In vitro evaluation of the resistance of three types of fixation to treat fractures of the mandibular angle

被引:9
|
作者
Gonzales, D. M. C. [1 ]
Spagnol, G. [1 ]
Sverzut, C. E. [1 ]
Trivellato, A. E. [1 ]
机构
[1] Univ Sao Paulo, Sch Dent Ribeirao Preto, Dept Oral & Maxillofacial Surg & Periodontol, Ave Cafe S-N,Fone 5516 33154137, BR-14040904 Ribeirao Preto, SP, Brazil
来源
关键词
Mandibular Fracture; Bone screws; Bone plates; 3D plate; TITANIUM MINIPLATE; PLATING TECHNIQUES; MANAGEMENT;
D O I
10.1016/j.bjoms.2016.09.017
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to compare the mechanical resistance of three different plates used to treat fractures of the mandibular angle: a regular 4-hole plate, a longer 4-hole plate (both positioned using the Champy technique), and a 3-dimensional plate positioned over the oblique line. Three equal groups of replicas of human dentate mandibles made out of polyurethane resin were used (n = 21 in each group). The force was applied perpendicular to the occlusal plane at a rate of 2 mm/minute at three different points: the first molar on the sectioned side; the first molar on the contralateral side; and between the central incisors. This was followed by a resistance-to-load test. The two varying factors( type of plate and site-of-load application) were tested by analysis of variance, and probabilities of less than 0.05 were accepted as significant. There were no significant differences between the subgroups, or between the mean values of the different types of plates (p = 0.925). The three types of plates showed similar mechanical behaviour, which showed that the 3-dimensional plates positioned over the oblique line can produce mechanical scores similar to those of conventional plates. (C) 2016 The British Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:136 / 140
页数:5
相关论文
共 50 条
  • [1] Biomechanical evaluation of new fixation devices for mandibular angle fractures
    Wittenberg, JM
    Mukherjee, DP
    Smith, BR
    Kruse, RN
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1997, 26 (01) : 68 - 73
  • [2] EVALUATION OF INTERNAL-FIXATION TECHNIQUES FOR MANDIBULAR ANGLE FRACTURES
    SHETTY, V
    BAKER, S
    CAPUTO, A
    BERTOLAMI, C
    [J]. JOURNAL OF DENTAL RESEARCH, 1990, 69 : 297 - 297
  • [3] LAG SCREW FIXATION OF MANDIBULAR ANGLE FRACTURES
    ELLIS, E
    GHALI, GE
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1991, 49 (03) : 234 - 243
  • [4] Fixation of mandibular angle fractures: Clinical studies
    Chrcanovic B.R.
    [J]. Oral and Maxillofacial Surgery, 2014, 18 (2) : 123 - 152
  • [5] COMPLICATIONS OF NONRIGID FIXATION OF MANDIBULAR ANGLE FRACTURES
    PASSERI, LA
    ELLIS, E
    SINN, DP
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1993, 51 (04) : 382 - 384
  • [6] MONOCORTICAL MINIPLATE FIXATION OF MANDIBULAR ANGLE FRACTURES
    LEVY, FE
    SMITH, RW
    ODLAND, RM
    MARENTETTE, LJ
    [J]. ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1991, 117 (02) : 149 - 154
  • [7] External skeletal fixation in fractures of the mandibular angle
    Converse, JM
    Waknitz, FW
    [J]. JOURNAL OF BONE AND JOINT SURGERY, 1942, 24 : 154 - 160
  • [8] Fixation of mandibular angle fractures: In vitro biomechanical assessments and computer-based studies
    Chrcanovic B.R.
    [J]. Oral and Maxillofacial Surgery, 2013, 17 (4) : 251 - 268
  • [9] Comparison of compressive strength between three different plates for mandibular angle fractures fixation
    Lyrio, Mariana C. Negreiros
    Monnazzi, Marcelo Silva
    De Moraes, Marcio
    Hochuli-Vieira, Eduardo
    Nunes Reis, Jose Mauricio
    Pereira-Filho, Valfrido Antonio
    [J]. JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2014, 42 (05) : E277 - E280
  • [10] TREATMENT OF MANDIBULAR ANGLE FRACTURES - PLATE AND SCREW FIXATION
    ASSAEL, LA
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1994, 52 (07) : 757 - 761