Quality assurance for MRI: practical experience

被引:68
|
作者
Firbank, MJ [1 ]
Harrison, RM
Williams, ED
Coulthard, A
机构
[1] Univ Newcastle Upon Tyne, Royal Victoria Infirm, Dept Radiol, Newcastle Upon Tyne NE1 4LP, Tyne & Wear, England
[2] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
[3] Sunderland Royal Hosp, Dept Reg Med Phys, Sunderland SR4 7TP, England
来源
BRITISH JOURNAL OF RADIOLOGY | 2000年 / 73卷 / 868期
关键词
D O I
10.1259/bjr.73.868.10844863
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to propose guidelines for quality assurance (QA) in MRI, based on a comprehensive assessment of QA parameters undertaken on a busy clinical MRI scanner over the course of 1 year. QA phantoms supplied by the scanner manufacturer were used together with the Eurospin MRI phantom set. Signal-to-noise ratio (SNR) and image uniformity were measured daily from spin echo images acquired using a quadrature send-receive head coil and from a gradient echo sequence using the Helmholtz body coil. The voltage of the transmit radiofrequency (RF) amplifier was noted. Monthly measurements of slice thickness, geometric distortion, slice position, image resolution and image ghosting were acquired using the head coil. In addition, SNR was measured monthly on a selection of commonly used coils. Apart from some drift of the RF amplifier voltage, all measurements were within acceptable limits and were stable over the course of 1 year. Satisfactory measurements of SNR were possible using the simple phantom supplied with the scanner. The SNR, geometric distortion and RF amplifier voltage are simple to determine and can be measured in less than 15 min by the scanner operator, using the scanner software. Weekly recording of these parameters is recommended for busy clinical MRI scanners, as this should allow deviations from acceptable limits to be identified early. Such inhouse checks can usefully be compared with the less frequent estimations performed by the service engineer. Comprehensive QA routines are discussed for systems used for quantitative measurements.
引用
收藏
页码:376 / 383
页数:8
相关论文
共 50 条
  • [41] QUALITY ASSURANCE TRAINING BASED ON USNRC EXPERIENCE
    VOLLMER, RH
    [J]. TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1975, 21 (JUN): : 84 - 85
  • [42] Quality Assurance in multi-site MRI trials
    Fox, Robert
    Zhou, Xiaopeng
    Sakaie, Ken
    Debbins, Josef
    Kirsch, John
    Lowe, Mark
    [J]. MULTIPLE SCLEROSIS JOURNAL, 2016, 22 : 40 - 41
  • [43] PERSPECTIVES ON THE SMRI MRI ECONOMICS AND QUALITY ASSURANCE COMMITTEE
    YOUNG, SW
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (01): : 133 - 134
  • [44] Stability Test for MRI Quality Assurance in Radiation Therapy
    McClain, B.
    Curcuru, A.
    Green, O.
    Mutic, S.
    Gach, H.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3077 - 3077
  • [45] MRI quality assurance for gamma knife stereotactic radiosurgery
    Damyanovich, A. Z.
    Zhang, B.
    Mikulis, D.
    Bissonnette, J-P.
    Bernstein, M.
    Menard, C.
    Jaffray, D. A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S2 - S2
  • [46] QUALITY ASSURANCE AND QUALITY CONTROL IN THE ANALYTICAL CHEMICAL LABORATORY. A PRACTICAL APPROACH
    Siepak, Jerzy
    [J]. ECOLOGICAL CHEMISTRY AND ENGINEERING A-CHEMIA I INZYNIERIA EKOLOGICZNA A, 2010, 17 (01): : 157 - 158
  • [47] Quality assurance of joint degree programs from the perspective of quality assurance agencies: experience in East Asia
    Hou , Yung-Chi
    Ince, Martin
    Tsai, Sandy
    Wang, Wayne
    Hung, Vicky
    Jiang, Chung Lin
    Chen, Karen Hui-Jung
    [J]. HIGHER EDUCATION RESEARCH & DEVELOPMENT, 2016, 35 (03) : 473 - 487
  • [48] A Practical Approach to Quality Assurance Implementation in Clinical Echocardiographic Laboratories
    Daubert, Melissa A.
    Barnhart, Huiman X.
    Yow, Eric
    Rabineau, Dawn
    Douglas, Pamela S.
    [J]. CIRCULATION, 2012, 126 (21)
  • [49] An Efficient and Practical MR Flow Quantification Quality Assurance Method
    Lin, C.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3255 - 3255
  • [50] The evaluation of quality assurance: developing and testing practical methods for managers
    Catsambas, TT
    Kelley, E
    Legros, S
    Massoud, R
    Bouchet, B