Interexaminer and intraexaminer reliability of the microperimeter MP-1

被引:32
|
作者
Weingessel, B. [1 ]
Sacu, S. [1 ]
Vecsei-Marlovits, P. V. [1 ]
Weingessel, A. [2 ]
Richter-Mueksch, S. [1 ]
Schmidt-Erfurth, U. [1 ]
机构
[1] Med Univ Vienna, Dept Ophthalmol, A-1090 Vienna, Austria
[2] Univ Technol, Dept Stat & Probabil Theory, Vienna, Austria
关键词
microperimetry; reliability; fixation stability; macular sensitivity; SCANNING LASER OPHTHALMOSCOPE; DIABETIC MACULAR EDEMA; OCULAR HYPERTENSION TREATMENT; OPTICAL COHERENCE TOMOGRAPHY; TEST-RETEST VARIABILITY; FUNDUS PERIMETRY; INTERRATER AGREEMENT; GLAUCOMA; REPRODUCIBILITY; THRESHOLD;
D O I
10.1038/eye.2008.237
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the interexaminer and intraexaminer reliability of macular microperimetry using the microperimeter MP-1. Methods Participants: Fifteen healthy volunteers younger than 40 years of age (Group 1), 15 healthy subjects over 60 years (Group 2), and five patients with age-related macular degeneration (Group 3). Observation procedure: Two examiners (E1 and E2) measured, in random order, interexaminer (E2-E1a) reliability. Another examination was undergone by one of the examiners a week later to evaluate the intraexaminer (E1b-E1a) reliability. Main outcome measures: Macular sensitivity (mean threshold (decibel)) and stability of fixation were determined using MP1 microperimetry. Agreement was analysed by means of Bland-Altman plots and by the determination of the intraclass correlation coefficient. Results The interexaminer (E2-E1a) and the intraexaminer (E1b-E1a) differences in the mean threshold values were not statistically significant (P = 0.850, 95% confidence Interval (CI) = -0.265 to 0.319; P = 0.246, 95% CI = -0.099 to 0.375, respectively). Limits of agreement and intraclass correlation coefficients also showed good agreement in each group. Conclusions A good reliability was found for the mean threshold values in all the three groups, indicating examiner-independent measurements. Eye (2009) 23, 1052-1058; doi:10.1038/eye.2008.237; published online 1 August 2008
引用
收藏
页码:1052 / 1058
页数:7
相关论文
共 50 条
  • [21] Fixation Characteristics of Severe Amblyopia with Eccentric Fixation and Central Fixation Assessed by the MP-1 Microperimeter
    Wang, Shu
    Tian, Tian
    Zou, Leilei
    Wu, Sujia
    Liu, Yan
    Wen, Wen
    Liu, Hong
    SEMINARS IN OPHTHALMOLOGY, 2021, 36 (5-6) : 360 - 365
  • [22] Retinal-Specific Measurement of Dark Adapted Visual Function Using the Mp-1 Microperimeter
    Crossland, M. D.
    Luong, V. A.
    Rubin, G. S.
    Fitzke, F. W.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [23] Interexaminer and intraexaminer reliability of pupil diameter measurement using the Colvard pupillometer
    Yoon, M
    Schmidt, G
    Lietman, T
    McLeod, SD
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U11 - U11
  • [24] Intersession test-retest variability of conventional and novel parameters using the MP-1 microperimeter
    Wong, Evan N.
    Mackey, David A.
    Morgan, William H.
    Chen, Fred Kuanfu
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 29 - 42
  • [25] Optical biometry of the anterior eye segment: Interexaminer and intraexaminer reliability of ACMaster
    Sacu, S
    Findl, O
    Buehl, W
    Kiss, B
    Gleiss, A
    Drexler, W
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (12): : 2334 - 2339
  • [26] A REVIEW OF INTRAEXAMINER AND INTEREXAMINER RELIABILITY OF STATIC SPINAL PALPATION: A LITERATURE SYNTHESIS
    Haneline, Michael T.
    Young, Morgan
    JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS, 2009, 32 (05) : 379 - 386
  • [27] Intraexaminer and interexaminer reliability for palpation of the cranial rhythmic impulse at the head and sacrum
    Moran, RW
    Gibbons, P
    JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS, 2001, 24 (03) : 183 - 190
  • [28] Changes in Retinal Sensitivity in the Natural History of Geographic Atrophy as Measured Using the Nidek MP-1 Microperimeter
    Meleth, A. D.
    Mettu, P.
    Ferris, F. L., III
    Chew, E. Y.
    Wong, W. T.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [29] Fixation Behavior in Primary Open Angle Glaucoma at Early and Moderate Stage Assessed by the MicroPerimeter MP-1
    Shi, Yan
    Liu, Mugen
    Wang, Xin
    Zhang, Chun
    Huang, Ping
    JOURNAL OF GLAUCOMA, 2013, 22 (02) : 169 - 173
  • [30] INTRAEXAMINER AND INTEREXAMINER RELIABILITY OF THE CHIROPRACTIC BIOPHYSICS LATERAL LUMBAR RADIOGRAPHIC MENSURATION PROCEDURE
    TROYANOVICH, SJ
    ROBERTSON, GA
    HARRISON, DD
    HOLLAND, B
    JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS, 1995, 18 (08) : 519 - 524