Testing the reliability of humidity generator through measurements traceable to calibration standards

被引:5
|
作者
Abd El-Galil, Doaa Mohamed [1 ]
Mahmoud, Essam [1 ]
机构
[1] NIS, Thermal Metrol Lab, PO 136, Giza 12211, Egypt
关键词
Relative humidity; Humidity generator; Stability; Uncertainty;
D O I
10.1016/j.measurement.2018.04.028
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper describes the metrological performance of NIS two-pressure humidity generator for testing the generator reliability after working for 13 years. In this work, humidity & temperature are observed and the measuring system is prepared for the calibration of humidity & temperature measurements. In the generator test chamber, relative humidity measurements are performed between 10% and 95% at fixed temperature of 25 degrees C and temperature measurements are performed between 10 degrees C and 65 degrees C at fixed humidity of 50% relative humidity. Results showed that the uniformity value is between +/- 0.13 degrees C and +/- 0.25 degrees C for temperature and between +/- 0.46% and +/- 0.62% for relative humidity. The value of stability is between 0.005 degrees C and 0.04 degrees C for temperature and 0.03% to 0.08% for relative humidity. Homogeneity value is in range between 0.09 degrees C and 0.12 degrees C for temperature, and between 0.43% and 0.52% for relative humidity, these satisfactory results showed the good performance of the generator. The expanded uncertainty of a coverage factor k=2 was found to be from +/- 0.2 degrees C to +/- 0.32 degrees C for temperature and +/- 0.8% to +/- 0.96% for relative humidity. The obtained results give us the confidence that NIS two-pressure humidity generator can be used as a high accurate humid air generator and reference instrument, for calibrating all types of dew-point meters and humidity sensors with high accuracy.
引用
收藏
页码:159 / 162
页数:4
相关论文
共 50 条
  • [21] The importance of calibration standards in jitter measurements
    Ishibe, K
    IEEE COMMUNICATIONS MAGAZINE, 2003, : S6 - S8
  • [22] Traceable measurements of the magnetic susceptibility of weakly-magnetic mass standards
    Pendrill, LR
    Davis, RS
    Neugebauer, L
    Källgren, H
    Thor, JE
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 523 - 524
  • [23] Frequency comb generator for calibration of laser frequency standards
    Merimaa, M
    Nyholm, K
    Vainio, M
    2005 Digest of the LEOS Summer Topical Meetings, 2005, : 91 - 92
  • [24] Standards, Measurements and Testing programme
    不详
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1998, 17 (01) : VI - VII
  • [25] ASSURING ACCURATE MEASUREMENTS AND TRACEABLE STANDARDS .2. MEASUREMENT ERRORS
    LANDVATER, J
    INDUSTRIAL RESEARCH & DEVELOPMENT, 1979, 21 (08): : 75 - 78
  • [26] Calibration method of flow meters for a divided flow humidity generator
    Takahashi, Chiharu
    Kitano, Hiroshi
    Sensors and Actuators, B: Chemical, 1996, B36 (1 -3 pt 2): : 522 - 527
  • [27] Calibration method of flow meters for a divided flow humidity generator
    Takahashi, C
    Kitano, H
    SENSORS AND ACTUATORS B-CHEMICAL, 1996, 36 (1-3) : 522 - 527
  • [28] Traceable calibration of AFM step height measurements for integrated circuit manufacturing
    Robert, James
    Banke, Bill
    Dixson, Ronald
    Strocchia-Rivera, Carlos
    METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XXII, PTS 1 AND 2, 2008, 6922 (1-2):
  • [30] Improving relative humidity measurements on Mars: new laboratory calibration measurements
    Hieta, Maria
    Jaakonaho, Iina
    Polkko, Jouni
    Lorek, Andreas
    Garland, Stephen
    de Vera, Jean-Pierre
    Genzer, Maria
    Harri, Ari-Matti
    GEOSCIENTIFIC INSTRUMENTATION METHODS AND DATA SYSTEMS, 2024, 13 (02) : 337 - 351