This paper describes the advances in the development and subsequent testing of an imaging device for three-dimensional ultrasound measurement of atherosclerotic plaque in the carotid artery. The embolization from the atherosclerotic carotid plaque is one of the most common causes of ischemic stroke and, therefore, we consider the measurement of the plaque as extremely important. The paper describes the proposed hardware for enhancing the standard ultrasonic probe to provide a possibility of accurate probe positioning and synchronization with the cardiac activity, allowing the precise plaque measurements that were impossible with the standard equipment. The synchronization signal is derived from the output signal of the patient monitor (electrocardiogram (ECG)), processed by a microcontroller-based system, generating the control commands for the linear motion moving the probe. The controlling algorithm synchronizes the movement with the ECG waveform to obtain clear images not disturbed by the heart activity.
机构:
AtheroPoint, Monitoring & Diagnost Div, Roseville, CA USAUniv Cagliari, AOU Cagliari, Dept Med Imaging, Cagliari, Italy
Suri, Jasjit S.
Wasserman, Bruce A.
论文数: 0引用数: 0
h-index: 0
机构:
Johns Hopkins Bloomberg Sch Publ Hlth, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD USAUniv Cagliari, AOU Cagliari, Dept Med Imaging, Cagliari, Italy
Wasserman, Bruce A.
Anzalone, Nicoletta
论文数: 0引用数: 0
h-index: 0
机构:
Ist Sci San Raffaele, Neuroradiol Unit, Via Olgettina 60, I-20132 Milan, Italy
Ist Sci San Raffaele, CERMAC, Via Olgettina 60, I-20132 Milan, Italy
Univ Vita Salute San Raffaele, Via Olgettina 60, I-20132 Milan, ItalyUniv Cagliari, AOU Cagliari, Dept Med Imaging, Cagliari, Italy