Five-field IMRT class solutions and dosimetric planning guidelines for implementing accelerated partial breast irradiation

被引:15
|
作者
Quirk, Sarah [1 ,3 ]
Grendarova, Petra [3 ,4 ]
Roumeliotis, Michael [1 ,2 ,3 ]
机构
[1] Tom Baker Canc Clin, Dept Med Phys, Calgary, AB, Canada
[2] Univ Calgary, Dept Phys & Astron, Calgary, AB, Canada
[3] Univ Calgary, Dept Oncol, Calgary, AB, Canada
[4] Tom Baker Canc Clin, Dept Oncol, Calgary, AB, Canada
关键词
INTENSITY-MODULATED RADIOTHERAPY; COUCH POSITION COMBINATIONS; COLLISION INDICATOR CHARTS; RADIATION-THERAPY; LUMPECTOMY CAVITY; TRIAL; APBI;
D O I
10.1016/j.prro.2017.09.009
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A comprehensive set of planning guidelines was developed to aid in achieving reproducible dosimetric results for external beam accelerated partial breast irradiation (APBI). The methodology for the development of class solutions for dosimetric planning of the APBI technique, including dose constraint recommendations, is presented for target coverage and conformity as well as normal tissues. Methods and materials: A conservative patient setup was simulated on a linear accelerator, and a comprehensive arrangement of gantry and couch angles was measured for clearance. This provided the foundation for available beam arrangements to develop reproducible and conformal 5-field intensity modulated radiation therapy partial breast plans. Forty patients were planned. Patient plans were assessed according to anatomy specific features, such as laterality and seroma location within the breast. Results: Clearance tables are presented to give permissible gantry and couch orientations according to measurements facilitated by patient simulation. Beam arrangement class solutions are presented for left- and right-sided APBI patients. Dosimetric recommendations are made based on the results of 40 patient plans. The median and range, describing target coverage and target conformity, are reported, as are normal tissue constraints ipsilateral lung ipsilateral breast, heart, liver, and contralateral breast. In all cases, the dose recommendations were at least as strict as multi-institutional APBI trials. In the case of ipsilateral lung and ipsilateral breast, the planning recommendations are more stringent. Conclusions: APBI using a 5-field intensity modulated radiation therapy technique was comprehensively developed and evaluated to provide recommendations yielding highly conformal and reproducible treatment plans. This provides a clear method to implement external beam APBI planning and delivery. (C) 2017 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:E99 / E107
页数:9
相关论文
共 50 条
  • [1] Dosimetric comparison of 3D-CRT to IMRT for percutaneous accelerated partial breast irradiation
    Duma, M. N.
    Hugo, C.
    Ertel, E.
    Winkler, C.
    Geinitz, H.
    Molls, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2010, 186 : 36 - 37
  • [2] Dosimetric Comparison of Robotic IMRT and Partial-Arc VMAT in Accelerated Partial Breast Irradiation (APBI) for Early Breast Cancer
    Kim, S. Y.
    Kim, J. H.
    Woo, J. Y.
    Baek, J. G.
    Park, K. S.
    Kim, M. J.
    Cho, J. H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E639 - E639
  • [3] Application of IMRT Class Solution for Improved Prone Accelerated Partial Breast Irradiation Dosimetry
    Quirk, S.
    Thind, K.
    Cao, J.
    Frederick, A.
    Grendarova, P.
    Craighead, P.
    Olivotto, I. A.
    Roumeliotis, M.
    MEDICAL PHYSICS, 2018, 45 (06) : E298 - E298
  • [4] Dosimetric Quality of External Beam Radiation Planning Techniques in Accelerated Partial Breast Irradiation
    Richards, Z.
    Weir, J.
    Henson, C.
    Ahmad, S.
    Herman, T.
    Herman, T. De La Fuente
    MEDICAL PHYSICS, 2019, 46 (06) : E623 - E623
  • [5] Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes
    Zaker Rana
    Nadim M. Nasr
    Huaying Ji
    Virginia Lorio
    Stephanie Akbari
    Molly Sebastian
    Mami Martin
    Robert L. Hong
    Radiation Oncology, 10
  • [6] Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes
    Rana, Zaker
    Nasr, Nadim M.
    Ji, Huaying
    Lorio, Virginia
    Akbari, Stephanie
    Sebastian, Molly
    Martin, Mami
    Hong, Robert L.
    RADIATION ONCOLOGY, 2015, 10
  • [7] Comparison of Static Fields IMRT and Helical IMRT for Accelerated Partial Breast Irradiation
    Simontacchi, G.
    Meattini, I.
    Marrazzo, L.
    Pallotta, S.
    Talamonti, C.
    Turkaj, A.
    Topulli, J.
    Monteleone, E.
    Cappelli, S.
    Masoni, T.
    Bucciolini, M.
    Livi, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S260 - S260
  • [8] A treatment planning study comparing whole breast radiotherapy against conformal, IMRT and tomotherapy for accelerated partial breast irradiation
    Oliver, M
    Chen, J
    Van Dyk, J
    Wong, E
    Perera, F
    MEDICAL PHYSICS, 2005, 32 (06) : 2031 - 2031
  • [9] Breath hold and volumetric IMRT for accelerated partial breast irradiation (APBI)
    Osman, S. O. S.
    Essers, M.
    Poortmans, P. H.
    EUROPEAN JOURNAL OF CANCER, 2013, 49 : S450 - S450
  • [10] Dosimetric evaluation of accelerated partial breast irradiation using VMAT technique
    Peng, Jean L.
    McDonald, Dan G.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3529 - S3531