Quality assurance for MRI: practical experience

被引:68
|
作者
Firbank, MJ [1 ]
Harrison, RM
Williams, ED
Coulthard, A
机构
[1] Univ Newcastle Upon Tyne, Royal Victoria Infirm, Dept Radiol, Newcastle Upon Tyne NE1 4LP, Tyne & Wear, England
[2] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
[3] Sunderland Royal Hosp, Dept Reg Med Phys, Sunderland SR4 7TP, England
来源
BRITISH JOURNAL OF RADIOLOGY | 2000年 / 73卷 / 868期
关键词
D O I
10.1259/bjr.73.868.10844863
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to propose guidelines for quality assurance (QA) in MRI, based on a comprehensive assessment of QA parameters undertaken on a busy clinical MRI scanner over the course of 1 year. QA phantoms supplied by the scanner manufacturer were used together with the Eurospin MRI phantom set. Signal-to-noise ratio (SNR) and image uniformity were measured daily from spin echo images acquired using a quadrature send-receive head coil and from a gradient echo sequence using the Helmholtz body coil. The voltage of the transmit radiofrequency (RF) amplifier was noted. Monthly measurements of slice thickness, geometric distortion, slice position, image resolution and image ghosting were acquired using the head coil. In addition, SNR was measured monthly on a selection of commonly used coils. Apart from some drift of the RF amplifier voltage, all measurements were within acceptable limits and were stable over the course of 1 year. Satisfactory measurements of SNR were possible using the simple phantom supplied with the scanner. The SNR, geometric distortion and RF amplifier voltage are simple to determine and can be measured in less than 15 min by the scanner operator, using the scanner software. Weekly recording of these parameters is recommended for busy clinical MRI scanners, as this should allow deviations from acceptable limits to be identified early. Such inhouse checks can usefully be compared with the less frequent estimations performed by the service engineer. Comprehensive QA routines are discussed for systems used for quantitative measurements.
引用
收藏
页码:376 / 383
页数:8
相关论文
共 50 条
  • [1] QUALITY ASSURANCE IN PROJECT MANAGEMENT -PRACTICAL EXPERIENCE
    Rakovic, Radoslav
    [J]. 16TH ISSAT INTERNATIONAL CONFERENCE ON RELIABILITY AND QUALITY IN DESIGN, 2010, : 372 - 376
  • [2] QUALITY ASSURANCE - OPPORTUNITY FOR PRACTICAL STUDENT EXPERIENCE
    BRAYMAN, SJ
    [J]. AMERICAN JOURNAL OF OCCUPATIONAL THERAPY, 1992, 46 (04): : 369 - 371
  • [3] SOME PRACTICAL EXPERIENCE WITH A SOFTWARE QUALITY ASSURANCE PROGRAM
    GUSTAFSON, GG
    KERR, RJ
    [J]. COMMUNICATIONS OF THE ACM, 1982, 25 (01) : 4 - 12
  • [4] How to teach practical Software Quality Assurance - An experience report
    Thompson, JB
    Edwards, HM
    [J]. FIRST ASIA-PACIFIC CONFERENCE ON QUALITY SOFTWARE, PROCEEDINGS, 2000, : 181 - 187
  • [5] PRACTICAL QUALITY ASSURANCE
    CONNELL, J
    BRICE, L
    [J]. DATAMATION, 1985, 31 (05): : 106 - &
  • [6] QUALITY ASSURANCE IN MRI
    PRICE, RR
    [J]. MEDICAL PHYSICS, 1986, 13 (04) : 624 - 624
  • [7] MRI - Quality assurance
    Lazeyras, Francois
    [J]. PHYSICS FOR MEDICAL IMAGING APPLICATIONS, 2007, 240 : 49 - 53
  • [8] QUALITY ASSURANCE IN CARDIAC-SURGERY - HISTORY, PRELIMINARIES, PRACTICAL EXPERIENCE
    KALMAR, P
    [J]. THORACIC AND CARDIOVASCULAR SURGEON, 1990, 38 (02): : 108 - 114
  • [9] PRACTICAL NUCLEAR QUALITY ASSURANCE
    SCHATZMAN, JT
    [J]. MATERIALS EVALUATION, 1976, 34 (09) : P16 - P16
  • [10] PRACTICAL QUALITY ASSURANCE FOR DIAGNOSTIC TELERADIOLOGY
    FREY, GD
    OLINGER, R
    SPICER, KM
    SCHENING, T
    CONWAY, WF
    YOUNG, JW
    [J]. RADIOLOGY, 1995, 197 : 449 - 449