Comparison of Tono-Pen AVIA Intraocular Pressure Measurements Performed at Limbus With Central Corneal Tono-Pen AVIA Intraocular Pressure

被引:5
|
作者
Sharma, Reetika [1 ]
Majumdar, Saptorishi [1 ]
Sobti, Amit [1 ]
Arora, Tarun [1 ]
Agarwal, Tushar [1 ]
Dada, Tanuj [1 ]
机构
[1] All India Inst Med Sci, Dr Rajendra Prasad Ctr Ophthalm Sci, Glaucoma Facil, New Delhi 110029, India
关键词
Tono-Pen AVIA; limbus; correlation; central cornea; GOLDMANN APPLANATION TONOMETER; CLINICAL-EVALUATION; CONTACT-LENSES; THICKNESS; EYES;
D O I
10.1097/ICO.0b013e31827c9d25
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare Tono-Pen AVIA (TPA) intraocular pressure (IOP) measurements performed at the limbus with central corneal Tono-Pen AVIA (TPA) intraocular pressure. Methods: Fifty-one patients attending our outpatient department with a best corrected visual acuity of >= 20/40, no corneal disease, and no astigmatism greater than 1.5 diopters underwent routine ophthalmological examination. Patients with ametropia greater than or equal to +/- 5 diopters, a history of recent contact lens wear, and central corneal thickness of <480 mu m or >550 mu m were excluded from the study. IOP was recorded first in the corneal center, followed by tonometry measurements obtained from the temporal, nasal, superior, and inferior limbus using TPA. IOP measurements were compared using the paired t test. Correlation was derived by Pearson correlation coefficient, and regression equations were derived. Results: The mean age of patients in our study was 43.86 +/- 15.39 years. The mean TPA IOP from the central cornea was 16.59 +/- 3.75 mm Hg. There was a statistically significant difference between Tono-Pen IOP from the central cornea and that recorded from all 4 limbal positions (P < 0.001). There was a positive correlation between the central and the limbal IOP obtained with Tono-Pen. Correlation was maximum for the temporal limbus. Conclusions: Limbal TPA IOP is significantly higher than central TPA IOP. There remains a good correlation between central and limbal IOP. This was found to be greatest for the temporal limbus.
引用
收藏
页码:943 / 946
页数:4
相关论文
共 50 条
  • [21] Intraocular pressure measurement in cynomolgus monkeys - Tono-Pen versus manometry
    Peterson, JA
    Kiland, JA
    Croft, MA
    Kaufman, PL
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (06) : 1197 - 1199
  • [22] TONO-PEN VS MANOMETER - COMPARATIVE INTRAOCULAR-PRESSURE MEASUREMENTS IN CADAVER EYES
    HESSEMER, V
    ROSSLER, R
    JACOBI, KW
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1989, 195 (01) : 23 - 27
  • [23] Measurements of Intraocular Pressure with Tono-Pen and Goldmann Applanation Tonometers in Subjects with Systemic Illness
    Hsu, Sheng-Yao
    Lin, Shu-Fang
    Hsu, Ai-Hsi
    TZU CHI MEDICAL JOURNAL, 2007, 19 (04): : 241 - 244
  • [24] Intraocular pressure measurements with the Tono-Pen XL® and Perkins® applanation tonometers in horses and cattle
    Andrade, Silvia Franco
    Palozzi, Rodrigo de Jesus
    Rolim, Rodrigo Duarte
    Cecilio Junqueira, Jose Ricardo
    do Valle, Heloisa Ferreira
    da Silva, Marcelo Kurak
    Maia, Marcelo Augusto
    Zampiere, Rodrigo Augusto
    CIENCIA RURAL, 2013, 43 (05): : 865 - 870
  • [25] 24-hour intraocular pressure (IOP) monitoring with Goldmann applanation (GAT) tonometry and Tono-Pen AVIA® in keratoconic eyes.
    Merula, Rafael Vidal
    Cronemberger, Sebastiao
    Veloso, Artur C.
    Diniz-Filho, Alberto
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [26] Comparison of Tono-Pen and Pneumotonometer Measurements from Scleral Locations with Intraocular Pressure Measurements from the Cornea
    Tai, Tak Yee Tania
    Harizman, Noga
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [27] Comparison of Tono-Pen and Goldmann applanation tonometers for measurement of intraocular pressure in healthy children
    Bradfield, Yasmin S.
    Kaminski, Brett M.
    Repka, Michael X.
    Melia, Michele
    JOURNAL OF AAPOS, 2012, 16 (03): : 242 - 248
  • [28] The influence of central corneal thickness and corneal curvature on Intraocular pressure measured by Tono-Pen and rebound tonometer in children
    Sahin, Afsun
    Basmak, Hikmet
    Yidirim, Nigun
    JOURNAL OF GLAUCOMA, 2008, 17 (01) : 57 - 61
  • [29] Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen
    Ma, Di
    Chen, Chong-Bo
    Liang, Jiajian
    Lu, Zhihao
    Chen, Haoyu
    Zhang, Mingzhi
    SCIENTIFIC REPORTS, 2016, 6
  • [30] Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen
    Di Ma
    Chong-Bo Chen
    Jiajian Liang
    Zhihao Lu
    Haoyu Chen
    Mingzhi Zhang
    Scientific Reports, 6