Accuracy evaluation of a 3D-printed individual template for needle guidance in head and neck brachytherapy

被引:50
|
作者
Huang, Ming-Wei [1 ]
Zhang, Jian-Guo [1 ]
Zheng, Lei [1 ]
Liu, Shu-Ming [1 ]
Yu, Guang-Yan [1 ]
机构
[1] Peking Univ, Dept Oral & Maxillofacial Surg, Sch & Hosp Stomatol, Beijing 100081, Peoples R China
关键词
3D-printed; individual template; brachytherapy; head and neck; PERMANENT PROSTATE IMPLANTS; DOSE-RATE BRACHYTHERAPY; INTERSTITIAL BRACHYTHERAPY; NAVIGATION SYSTEM; MALIGNANT-TUMORS; IMAGE-GUIDANCE; VAGINAL TUMORS; PLACEMENT; INSERTION; CANCER;
D O I
10.1093/jrr/rrw033
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
To transfer the preplan for the head and neck brachytherapy to the clinical implantation procedure, a preplanbased 3D-printed individual template for needle insertion guidance had previously been designed and used. The accuracy of needle insertion using this kind template was assessed in vivo. In the study, 25 patients with head and neck tumors were implanted with I-125 radioactive seeds under the guidance of the 3D-printed individual template. Patients were divided into four groups based on the site of needle insertion: the parotid and masseter region group (nine patients); the maxillary and paranasal region group (eight patients); the submandibular and upper neck area group (five patients); and the retromandibular region group (six patients). The distance and angular deviations between the preplanned and placed needles were compared, and the complications and time required for needle insertion were assessed. The mean entrance point distance deviation for all 619 needles was 1.18 +/- 0.81 mm, varying from 0.857 +/- 0.545 to 1.930 +/- 0.843 mm at different sites. The mean angular deviation was 2.08 +/- 1.07 degrees, varying from 1.85 +/- 0.93 to 2.73 +/- 1.18 degrees at different sites. All needles were manually inserted to their preplanned positions in a single attempt, and the mean time to insert one needle was 7.5 s. No anatomical complications related to inaccurately placed implants were observed. Using the 3D-printed individual template for the implantation of I-125 radioactive seeds in the head and neck region can accurately transfer a CT-based preplan to the brachytherapy needle insertion procedure. Moreover, the addition of individual template guidance can reduce the time required for implantation and minimize the damage to normal tissues.
引用
收藏
页码:662 / 667
页数:6
相关论文
共 50 条
  • [21] Design of Custom, 3D-Printed Surface Brachytherapy Applicators
    Chytyk-Praznik, K.
    Oliver, P.
    Allan, J.
    Best, L.
    Robar, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E737 - E737
  • [22] 3D-printed Salt Template for bioabsorbable Bone Implants
    不详
    ORTHOPADE, 2019, 48 (10): : 848 - 848
  • [23] Design of 3D-printed universal oral stent for tongue immobilization in head and neck radiotherapy
    Mundee, Thunpisit
    Jongwannasiri, Chavin
    Fuangrod, Todsaporn
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (01)
  • [24] Improved accuracy of 3D-printed navigational template during complicated tibial plateau fracture surgery
    Huang, Huajun
    Hsieh, Ming-Fa
    Zhang, Guodong
    Ouyang, Hanbin
    Zeng, Canjun
    Yan, Bin
    Xu, Jing
    Yang, Yang
    Wu, Zhanglin
    Huang, Wenhua
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (01) : 109 - 117
  • [25] Improved accuracy of 3D-printed navigational template during complicated tibial plateau fracture surgery
    Huajun Huang
    Ming-Fa Hsieh
    Guodong Zhang
    Hanbin Ouyang
    Canjun Zeng
    Bin Yan
    Jing Xu
    Yang Yang
    Zhanglin Wu
    Wenhua Huang
    Australasian Physical & Engineering Sciences in Medicine, 2015, 38 : 109 - 117
  • [26] 3D-Printed Linear Positioner with Micrometer Accuracy
    Kuo, Yin-Yen
    Cheng, Chiao-Hua
    Hung, Shao-Kang
    2016 THE 3RD INTERNATIONAL CONFERENCE ON MECHATRONICS AND MECHANICAL ENGINEERING (ICMME 2016), 2017, 95
  • [27] Accuracy assessment of 3D-printed tooth replicas
    Sokolowski, Armin Andreas
    Sokolowski, Alwin Alan
    Kammerhofer, Julian
    Madreiter-Sokolowski, Corina Theresia
    Payer, Michael
    Koller, Martin
    Jakse, Norbert
    Wegscheider, Walther Alfred
    INTERNATIONAL JOURNAL OF COMPUTERIZED DENTISTRY, 2019, 22 (04) : 321 - 329
  • [28] The Accuracy of 3D-Printed Fixed Dental Restorations
    Alammar, Amirah
    Att, Wael
    Beuer, Florian
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2024,
  • [29] Tiny 3D-printed lens threads syringe needle
    不详
    PHYSICS WORLD, 2016, 29 (08) : 5 - 5
  • [30] Evaluation of the accuracy of digital and 3D-printed casts compared with conventional stone casts
    Ellakany, Passent
    Al-Harbi, Fahad
    El Tantawi, Maha
    Mohsen, Cherif
    JOURNAL OF PROSTHETIC DENTISTRY, 2022, 127 (03): : 438 - 444