In this work, we present a setup for high-power single shot stroboscopic digital holography and demonstrate it in an application on rabbit eardrum vibration measurement. The setup is able to make full-field time-resolved measurements of vibrating surfaces with a precision in the nanometer range in a broad frequency range. The height displacement of the measured object is visualized over the entire surface as a function of time. Vibration magnitude and phase maps can be extracted from these data, the latter proving to be very useful to reveal phase delays across the surface. Such deviations from modal motion indicate energy losses due to internal damping, in contrast to purely elastic mechanics. This is of great interest in middle ear mechanics and finite element modelling. In our setup, short laser pulses are fired at selected instants within the surface vibration period and are recorded by a CCD camera. The timing of the pulses and the exposure of the camera are synchronized to the vibration phase by a microprocessor. The high-power frequency-doubled Nd:YAG laser produces pulses containing up to 5 mJ of energy, which is amply sufficient to record single-shot holograms. As the laser pulse length is 8 ns and the smallest time step of the trigger electronics is 1 mu s, vibration measurements of frequencies up to 250 kHz are achievable through this method, provided that the maximum vibration amplitude exceeds a few nanometers. In our application, middle ear mechanics, measuring frequencies extend from 5 Hz to 20 kHz. The experimental setup will be presented, as well as results of measurements on a stretched circular rubber membrane and a rabbit's eardrum. Two of the challenges when measuring biological tissues, such as the eardrum, are low reflectivity and fast dehydration. To increase reflectivity, a coating is applied and to counteract the undesirable effects of tissue dehydration, the measurement setup and software have been optimized for speed without compromising on the quality. Results of a repeatability test will be presented as well. Since the method measures the membrane motion as a function of time in small time steps, we do not only measure vibration amplitude like in time-average holography, but we can also measure non-linear elastic and transient behaviour. In conclusion, the combination of good spatial, depth and time resolution with the fast data acquisition and very wide frequency range make our technique applicable in a number of fields, including biological tissue vibrations.
引用
收藏
页码:323 / 329
页数:7
相关论文
共 50 条
  • [41] Developments in Opto-Electronic Holography (OEH) for measurement of sound-induced eardrum motion
    Dobrev, Ivo
    Cheng, Jeffrey T.
    Khaleghi, Morteza
    Guignard, Jeremie
    Ravicz, Michael E.
    Rosowski, John J.
    Furlong, Cosme
    OPTICAL MEASUREMENT TECHNIQUES FOR STRUCTURES & SYSTEMS III, 2016, : 73 - 83
  • [42] Vibration Modes Study of Defective Eardrum Realized Using Digital Holographic Endoscopy
    Ding, Jianwen
    Zhou, Wenjing
    Yu, Yingjie
    CHINESE JOURNAL OF LASERS-ZHONGGUO JIGUANG, 2023, 50 (15):
  • [43] Dynamic Motion of Human Tympanic Membrane: Stroboscopic Holography Measurement and Finite Element Analysis
    Aernouts, J.
    Cheng, J. T.
    Furlong, C.
    Dirckx, J. J. J.
    Rosowski, J.
    OPTICAL MEASUREMENT TECHNIQUES FOR STRUCTURES & SYSTEMS2 (OPTIMESS2012), 2013, : 1 - 12
  • [44] Quantitative vibration amplitude measurement using stroboscopic phase shifting ESPI
    Jung, Hyun-Chul
    Kim, Koung-Suk
    ADVANCED NONDESTRUCTIVE EVALUATION I, PTS 1 AND 2, PROCEEDINGS, 2006, 321-323 : 95 - 98
  • [45] Possibility to break through limitation of measurement range in dual-wavelength digital holography
    李拓
    雷文秀
    孙鑫凯
    董军
    陶冶
    史祎诗
    Chinese Physics B, 2021, (09) : 137 - 145
  • [46] Possibility to break through limitation of measurement range in dual-wavelength digital holography*
    Li, Tuo
    Lei, Wen-Xiu
    Sun, Xin-Kai
    Dong, Jun
    Tao, Ye
    Shi, Yi-Shi
    CHINESE PHYSICS B, 2021, 30 (09)
  • [47] Wave front synthesis for comparative measurement in digital holography and TV holography
    Kornis, Janos
    Szabo, Andrds
    Zobory, Istvan
    SPECKLE06: SPECKLES, FROM GRAINS TO FLOWERS, 2006, 6341
  • [49] Measurement of Piezoelectric Transformer Vibrations by Digital Holography
    Psota, Pavel
    Ledl, Vit
    Dolecek, Roman
    Erhart, Jiri
    Kopecky, Vaclav
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (09) : 1962 - 1968
  • [50] Application of digital holography in temperature distribution measurement
    Wang, Guangjun
    Li, Yan
    Wang, Dayong
    Zhao, Jie
    HOLOGRAPHY, DIFFRACTIVE OPTICS, AND APPLICATIONS IV, 2010, 7848