Pulse oximeter plethysmograph waveform and automated oscillometric sphygmomanometer for ankle-brachial index measurement

被引:1
|
作者
Arnold, Cosby G. [1 ]
Walker, J. Richard [2 ]
Metter, E. Jeffrey [3 ]
Young, Shane [2 ]
Brady, Mark F. [2 ]
机构
[1] Univ Colorado, Sch Med, Dept Emergency Med, Denver, CO USA
[2] Univ Tennessee, Hlth Sci Ctr, Dept Emergency Med, Memphis, TN USA
[3] Univ Tennessee, Hlth Sci Ctr, Dept Neurol, Memphis, TN USA
来源
关键词
Ankle-brachial index; Pulse oximetry; Plethysmography; Vascular sufficiency;
D O I
10.1016/j.ajem.2020.10.042
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: There are limited non-invasive methods to assess lower extremity arterial injuries in the emergency department (ED) and pre-hospital setting. The ankle-brachial index (ABI) requires careful auscultation by Doppler, an approach made difficult in noisy environments. We sought to determine the agreement of the ABI measured using the pulse oximeter plethysmograph waveform (Pleth) with auscultation by Doppler in a controlled setting. A secondary outcome sought to examine the agreement of ABI by automated oscillometric sphygmomanometer (AOS) with Doppler. Methods: We measured blood pressure in the right upper and lower extremities of healthy volunteers using: (1) Doppler and manual sphygmomanometer; (2) Pleth and manual sphygmomanometer; and (3) AOS. The Bland-Altman approach to assessing agreement between methods was used comparing mean differences between ABI pairs to their means for Doppler versus Pleth and Doppler versus AOS. The intraclass correlation coefficient (ICC) from mixed effects models examined intraand inter-rater reliability. Results: Among 100 participants with normal ABI the mean ABI (95%CI) were Doppler 1.11 (0.90-1.33), Pleth 1.10 (0.91-1.30), and AOS 1.10 (0.90-1.30). The ABI difference (95% CI for limits of agreement) were 0.01 (-0.20,0.18) for Doppler-Pleth and 0.02 (-0.26, 0.22) for Doppler-AOS. The ICC for the Doppler-Pleth comparison (ICC = 0.56, 95% CI 0.47-0.63) was greater than for the Doppler-AOS (ICC = 0.32, 95% CI 0.19-0.43). Conclusions: The ABI measured using the Pleth has a high level of agreement with measurement by Doppler. The AOS and Doppler have good agreement with greater measurement variability. Pleth and AOS may be reasonable alternatives to Doppler for ABI. (c) 2020 Elsevier Inc. All rights reserved.
引用
收藏
页码:162 / 165
页数:4
相关论文
共 50 条
  • [21] An algorithm to determine the ankle-brachial pressure index using the oscillometric blood pressure measurement method
    Meža, Marko
    Šušterič, Jakob
    Krivic, Tomo
    Tasič, Jurij F.
    Elektrotehniski Vestnik/Electrotechnical Review, 2011, 78 (03): : 136 - 141
  • [22] Validation study of automated oscillometric measurement of the ankle-brachial index for lower arterial occlusive disease by comparison with computed tomography angiography
    Ichihashi, Shigeo
    Hashimoto, Tomoko
    Iwakoshi, Shinichi
    Kichikawa, Kimihiko
    HYPERTENSION RESEARCH, 2014, 37 (06) : 591 - 594
  • [23] Pitfalls in the ankle-brachial index and brachial-ankle pulse wave velocity
    Ato, Dai
    VASCULAR HEALTH AND RISK MANAGEMENT, 2018, 14 : 41 - 62
  • [24] Validation study of automated oscillometric measurement of the ankle-brachial index for lower arterial occlusive disease by comparison with computed tomography angiography
    Shigeo Ichihashi
    Tomoko Hashimoto
    Shinichi Iwakoshi
    Kimihiko Kichikawa
    Hypertension Research, 2014, 37 : 591 - 594
  • [25] Comparison of the automated oscillometric method with Doppler ultrasound method to access the Ankle-Brachial Pressure Index (ABPI)
    Hamel, J. -F.
    Tanguy, M.
    Foucaud, D.
    Fanello, S.
    JOURNAL DES MALADIES VASCULAIRES, 2010, 35 (03) : 169 - 174
  • [26] Oscillometric measurement of the ankle-brachial index and the estimated carotid-femoral pulse wave velocity improves the sensitivity of an automated device in screening peripheral artery disease
    Fendrik, Krisztina
    Biro, Katalin
    Endrei, Dora
    Koltai, Katalin
    Sandor, Barbara
    Toth, Kalman
    Kesmarky, Gabor
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
  • [27] Teaching and evaluating measurement of the ankle-brachial index
    Aboyans, Victor
    Lacroix, Philippe
    VASCULAR MEDICINE, 2018, 23 (02) : 114 - 115
  • [28] Comparison of Ankle-Brachial Index Measured by an Automated Oscillometric Apparatus with that by Standard Doppler Technique in Vascular Patients
    Korno, M.
    Eldrup, N.
    Sillesen, H.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2009, 38 (05) : 610 - 615
  • [29] Automated plethysmographic measurement of the ankle-brachial index: a comparison with the doppler ultrasound method
    Davies, Jane H.
    Williams, E. Mark
    HYPERTENSION RESEARCH, 2016, 39 (02) : 100 - 106
  • [30] Automated plethysmographic measurement of the ankle-brachial index: a comparison with the doppler ultrasound method
    Jane H Davies
    E Mark Williams
    Hypertension Research, 2016, 39 : 100 - 106