In situ determination of the dissolved oxygen in water

被引:0
|
作者
Yang, R [1 ]
Chen, HW [1 ]
Zhang, HQ [1 ]
Yu, AM [1 ]
Jin, QH [1 ]
机构
[1] Jilin Univ, Coll Chem, Inst Analyt Sci, Chungchun R&D Ctr Analyt Instruments, Changchun 130023, Peoples R China
来源
关键词
dissolved oxygen; in situ determination; portable photometer;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The indigo carmine was reduced by using cheap, easily attainable and non-poisonous reagents. Therefore, a special reagent was developed to measure the concentration of dissolved oxygen in water in situ. The molar absorption coefficient of the method is 1.35x10(4), and its dynamic linear range is 0-27 mg/L linearity coefficient of 0.997. The most oxidizing species, such as Cl-2, ClO-, Cu2+, Fe3+, V5+, and As( V), do not interfere the determination. It takes no more than 1 min to analyze one sample. The method was applied to determining some practical samples and all the results obtained were satisfactory.
引用
收藏
页码:1501 / 1503
页数:3
相关论文
共 16 条
  • [1] ANTHONY V, 1982, OPER RES, V30, P1
  • [2] COLORIMETRIC DETERMINATION OF LOW CONCENTRATIONS OF DISSOLVED OXYGEN IN WATER
    BUCHOFF, LS
    INGBER, NM
    BRADY, JH
    [J]. ANALYTICAL CHEMISTRY, 1955, 27 (09) : 1401 - 1404
  • [3] CASTANEDA HB, 1982, AM LAB, P119
  • [4] CHEN HW, 2000, ANAL INSTRUM, P17
  • [5] CHEN HW, 2000, ANAL INSTRUM, P28
  • [6] Clark LC, 1959, U.S. Patent, Patent No. 2913386
  • [7] *COMP GROUP PRACT, 2001, PRACT CHEM HDB, P622
  • [8] DONNA I, 1992, J BIO CYCLE, V33, P62
  • [9] HITCHMAN ML, 1978, MEASUREMENT DISSOLVE, P160
  • [10] HITCHMAN ML, 1978, MEASUREMENT DISSOLVE, P71