NUMERICAL ANALYSIS ON THE INFLUENCE OF TAPER JUNCTION CORROSION PROFILE WITH EMPHASIS ON TI-ALLOY MODULAR HIP ARTHROPLASTY

被引:0
|
作者
Novotny, Adam [1 ]
Paliwal, Manish [1 ]
机构
[1] Coll New Jersey, Dept Mech Engn, Ewing, NJ 08628 USA
关键词
RETRIEVAL;
D O I
10.1115/IMECE2017-71907
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Modular designs give orthopaedic surgeons a greater flexibility to custom fit the implant to the patients bone while performing total hip arthroplasty. Titanium alloy (Ti6Al4V (ASTM F-136)) is typically used for modular hip implant stems. This highly corrosion resistant alloy forms passive surface oxide films spontaneously. However, with modular designs, micro-motion may occur at the taper junctions during mechanical loading. Crevices between the taper junctions may allow the body fluids to enter and remain stagnant. These conditions make the modular tapers susceptible to fatigue and mechanically assisted crevice corrosion. The in vivo degradation of metal alloy implants compromises the structural integrity. The influence of stress corrosion induced pits of a titanium-alloy modular implant in cement less total hip arthroplasty was numerically investigated. The effect of pit geometry parameters - cylindrical, conical, and hemispherical dimple are compared and discussed in terms of taper performance.
引用
收藏
页数:6
相关论文
共 37 条
  • [31] Model-Based Roentgen Stereophotogrammetric Analysis to Monitor the Head-Taper Junction in Total Hip Arthroplasty in Vivo-And They Do Move
    Xu, Jing
    Sonntag, Robert
    Kretzer, J. Philippe
    Taylor, Dominic
    Forst, Raimund
    Seehaus, Frank
    MATERIALS, 2020, 13 (07)
  • [32] Microgrooved Surface Topography Does Not Influence Fretting Corrosion of Tapers in Total Hip Arthroplasty: Classification and Retrieval Analysis
    Arnholt, Christina
    Underwood, Richard
    MacDonald, Daniel W.
    Higgs, Genymphas B.
    Chen, Antonia F.
    Klein, Gregg
    Hamlin, Brain
    Lee, Gwo-Chin
    Mont, Michael
    Cates, Harlod
    Malkani, Arthur
    Kraay, Matthew
    Rimnac, Clare
    Kurtz, Steven M.
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 99 - 112
  • [33] Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material
    Grupp, Thomas M.
    Weik, Thomas
    Bloemer, Wilhelm
    Knaebel, Hanns-Peter
    BMC MUSCULOSKELETAL DISORDERS, 2010, 11
  • [34] Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material
    Thomas M Grupp
    Thomas Weik
    Wilhelm Bloemer
    Hanns-Peter Knaebel
    BMC Musculoskeletal Disorders, 11
  • [35] Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
    Martelli, Andrea
    Erani, Paolo
    Pazzagli, Nicola
    Cannillo, Valeria
    Baleani, Massimiliano
    MATERIALS, 2023, 16 (03)
  • [36] Experimental and numerical analysis of tensile properties of Ti-13Nb-13Zr alloy and determination of influence of anodization process
    Barjaktarevic, Dragana
    Medjo, Bojan
    Gubeljak, Nenad
    Cvijovic-Alagic, Ivana
    Stefane, Primoz
    Djokic, Veljko
    Rakin, Marko
    1ST VIRTUAL EUROPEAN CONFERENCE ON FRACTURE - VECF1, 2020, 28 : 2187 - 2194
  • [37] A Finite Element Analysis Study of Influence of Femoral Stem Material in Stress Shielding in a Model of Uncemented Total Hip Arthroplasty: Ti-6Al-4V versus Carbon Fibre-Reinforced PEEK Composite
    Ceddia, Mario
    Solarino, Giuseppe
    Giannini, Giorgio
    De Giosa, Giuseppe
    Tucci, Maria
    Trentadue, Bartolomeo
    JOURNAL OF COMPOSITES SCIENCE, 2024, 8 (07):