Investigating high speed phenomena in laser plasma interactions using dilation x-ray imager

被引:76
|
作者
Nagel, S. R. [1 ]
Hilsabeck, T. J. [2 ]
Bell, P. M. [1 ]
Bradley, D. K. [1 ]
Ayers, M. J. [1 ]
Piston, K. [1 ]
Felker, B. [1 ]
Kilkenny, J. D. [2 ]
Chung, T. [2 ]
Sammuli, B. [2 ]
Hares, J. D. [3 ]
Dymoke-Bradshaw, A. K. L. [3 ]
机构
[1] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
[2] Gen Atom Co, San Diego, CA 92186 USA
[3] Kentech Instruments Ltd, Wallingford OX10, Oxon, England
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2014年 / 85卷 / 11期
关键词
D O I
10.1063/1.4890396
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The DIlation X-ray Imager (DIXI) is a new, high-speed x-ray framing camera at the National Ignition Facility (NIF) sensitive to x-rays in the range of approximate to 2-17 keV. DIXI uses the pulse-dilation technique to achieve a temporal resolution of less than 10 ps, a approximate to 10x improvement over conventional framing cameras currently employed on the NIF (approximate to 100 ps resolution), and otherwise only attainable with 1D streaked imaging. The pulse-dilation technique utilizes a voltage ramp to impart a velocity gradient on the signal-bearing electrons. The temporal response, spatial resolution, and x-ray sensitivity of DIXI are characterized with a short x-ray impulse generated using the COMET laser facility at Lawrence Livermore National Laboratory. At the NIF a pinhole array at 10 cm from target chamber center (tcc) projects images onto the photocathode situated outside the NIF chamber wall with a magnification of approximate to 64x. DIXI will provide important capabilities for warm-dense-matter physics, high-energy-density science, and inertial confinement fusion, adding important capabilities to temporally resolve hot-spot formation, x-ray emission, fuel motion, and mix levels in the hot-spot at neutron yields of up to 10(17). We present characterization data as well as first results on electron-transport phenomena in buried-layer foil experiments. (C) 2014 AIP Publishing LLC.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] X-ray wavelength optimization of the laser plasma X-ray lithography source
    Chaker, M.
    Boily, S.
    Lafontaine, B.
    Kieffer, J.C.
    Pepin, H.
    Toubhans, I.
    Fabbro, R.
    Microelectronic Engineering, 1990, 10 (02) : 91 - 105
  • [42] High power laser-plasma x-ray source for lithography
    Gaeta, CJ
    Rieger, H
    Turcu, ICE
    Forber, RA
    Cassidy, KL
    Campeau, SM
    Powers, MJ
    Maldonado, JR
    Morris, JH
    Foster, RM
    Smith, HI
    Lim, MH
    EMERGING LITHOGRAPHIC TECHNOLOGIES VI, PTS 1 AND 2, 2002, 4688 : 818 - 833
  • [43] X-ray microscopy and imaging of Caenorhabditis elegans nematode using a laser plasma pulsed x-ray source
    Poletti, G
    Orsini, F
    Ullschmied, J
    Skala, J
    Kralikova, B
    Pfeifer, M
    Kadlec, C
    Mocek, T
    Präg, A
    Cotelli, F
    Lamia, CL
    Batani, D
    Bernardinello, A
    Desai, T
    Zullini, A
    LASER-GENERATED AND OTHER LABORATORY X-RAY AND EUV SOURCES, OPTICS, AND APPLICATIONS, 2003, 5196 : 375 - 382
  • [44] Theoretical study on temporal and spatial performance of magnetic solenoid used in dilation x-ray imager
    Wang, Qiangqiang
    Cao, Zhurong
    Chen, Tao
    Deng, Keli
    Deng, Bo
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (07):
  • [45] Note: Diagnosing femtosecond laser-solid interactions with monochromatic Kα imager and x-ray pinhole camera
    Lin, X. X.
    Li, Y. T.
    Liu, F.
    Liu, B. C.
    Du, F.
    Wang, S. J.
    Chen, L. M.
    Zhang, L.
    Zheng, Y.
    Liu, X.
    Liu, L.
    Wang, Z. H.
    Ma, J. L.
    Wei, Z. Y.
    Zhang, J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (03):
  • [46] Observations of X-Ray Radiation Pressure Force Using High-Speed Diffracted X-Ray Tracking
    Sasaki, Yuji C.
    Suzuki, Yasuhito
    Sekiguchi, Hiroshi
    Ohta, Noboru
    Mastuo, Tatsuhito
    Yagi, Naoto
    BIOPHYSICAL JOURNAL, 2011, 100 (03) : 312 - 312
  • [47] X-RAY COMPTON SCATTER IMAGING USING A HIGH-SPEED FLYING SPOT X-RAY TUBE
    TOWE, BC
    JACOBS, AM
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (10) : 717 - 721
  • [48] Development of High-speed X-ray CT Inspection System Using X-ray Line Sensor
    Suzuki, Daisuke
    Noguchi, Kenji
    Murakoshi, Takayuki
    Teramoto, Atsushi
    2013 IEEE 3RD CPMT SYMPOSIUM JAPAN (ICSJ 2013), 2013,
  • [49] Material discriminated X-ray CT by using conventional microfocus X-ray tube and CdTe imager
    Onishi, Y.
    Nakashima, T.
    Koike, A.
    Morii, H.
    Neo, Y.
    Mimura, H.
    Aoki, T.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 1170 - 1174
  • [50] Laser-plasma x-ray sources using ceramic target
    Takemura, Yuichiro
    Yamaguchii, Naohiro
    Hara, Tamio
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2008, 47 (04) : 2275 - 2278