The study of reconstruction image quality resulting from geometric error in micro-CT system

被引:0
|
作者
Xi, Dongxing [1 ,2 ]
Gui, Jianbao [1 ,2 ]
Zou, Jing [1 ,2 ]
Rong, Junyan [1 ,2 ]
Hu, Zhanli [1 ,2 ]
Liu, Xin [1 ,2 ]
Zheng, Hairong [1 ,2 ]
机构
[1] Chinese Acad Sci, Paul C Lauterbur Res Ctr Biomed Imaging, Shenzhen Inst Adv Technol, Inst Biomed & Hlth Engn, Shenzhen, Peoples R China
[2] Chinese Acad Sci, Key Lab Biomed Informat & Hlth Engn, Beijing 100864, Peoples R China
关键词
cone-beam CT; geometrical parameter; geometrical calibration; three-dimensional reconstruction; BEAM; CALIBRATION; TOMOGRAPHY; ALGORITHM;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The reconstruction images often suffer from artifacts caused by misaligned geometry of flat detector in micro-CT. This paper mainly simulates the causality between the reconstruction image quality and the flat detector geometry by the eight-spheres phantom. The simulation results show that the most influential parameter is beta (the rotation angle of the detector plane along the center point O) among the five detector parameters, next is a, b, theta and a degressively. We should control a under 7 degrees and a, b offset lower than 8 pixels, when adjusting micro-CT systems. We also show that the relevant correction is efficient to improve the reconstruction image quality affected by the tiny geometric error.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] The impact of spectral filtration on image quality in micro-CT system
    Ren, Liqiang
    Ghani, Muhammad U.
    Wu, Di
    Zheng, Bin
    Chen, Yong
    Yang, Kai
    Wu, Xizeng
    Liu, Hong
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 301 - 315
  • [2] Implementation of short-scan reconstruction with compensation for geometric alignment for a micro-CT system
    Sawyer, SA
    Frey, EC
    He, B
    Wang, YC
    Tsui, BMW
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2981 - 2984
  • [3] Correction of micro-CT image geometric artefacts based on marker
    Tao, Huanjie
    Lu, Xiaobo
    [J]. IET IMAGE PROCESSING, 2018, 12 (06) : 888 - 895
  • [4] Image Reconstruction from A Reduced Number of Projections in Micro-CT Specimen Imaging
    Han, Xiao
    Bian, Junguo
    Eaker, Diane R.
    Kline, Timothy L.
    Sidky, Emil Y.
    Ritman, Erik L.
    Pan, Xiaochuan
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3466 - 3469
  • [5] Image Reconstruction for Hybrid True-Color Micro-CT
    Xu, Qiong
    Yu, Hengyong
    Bennett, James
    He, Peng
    Zainon, Rafidah
    Doesburg, Robert
    Opie, Alex
    Walsh, Mike
    Shen, Haiou
    Butler, Anthony
    Butler, Phillip
    Mou, Xuanqin
    Wang, Ge
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (06) : 1711 - 1719
  • [6] Assessment of image quality parameters of a novel micro-CT system compared to a conventional CT geometry
    Kumar, K.
    Nezhad, N. Saeid
    Mueller, B. H.
    Tischenko, O.
    Hoeschen, C.
    [J]. MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [7] Geometric Calibration of a Micro-CT System and Performance for Insect Imaging
    Hu, Zhanli
    Gui, Jianbao
    Zou, Jing
    Rong, Junyan
    Zhang, Qiyang
    Zheng, Hairong
    Xia, Dan
    [J]. IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2011, 15 (04): : 655 - 660
  • [8] Investigations into the Geometric Calibration and Systematic Effects of a Micro-CT System
    Hardner, Matthias
    Liebold, Frank
    Wagner, Franz
    Maas, Hans-Gerd
    [J]. SENSORS, 2024, 24 (16)
  • [9] Geometric calibration for a dual tube/detector micro-CT system
    Johnston, Samuel M.
    Johnson, G. Allan
    Badeaa, Cristian T.
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 1820 - 1829
  • [10] Improved Dynamic Micro-CT Imaging Binning for Optimized Image Reconstruction
    Ismail, M. K.
    Xin, Y.
    Hamedani, H.
    Amzajerdian, F.
    Baron, R.
    Loza, L.
    Ruppert, K.
    Kadlecek, S.
    Rizi, R. R.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2023, 207