Joachim
    Scheyerer, Max Joseph
    Konieczny, Markus Rafael
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01)
  • [5] Navigation versus fluoroscopy in minimalinvasive iliosacral screw placement
    Max Prost
    Roman Taday
    Carl Christoph Paul Beyersdorf
    David Latz
    Joachim Windolf
    Max Joseph Scheyerer
    Markus Rafael Konieczny
    Journal of Orthopaedic Surgery and Research, 19
  • [6] Comparison of fluoroscopy and fluoroscopy-based 2D computer navigation for iliosacral screw placement: a retrospective study
    Madeja, Roman
    Pometlova, Jana
    Osemlak, Pawel
    Voves, Jiri
    Bialy, Lubor
    Vrtkova, Adela
    Pleva, Leopold
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2022, 48 (06) : 4897 - 4902
  • [7] Comparison of fluoroscopy and fluoroscopy-based 2D computer navigation for iliosacral screw placement: a retrospective study
    Roman Madeja
    Jana Pometlová
    Pawel Osemlak
    Jiří Voves
    Lubor Bialy
    Adéla Vrtková
    Leopold Pleva
    European Journal of Trauma and Emergency Surgery, 2022, 48 : 4897 - 4902
  • [8] Evaluation of 2D and 3D navigation for iliosacral screw fixation
    Daniel Behrendt
    Maria Mütze
    Hanno Steinke
    Martin Koestler
    Christoph Josten
    Jörg Böhme
    International Journal of Computer Assisted Radiology and Surgery, 2012, 7 : 249 - 255
  • [9] Evaluation of 2D and 3D navigation for iliosacral screw fixation
    Behrendt, Daniel
    Muetze, Maria
    Steinke, Hanno
    Koestler, Martin
    Josten, Christoph
    Boehme, Jorg
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (02) : 249 - 255
  • [10] Application of a CT-3D fluoroscopy matching navigation system to the pelvic and femoral regions
    Takao, Masaki
    Nishii, Takashi
    Sakai, Takashi
    Sugano, Nobuhiko
    COMPUTER AIDED SURGERY, 2012, 17 (02) : 69 - 76