Evaluation of a total reflection X-ray fluorescence spectrometer in the determination of arsenic and trace metals in environmental samples

被引:12
|
作者
Juvonen, Riitta [1 ]
Parviainen, Annika [1 ]
Loukola-Ruskeeniemi, Kirsti [1 ]
机构
[1] Helsinki Univ Technol, Dept Civil & Environm Engn, Geoenvironm Technol Sect, FI-02015 Helsinki, Finland
关键词
TXRF; total reflection X-ray fluorescence; arsenic; trace metals; mine tailings; surface water; acid mine drainage; ELEMENT DETERMINATION;
D O I
10.1144/1467-7873/09-205
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
Suitability of the PicoTAX total reflection X-ray fluorescence (TXRF) spectrometer was Studied for determination of As and trace metals in the Haveri Au-Cu sulphide mine tailings leached with aqua regia and in water affected by acid mine drainage. Results for 12 water and 16 tailings samples were compared with accredited methods, inductively coupled plasma-atomic emission spectrometry (ICP-AES) and inductively Coupled plasma-mass spectrometry (ICP-MS). Certified reference samples analysed were QC Metal LL from Eurofins A/S, and groundwater CRM 609 and lake sediment BCR-280R from the Institute for Reference Materials and Measurements (IRMM) of the European Commission. Approximate method detection limits (MDL) for water samples, based on reference sample analyses, are in the range of 20 mu g l(-1) for As, Cr, Pb and Zn and 5-10 mu g l(-1) for Co, Cu, Ni, V and Mn. MDLs for aqua regia leach, calculated using reagent blank values, were <= 20 (mg kg(-1)) for As, Co, Cr, Cu, Mn, Ni, V, and Zn. Random error of determination varied from 12 to 20%. Inter-element interferences include: As and Pb on each other, and Fe on Cr and Co. Using the Mo anode of the X-ray tube excludes the determination of Cd, Mo, Nb, Tc and Zr. TXRF can be recommended merely as a preliminary technique for selecting samples.
引用
收藏
页码:173 / 178
页数:6
相关论文
共 50 条
  • [31] Determination of trace elements in uranium oxide by Total Reflection X-ray Fluorescence spectrometry
    Misra, NL
    Mudher, KDS
    Adya, VC
    Rajeswari, B
    Venugopal, V
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (06) : 834 - 840
  • [32] Separation and enrichment of palladium and gold in biological and environmental samples, adapted to the determination by total reflection X-ray fluorescence
    Messerschmidt, J
    von Bohlen, A
    Alt, F
    Klockenkämper, R
    [J]. ANALYST, 2000, 125 (03) : 397 - 399
  • [33] Intelligent Simultaneous Quantitative Online Analysis of Environmental Trace Heavy Metals with Total-Reflection X-Ray Fluorescence
    Ma, Junjie
    Wang, Yeyao
    Yang, Qi
    Liu, Yubing
    Shi, Ping
    [J]. SENSORS, 2015, 15 (05): : 10650 - 10675
  • [34] Rapid Screening of Heavy Metals and Trace Elements in Environmental Samples Using Portable X-Ray Fluorescence Spectrometer, A Comparative Study
    Jacqueline Q. McComb
    Christian Rogers
    Fengxiang X. Han
    Paul B. Tchounwou
    [J]. Water, Air, & Soil Pollution, 2014, 225
  • [35] Element determination in medieval soil samples by total reflection X-ray fluorescence analysis
    von Bohlen, A
    Brink-Kloke, H
    Althoff, C
    [J]. ANALYTICA CHIMICA ACTA, 2003, 480 (02) : 327 - 335
  • [36] Rapid Screening of Heavy Metals and Trace Elements in Environmental Samples Using Portable X-Ray Fluorescence Spectrometer, A Comparative Study
    McComb, Jacqueline Q.
    Rogers, Christian
    Han, Fengxiang X.
    Tchounwou, Paul B.
    [J]. WATER AIR AND SOIL POLLUTION, 2014, 225 (12):
  • [37] Total reflection x-ray fluorescence spectrometer with parallel primary beam
    Korotkikh, EM
    [J]. X-RAY SPECTROMETRY, 2006, 35 (02) : 116 - 119
  • [38] Matrix effect on the detection limit and accuracy in total reflection X-ray fluorescence analysis of trace elements in environmental and biological samples
    Karjou, J.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2007, 62 (02) : 177 - 181
  • [39] Determination of trace and minor elements in Slovenian honey by total reflection X-ray fluorescence spectroscopy
    Golob, T
    Dobersek, U
    Kump, P
    Necemer, M
    [J]. FOOD CHEMISTRY, 2005, 91 (04) : 593 - 600
  • [40] Trace element determination in diesel particulates by total-reflection X-ray fluorescence analysis
    Vilhunen, JK
    von Bohlen, A
    Schmeling, M
    Rantanen, L
    Mikkonen, S
    Klockenkämper, R
    Klockow, D
    [J]. MIKROCHIMICA ACTA, 1999, 131 (3-4) : 219 - 223