Adjustable buttons for ACL graft cortical fixation partially fail with cyclic loading and unloading

被引:20
|
作者
Glasbrenner, J. [1 ]
Domnick, C. [1 ]
Raschke, M. J. [1 ]
Willinghoefer, T. [1 ]
Kittl, C. [1 ]
Michel, P. [1 ]
Waehnert, D. [1 ]
Herbort, Mirco [1 ]
机构
[1] Westphalian Wilhelms Univ, Dept Trauma Hand & Reconstruct Surg, Waldeyer Str 1, D-48149 Munster, Germany
关键词
Anterior cruciate ligament reconstruction; Cortical fixation; Adjustable button; Cyclic loading; ANTERIOR CRUCIATE LIGAMENT; INTERFERENCE SCREW FIXATION; SOFT-TISSUE GRAFT; SUSPENSION DEVICES; TIBIAL FIXATION; FIXED-LOOP; RECONSTRUCTION; TUNNEL; TRENDS; LENGTH;
D O I
10.1007/s00167-018-5262-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
PurposeDespite good initial pullout strength, it is unclear whether adjustable button (AB) devices for anterior cruciate ligament (ACL) soft-tissue graft fixation, which are based on the Chinese finger trap technique, resist cyclic loading. Furthermore, they have never been tested in a cyclic protocol including complete unloading. It was hypothesized, that the displacement of AB devices with the Chinese finger trap technique would be greater than that of continuous suture loop devices and other available AB mechanisms in a cyclic loading with complete unloading protocol.MethodsACL reconstruction was performed in a porcine knee model using three different types of cortical fixation devices: two different AB devices that use the Chinese finger trap design, one AB device that uses a locked suture loop mechanism and two different continuous loop devices as control groups (n=40). Specimens were mounted in a material-testing machine (Instron Inc.) that permitted 2500 loading and complete unloading cycles to a maximum of 250N, as well as continuous elongation recording. A one-way ANOVA was performed for statistical analysis.ResultsThe displacement of ABs with a Chinese finger trap loop (mean 8.1; SD 1.5mm and mean 6.1; SD 1.4mm) was significantly greater than that of AB with a locked suture loop (mean 4.7; SD 1.0mm; p<0.05) and devices with a continuous loop (mean 4.1; SD 0.5mm and mean 4.4, SD 0.3mm; p<0.01). No significant differences were detected between the ABs with a locked suture loop and the continuous loops.ConclusionCyclic loading and unloading of AB using the Chinese finger trap technique leads to significantly greater construct lengthening compared with other devices. Complete unloading of the ACL is very likely to occur during rehabilitation after ACL reconstruction. Lengthening of the AB device due to cyclic loading might be a potential mode of failure of the ACL graft fixation. Therefore, when using an AB femoral fixation technique, a locked suture loop design or a careful rehabilitation protocol should be considered.
引用
收藏
页码:2530 / 2536
页数:7
相关论文
共 37 条
  • [1] Adjustable buttons for ACL graft cortical fixation partially fail with cyclic loading and unloading
    J. Glasbrenner
    C. Domnick
    M. J. Raschke
    T. Willinghöfer
    C. Kittl
    P. Michel
    D. Wähnert
    Mirco Herbort
    Knee Surgery, Sports Traumatology, Arthroscopy, 2019, 27 : 2530 - 2536
  • [2] ACL graft migration under cyclic loading
    Christian Staerke
    Andreas Möhwald
    Karl-Heinz Gröbel
    Carsten Bochwitz
    Roland Becker
    Knee Surgery, Sports Traumatology, Arthroscopy, 2010, 18 : 1065 - 1070
  • [3] ACL graft migration under cyclic loading
    Staerke, Christian
    Moehwald, Andreas
    Groebel, Karl-Heinz
    Bochwitz, Carsten
    Becker, Roland
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2010, 18 (08) : 1065 - 1070
  • [4] A comparison of three adjustable cortical button ACL fixation devices
    Pasquali, Meghan
    Plante, Matthew J.
    Monchik, Keith O.
    Spenciner, David B.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2017, 25 (05) : 1613 - 1616
  • [5] A comparison of three adjustable cortical button ACL fixation devices
    Meghan Pasquali
    Matthew J. Plante
    Keith O. Monchik
    David B. Spenciner
    Knee Surgery, Sports Traumatology, Arthroscopy, 2017, 25 : 1613 - 1616
  • [6] Arthroscopic ACL Avulsion Fixation With Adjustable Loop Length Cortical Endobutton
    Firoozabadi, Mohammad Ayati
    Dehghaniashkezari, Davood
    Seyedtabaei, Seyed Mohammad Milad
    Irani, Pouya Tabatabaei
    Khaledian, Homayoon
    Savodji, Pantea Bozorgi
    Mortazavi, S. M. Javad
    ARTHROSCOPY TECHNIQUES, 2024, 13 (01):
  • [7] Biomechanical testing of fixed and adjustable femoral cortical suspension devices for ACL reconstruction under high loads and extended cyclic loading
    Sarvpreet Singh
    Sonia Ramos-Pascual
    Kinga Czerbak
    Muzaffar Malik
    Peter J. Schranz
    Anthony W. Miles
    Vipul Mandalia
    Journal of Experimental Orthopaedics, 7
  • [8] Biomechanical testing of fixed and adjustable femoral cortical suspension devices for ACL reconstruction under high loads and extended cyclic loading
    Singh, Sarvpreet
    Ramos-Pascual, Sonia
    Czerbak, Kinga
    Malik, Muzaffar
    Schranz, Peter J.
    Miles, Anthony W.
    Mandalia, Vipul
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2020, 7 (01)
  • [9] Tibial tunnel expansion does not correlate with four-strand graft maturation after ACL reconstruction using adjustable cortical suspensory fixation
    Biset, Alexandre
    Douiri, Adil
    Robinson, James R.
    Laboudie, Pierre
    Colombet, Philippe
    Graveleau, Nicolas
    Bouguennec, Nicolas
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2023, 31 (05) : 1761 - 1770
  • [10] Tibial tunnel expansion does not correlate with four-strand graft maturation after ACL reconstruction using adjustable cortical suspensory fixation
    Alexandre Biset
    Adil Douiri
    James R. Robinson
    Pierre Laboudie
    Philippe Colombet
    Nicolas Graveleau
    Nicolas Bouguennec
    Knee Surgery, Sports Traumatology, Arthroscopy, 2023, 31 : 1761 - 1770