Moisture content determination of pharmaceutical pellets by near infrared spectroscopy: Method development and validation

被引:76
|
作者
Mantanus, J. [1 ]
Ziemons, E. [1 ]
Lebrun, P. [1 ]
Rozet, E. [1 ]
Klinkenberg, R. [2 ]
Streel, B. [2 ]
Evrard, B. [3 ]
Hubert, Ph. [1 ]
机构
[1] Univ Liege, CIRM, Analyt Chem Lab, B-4000 Liege, Belgium
[2] Galephar Res Ctr MF, B-6900 Marche En Famenne, Belgium
[3] Univ Liege, CIRM, Pharmaceut Technol Lab, B-4000 Liege, Belgium
关键词
Moisture; Near infrared spectroscopy; Pharmaceutical pellets; Validation; Accuracy profile; PAT; QUANTITATIVE ANALYTICAL PROCEDURES; PROCESS ANALYTICAL TECHNOLOGY; SFSTP PROPOSAL; HARMONIZATION; GRANULATION; STRATEGIES; ASSAY; SPECTROPHOTOMETRY; CHEMOMETRICS; TABLETS;
D O I
10.1016/j.aca.2008.12.031
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of the present study was to develop and validate a near infrared method able to accurately determine a moisture content of pharmaceutical pellets ranging from 1% to 8% in order to check their moisture content conformity. A calibration and validation set were designed for the conception and evaluation of the method adequacy. An experimental protocol was then followed, involving two operators, independent production campaign batches and different temperatures for data acquisition. On the basis of this protocol, prediction models based on partial least squares (PLS) regression were then carried out. Conventional criteria such as the R-2, the root mean square errors of calibration and prediction (RMSEC and RMSEP) as well as the number of PLS factors enabled the selection of three preliminary models. However, such criteria did not clearly demonstrate the model's ability to give accurate predictions over the whole analyzed water content range. Consequently, a novel approach based on accuracy profiles which allow the selection of the most fitted model for purpose was used. According to this novel approach, the model using multiplicative scatter correction (MSC) pre-treatment was obviously the most suitable. Indeed, the resulting accuracy profile clearly showed that this model was able to determine moisture content over the range of 1-8% with a very acceptable accuracy. The present study confirmed that NIR spectroscopy could be used in the PAT concept as a non-invasive, non-destructive and fast technique for moisture content determination in pharmaceutical pellets. In addition, facing the limit of the classical and commonly used criteria, the use of accuracy profiles proved to be useful as a powerful decision tool to demonstrate the suitability of the proposed analytical method. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:186 / 192
页数:7
相关论文
共 50 条
  • [31] Determination of flow properties of pharmaceutical powders by near infrared spectroscopy
    Sarraguca, Mafalda C.
    Cruz, Ana V.
    Soares, Sandra O.
    Amaral, Helena R.
    Costa, Paulo C.
    Lopes, Joao A.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2010, 52 (04) : 484 - 492
  • [32] Moisture determination in hygroscopic drug substances by near infrared spectroscopy
    Zhou, XJ
    Hines, P
    Borer, MW
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1998, 17 (02) : 219 - 225
  • [33] Determination of Ribavirin and Moisture in Pharmaceuticals by Near-Infrared Spectroscopy
    Yang, Hongqin
    Liu, Yanxin
    Huang, Yanmei
    Tang, Bin
    Guo, Dan
    Li, Hui
    ANALYTICAL LETTERS, 2016, 49 (13) : 2077 - 2091
  • [34] The legume content in multispecies mixtures as estimated with near infrared reflectance spectroscopy: Method validation
    Locher, F
    Heuwinkel, H
    Gutser, R
    Schmidhalter, U
    AGRONOMY JOURNAL, 2005, 97 (01) : 18 - 25
  • [35] Development of a Near Infrared Spectroscopy method for the in-line quantitative bilastine drug determination during pharmaceutical powders blending
    Biagi, Diletta
    Nencioni, Paolo
    Valleri, Maurizio
    Calamassi, Niccolo
    Mura, Paola
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2021, 204
  • [36] Determination of Protein Content in Pea by Near Infrared Spectroscopy
    Lee, Jin Hwan
    Choung, Myoung-Gun
    FOOD SCIENCE AND BIOTECHNOLOGY, 2009, 18 (01) : 60 - 65
  • [37] Active content determination of pharmaceutical tablets using near infrared spectroscopy as Process Analytical Technology tool
    Chavez, Pierre-Francois
    Sacre, Pierre-Yves
    De Bleye, Charlotte
    Netchacovitch, Lauranne
    Mantanus, Jerome
    Motte, Henri
    Schubert, Martin
    Hubert, Philippe
    Ziemons, Eric
    TALANTA, 2015, 144 : 1352 - 1359
  • [38] Calibration models for determination of water content in pharmaceutical excipients using near-infrared spectroscopy (NIRS)
    Beyer, J
    Steffens, KJ
    PHARMAZEUTISCHE INDUSTRIE, 2003, 65 (02): : 186 - 192
  • [39] Determination of moisture content and density of fresh-sawn red oak lumber by near infrared spectroscopy
    Defo, Maurice
    Taylor, Adam M.
    Bond, Brian
    FOREST PRODUCTS JOURNAL, 2007, 57 (05) : 68 - 72
  • [40] DETERMINATION OF THE MOISTURE-CONTENT OF STARCH USING NEAR-INFRARED PHOTO-ACOUSTIC SPECTROSCOPY
    BELTON, PS
    TANNER, SF
    ANALYST, 1983, 108 (1286) : 591 - 596