The impact of custom cutting guides on patient satisfaction and residual symptoms following total knee arthroplasty

被引:8
|
作者
Nam, Denis [1 ]
Nunley, Ryan M. [1 ]
Berend, Keith R. [2 ]
Lombardi, Adolph V. [2 ]
Barrack, Robert L. [1 ]
机构
[1] Washington Univ, Sch Med, Barnes Jewish Hosp, Dept Orthoped Surg, 660 S Euclid Ave,Campus Box 8233, St Louis, MO 63110 USA
[2] Joint Implant Surg Inc, 7277 Smiths Mill Rd,Suite 200, New Albany, OH 43054 USA
来源
KNEE | 2016年 / 23卷 / 01期
关键词
Total knee arthroplasty; Custom cutting guides; Patient satisfaction; Alignment; Standard instrumentation; FUNCTIONAL SCALE; INSTRUMENTATION; ALIGNMENT; OUTCOMES; BLOCKS; PAIN;
D O I
10.1016/j.knee.2015.08.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Custom cutting guides (CCGs) in total knee arthroplasty (TKA) use preoperative three-dimensional (3-D) imaging to manufacture cutting blocks specific to a patient's anatomy. The purpose of this study was to evaluate the impact of CCGs versus standard intramedullary and extramedullary guides on patient-reported satisfaction and residual symptoms following TKA. Methods: A retrospective, multicenter study was performed to compare a magnetic resonance imaging-based CCG system versus standard instrumentation. All patients received the same, cemented, fixed-bearing, cruciate-retaining component, and had a primary diagnosis of osteoarthritis. Data was collected by an independent, third party survey center blinded to surgical technique that administered telephone questionnaires assessing patient satisfaction and symptoms. Patient age, gender, minority status, education level, income, length of follow-up, and pre-arthritic UCLA scores were considered potential confounders and accounted for using multivariate logistic regression analyses. Results: 448 patients (107 CCGs, 341 standard) were successfully interviewed. At a mean follow-up of three years, there was no difference in percentage of patients reporting their knee to feel "normal" (74% CCG versus 78% standard, p = 0.37). Residual symptoms including knee stiffness (37% CCG versus 28% standard, p = 0.08) and difficulty getting in and out of car (34% CCG versus 30% standard, p = 0.40) remained high. Multivariate regression analyses demonstrated no differences between the two cohorts for both patient-reported satisfaction and residual symptoms (odds ratios 0.72 to 1.48; p = 0.10 to 0.81). Conclusion: When interviewed by an independent, blinded third party, the use of CCGs in TKA did not improve patient-reported satisfaction or residual symptoms versus the use of standard alignment guides. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:144 / 148
页数:5
相关论文
共 50 条
  • [21] The Mark Coventry Award Custom Cutting Guides Do Not Improve Total Knee Arthroplasty Clinical Outcomes at 2 Years Followup
    Nam, Denis
    Park, Andrew
    Stambough, Jeffrey B.
    Johnson, Staci R.
    Nunley, Ryan M.
    Barrack, Robert L.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (01) : 40 - 46
  • [22] Patient Satisfaction after Total Knee and Hip Arthroplasty
    Lau, Rick L.
    Gandhi, Rajiv
    Mahomed, Safiyyah
    Mahomed, Nizar
    CLINICS IN GERIATRIC MEDICINE, 2012, 28 (03) : 349 - +
  • [23] Does Preoperative Opioid Consumption Influence Patient Satisfaction following Total Knee Arthroplasty?
    Smith, Austin F.
    Smith, Nolan S.
    Smith, Langan S.
    Yakkanti, Madhusudhan R.
    Malkani, Arthur L.
    JOURNAL OF KNEE SURGERY, 2023, 36 (13) : 1374 - 1379
  • [24] Do manual cutting guides for total knee arthroplasty introduce systematic error?
    Erik Hohmann
    Kevin Tetsworth
    International Orthopaedics, 2016, 40 : 277 - 284
  • [25] Do manual cutting guides for total knee arthroplasty introduce systematic error?
    Hohmann, Erik
    Tetsworth, Kevin
    INTERNATIONAL ORTHOPAEDICS, 2016, 40 (02) : 277 - 284
  • [26] Predictors of Patient Satisfaction Following Unicondylar Knee Arthroplasty
    Plancher, Kevin D.
    Bishai, S. K.
    Areson, D. G.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2008, 40 (05): : S144 - S144
  • [27] Advantages of patient-specific cutting guides with disposable instrumentation in total knee arthroplasty: a case control study
    Kevin Moerenhout
    Behrang Allami
    Georgios Gkagkalis
    Olivier Guyen
    Brigitte M. Jolles
    Journal of Orthopaedic Surgery and Research, 16
  • [28] Advantages of patient-specific cutting guides with disposable instrumentation in total knee arthroplasty: a case control study
    Moerenhout, Kevin
    Allami, Behrang
    Gkagkalis, Georgios
    Guyen, Olivier
    Jolles, Brigitte M.
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
  • [29] Patient specific cutting guides versus an imageless, computer-assisted surgery system in total knee arthroplasty
    Nam, Denis
    Maher, Patrick A.
    Rebolledo, Brian J.
    Nawabi, Danyal H.
    McLawhorn, Alexander S.
    Pearle, Andrew D.
    KNEE, 2013, 20 (04): : 263 - 267
  • [30] Assessing the accuracy of patient-specific guides for total knee arthroplasty
    Seon, Jong-Keun
    Park, Hyeong-Won
    Yoo, Seung-Hyun
    Song, Eun-Kyoo
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2016, 24 (11) : 3678 - 3683