Reduction of Motion Artifacts in Carotid MRI Using Free-Induction Decay Navigators
被引:13
|
作者:
Dyverfeldt, Petter
论文数: 0引用数: 0
h-index: 0
机构:
Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USAUniv Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Dyverfeldt, Petter
[1
]
Deshpande, Vibhas S.
论文数: 0引用数: 0
h-index: 0
机构:
Siemens Med Solut USA Inc, San Francisco, CA USAUniv Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Deshpande, Vibhas S.
[2
]
Kober, Tobias
论文数: 0引用数: 0
h-index: 0
机构:
Siemens Healthcare Sector IM S AW, Adv Clin Imaging Technol, Lausanne, SwitzerlandUniv Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Kober, Tobias
[3
]
Krueger, Gunnar
论文数: 0引用数: 0
h-index: 0
机构:
Siemens Healthcare Sector IM S AW, Adv Clin Imaging Technol, Lausanne, SwitzerlandUniv Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Krueger, Gunnar
[3
]
Saloner, David
论文数: 0引用数: 0
h-index: 0
机构:
Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Vet Affairs Med Ctr, San Francisco, CA 94121 USAUniv Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
Saloner, David
[1
,4
]
机构:
[1] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94143 USA
[2] Siemens Med Solut USA Inc, San Francisco, CA USA
[3] Siemens Healthcare Sector IM S AW, Adv Clin Imaging Technol, Lausanne, Switzerland
[4] Vet Affairs Med Ctr, San Francisco, CA 94121 USA
Purpose: To develop a framework for prospective free-induction decay (FID)-based navigator gating for suppression of motion artifacts in carotid magnetic resonance imaging (MRI) and to assess its capability in vivo. Materials and Methods: An FID-navigator, comprising a spatially selective low flip-angle sinc-pulse followed by an analog-to-digital converter (ADC) readout, was added to a conventional turbo spin-echo (TSE) sequence. Real-time navigator processing delivered accept/reject-and-reacquire decisions to the sequence. In this Institutional Review Board (IRB)-approved study, seven volunteers were scanned with a 2D T2-weighted TSE sequence. A reference scan with volunteers instructed to minimize motion as well as nongated and gated scans with volunteers instructed to perform different motion tasks were performed in each subject. Multiple image quality measures were employed to quantify the effect of gating. Results: There was no significant difference in lumen-to-wall sharpness (2.3 +/- 0.3 vs. 2.3 +/- 0.4), contrast-to-noise ratio (CNR) (9.0 +/- 2.0 vs. 8.5 +/- 2.0), or image quality score (3.1 +/- 0.9 vs. 2.6 +/- 1.2) between the reference and gated images. For images acquired during motion, all image quality measures were higher (P < 0.05) in the gated compared to nongated images (sharpness: 2.3 +/- 0.4 vs. 1.8 +/- 0.5, CNR: 8.5 +/- 2.0 vs. 7.2 +/- 2.0, score: 2.6 +/- 1.2 vs. 1.8 +/- 1.0). Conclusion: Artifacts caused by the employed motion tasks deteriorated image quality in the nongated scans. These artifacts were alleviated with the proposed FID-navigator.