Analyzing 3.0T MRI-scanners for implementation in radiosurgery

被引:0
|
作者
Mack, Andreas [1 ]
Wolff, Robert [1 ]
Scheib, Stefan G. [1 ]
Rieker, Marcus [1 ]
Weltz, Dirk [1 ]
Mack, Guenther [1 ]
Kreiner, Hans-Juerg [1 ]
Pilatus, Ulrich [1 ]
Zanella, Friedhelm E. [1 ]
Boettcher, Heinz D. [1 ]
Seifert, Volker [1 ]
机构
[1] Gamma Knife Ctr, Frankfurt, Germany
来源
RADIOSURGERY, VOL 6 | 2006年 / 6卷
关键词
D O I
10.1159/000093716
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Given the high mechanical accuracy of the Leksell Gamma Knife', the most sensitive technical factor having an influence on the overall precision of radiosurgery is the imaging (mainly magnetic resonance imaging (MRI)) study. The new generation of MRI-Scanners with field strength up to 3.0 T deliver promising image quality concerning anatomical resolution and contrast but critical features are the sensitivity to susceptibility and chemical shift effects both creating artefacts. Methods: The 3.0 T MRI-Scanner (Siemens 'Trio') was analyzed and compared to a Siemens Magnetom 'Expert' 1.0 T and to a Philips 'Gyroscan' 1.5 T. The analyses was performed in three steps: (1) The evaluation of the magnitude of error was performed within transversal slices in two orientations (axial/coronal) by using a cylindrical phantom with an embedded grid. (2) The deviations were determined for 21 targets in a slab phantom with known geometrical positions within the stereotactic frame. (3) Distortions caused by chemical shift and/or susceptibility effects were analyzed in a head-phantom. An in-house developed software was used for data analyses. Results: The 3.0 T MRI-Scanner was analyzed using sequences in axial and coronal orientations. The mean deviation was < 0.3 mm in axial and < 0.4 mm in coronal orientation. For the known targets the maximum deviation came up to 1.16 mm (far from the center). Due to inhomogeneities a shift in z-direction up to 1.5 mm was observed for a data set which was shown compressed by 1.2 mm. By optimizing these parameters in the protocol these inaccuracies could be reduced to < 1. 1 mm. Conclusions: The scanner showed up with sophisticating anatomical contrast and resolution in comparison to the established 1.0 T and 1.5 T scanners. However, due to the high-field strength the field within the head coil is very sensitive to inhomogeneities and therefore 3.0 T imaging data still have to be handled with care. Copyright (c) 2006 S. Karger AG, Basel.
引用
收藏
页码:60 / 70
页数:11
相关论文
共 50 条
  • [21] Analysis of ulcerative colitis with 3.0T MRI diffusion weighted imaging
    Zhao, Xin
    Sun, Haoran
    Cao, Xiaocang
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2013, 28 : 592 - 593
  • [22] Design and Magnetic Field Simulation of 3.0T MRI Superconducting Magnet
    Yue, Zengkai
    Sun, Zhizhou
    Liu, Zhizhen
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (02)
  • [23] 3.0T MRI系统图像质量控制检测
    高万河
    赵艳琼
    潘文哲
    中国医学装备, 2017, 14 (07) : 59 - 62
  • [24] 3.0T MRI肝脏动态增强扫描的护理
    许琴
    钟丽华
    杨妹
    华夏医学, 2015, 28 (04) : 119 - 120
  • [25] Reproducibility of the carotid vessel wall area and dimensions at 3.0T MRI
    Dehnavi, R. Alizadeh
    Doornbos, J.
    Tamsma, J. T.
    Stuber, M.
    Putter, H.
    Lamb, H. J.
    De Roos, A.
    ATHEROSCLEROSIS SUPPLEMENTS, 2006, 7 (03) : 267 - 267
  • [26] ANGIOGENESIS OF OVARIAN CANCER: CHARACTERIZATION BY 3.0T DIFFUSION WEIGHTED MRI
    Lehikoinen, A.
    Anttila, M.
    Rautiainen, S.
    Arponen, O.
    Kivela, A.
    Hamalainen, K.
    Kosma, V. M.
    Yla-Herttuala, S.
    Vanninen, R.
    Sallinen, H.
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2015, 25 (09) : 484 - 484
  • [27] 3.0T MRI对直肠癌T分期的价值
    陈腾飞
    刘艺超
    章辉庆
    世界最新医学信息文摘, 2018, 18 (88) : 224 - 225
  • [28] Evaluation of Cardiac Implantable Electronic Devices-Induced Distortion on 0.35T and 3.0T MR Scanners
    Omidi, A.
    Weiss, E.
    Al Khalifa, M.
    Kim, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : E173 - E173
  • [29] 近侧指间关节损伤的3.0T MRI表现
    李振芝
    王晓华
    于进超
    乔慧洁
    实用手外科杂志, 2021, 35 (03) : 345 - 347
  • [30] Diagnostic Value of 3.0T MRI in Cartilage Injury Grading of Knee Osteoarthritis
    Zhang, Xiaoming
    Tong, Haiyang
    Zhang, Jin
    Xu, Jinyong
    Xia, Shiyong
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2020, 10 (12) : 2979 - 2984