Multi-element quantification of ancient/historic glasses by laser ablation inductively coupled plasma mass spectrometry using sum normalization calibration

被引:47
|
作者
van Elteren, Johannes T. [1 ]
Tennent, Norman H. [2 ,3 ]
Selih, Vid S. [1 ]
机构
[1] Natl Inst Chem, Ljubljana 1001, Slovenia
[2] Fyne Conservat Serv, St Catherines PA25 8BA, Argyll, Scotland
[3] Univ Amsterdam, Fac Humanities, NL-1012 GC Amsterdam, Netherlands
关键词
Laser ablation; Glasses; Calibration; Internal standardization; Interferences; Oxides normalization; Scaling algorithm; LA-ICP-MS; QUANTITATIVE-DETERMINATION; TRACE; SAMPLES; AEROSOLS; STANDARD; ELEMENTS; TOOL;
D O I
10.1016/j.aca.2009.04.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) for quantitative analysis of ancient/historic glasses is subject to calibration issues which have been addressed in this work. Since ancient/historic glasses have widely ranging matrix compositions, a complementary analysis by an alternative method is generally employed to determine at least one major element which can be used as an internal standard. We demonstrate that such a complementary analysis is unnecessary using a so-called Sum normalization calibration technique (mathematically formulated) by simultaneous measurement of 54 elements and normalizing them to 100% [w/w] based on their corresponding oxide concentrations. The crux of this approach is that by assuming a random internal standard concentration of a particular major oxide, e.g. SiO2, the normalization algorithm varies the internal standard concentration until the cumulated concentrations of all 54 elemental oxides reach 100% [w/w]. The fact that 54 elements are measured simultaneously predetermines the laser ablation mode to rastering. Nine glass standards. some replicating historic compositions, were used for calibration. The linearity of the calibration graphs (forced through the origin) represented by the relative standard deviations in the slope were between 0.1 and 6.6% using SiO2 as an internal standard. This allows high-accuracy determination of elemental oxides as confirmed by good agreement between found and reported values for Major and minor elemental oxides in some synthetic glasses with typical medieval composition (European Science Foundation 151 and 158). Also for trace elemental concentrations of lanthanides in a reference glass (P&H Developments Ltd. DLH7. a base glass composition with nominally 75 mu gg(-1) elements added) accurate data were obtained. Interferences from polyatomic species and doubly charged species on the masses of trace elements are possible. depending on the base composition of the glass. with Ba and Sb glasses showing potential interferences on some lanthanides. We showed that they may be reduced to a great extent by using an Octopole Reaction System although the overall sensitivity decreases which may be a problem for some low-level determinations. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 50 条
  • [31] Calibration of laser ablation inductively coupled plasma mass spectrometry for quantitative measurements of lead in bone
    Bellis, David J.
    Hetter, Katherine M.
    Jones, Joseph
    Amarasiriwardena, Dula
    Parsons, Patrick J.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (09) : 948 - 954
  • [32] Calibration of laser ablation inductively coupled plasma mass spectrometry using standard additions with dried solution aerosols
    Leach, JJ
    Allen, LA
    Aeschliman, DB
    Houk, RS
    ANALYTICAL CHEMISTRY, 1999, 71 (02) : 440 - 445
  • [33] Quantification of liver iron overload disease with laser ablation inductively coupled plasma mass spectrometry
    Philipp Kim
    Sabine Weiskirchen
    Ricarda Uerlings
    Astrid Kueppers
    Florian Stellmacher
    André Viveiros
    Heinz Zoller
    Ralf Weiskirchen
    BMC Medical Imaging, 18
  • [34] Novel quantification strategies using spinning sampling chamber laser ablation inductively coupled plasma mass spectrometry
    Claverie, Fanny
    Malherbe, Julien
    Bier, Naomi
    Molloy, John L.
    Long, Stephen E.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2012, 244
  • [35] Quantification of liver iron overload disease with laser ablation inductively coupled plasma mass spectrometry
    Kim, Philipp
    Weiskirchen, Sabine
    Uerlings, Ricarda
    Kueppers, Astrid
    Stellmacher, Florian
    Viveiros, Andre
    Zoller, Heinz
    Weiskirchen, Ralf
    BMC MEDICAL IMAGING, 2018, 18
  • [36] Multi-element analysis of size-segregated fine and ultrafine particulate via Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry
    Rovelli, Sabrina
    Nischkauer, Winfried
    Cavallo, Domenico M.
    Limbeck, Andreas
    ANALYTICA CHIMICA ACTA, 2018, 1043 : 11 - 19
  • [37] Neon inductively coupled plasma for laser ablation-inductively coupled plasma-mass spectrometry
    Petibon, CM
    Longerich, HP
    Horn, I
    Tubrett, MN
    APPLIED SPECTROSCOPY, 2002, 56 (05) : 658 - 664
  • [38] Inter-element fractionation and correction in laser ablation inductively coupled plasma mass spectrometry
    Chen, ZX
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (12) : 1823 - 1828
  • [40] Study of element distributions in weathered marble crusts using laser ablation inductively coupled plasma mass spectrometry
    Ulens, Katia
    Moens, Luc
    Dams, Richard
    Van Winckel, Stefaan
    Vandevelde, Leon
    Journal of analytical atomic spectrometry, 1994, 9 (11): : 1243 - 1248