NUMERICAL ANALYSIS ON THE INFLUENCE OF TAPER JUNCTION CORROSION PROFILE WITH EMPHASIS ON TI-ALLOY MODULAR HIP ARTHROPLASTY

被引:0
|
作者
Novotny, Adam [1 ]
Paliwal, Manish [1 ]
机构
[1] Coll New Jersey, Dept Mech Engn, Ewing, NJ 08628 USA
关键词
RETRIEVAL;
D O I
10.1115/IMECE2017-71907
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Modular designs give orthopaedic surgeons a greater flexibility to custom fit the implant to the patients bone while performing total hip arthroplasty. Titanium alloy (Ti6Al4V (ASTM F-136)) is typically used for modular hip implant stems. This highly corrosion resistant alloy forms passive surface oxide films spontaneously. However, with modular designs, micro-motion may occur at the taper junctions during mechanical loading. Crevices between the taper junctions may allow the body fluids to enter and remain stagnant. These conditions make the modular tapers susceptible to fatigue and mechanically assisted crevice corrosion. The in vivo degradation of metal alloy implants compromises the structural integrity. The influence of stress corrosion induced pits of a titanium-alloy modular implant in cement less total hip arthroplasty was numerically investigated. The effect of pit geometry parameters - cylindrical, conical, and hemispherical dimple are compared and discussed in terms of taper performance.
引用
收藏
页数:6
相关论文
共 37 条
  • [1] NUMERICAL ANALYSIS ON STRESS INDUCED CORROSION WITH EMPHASIS ON TI-ALLOY TOTAL HIP ARTHROPLASTY STEM FAILURE
    Paliwal, Manish
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2013, VOL 3A, 2014,
  • [2] Modular Taper Junction Corrosion and Failure: How to Approach a Recalled Total Hip Arthroplasty Implant
    Pivec, Robert
    Meneghini, R. Michael
    Hozack, William J.
    Westrich, Geoffrey H.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (01): : 1 - 6
  • [3] Long-term survivorship of an exchangeable-neck hip prosthesis with a Ti-alloy/Ti-alloy neck–stem junction
    Massimiliano Baleani
    Aldo Toni
    Cristina Ancarani
    Susanna Stea
    Barbara Bordini
    Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 3649 - 3657
  • [4] Failure analysis of two Ti-alloy total hip arthroplasty femoral stems fractured in vivo
    Magnissalis, EA
    Zinelis, S
    Karachalios, T
    Hartofilakidis, G
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 66B (01) : 299 - 305
  • [5] Letter to the Editor: In Response to "Modular Taper Junction Corrosion and Failure: How to Approach a Recalled Total Hip Arthroplasty Implant"
    Davies, Joseph F.
    Leone, William A.
    Nessler, Joseph P.
    Perona, Paul G.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (02): : 449 - 450
  • [6] Long-term survivorship of an exchangeable-neck hip prosthesis with a Ti-alloy/Ti-alloy neck-stem junction
    Baleani, Massimiliano
    Toni, Aldo
    Ancarani, Cristina
    Stea, Susanna
    Bordini, Barbara
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (06) : 3649 - 3657
  • [7] Letter to the Editor: In Response to "Modular Taper Junction Corrosion and Failure: How to Approach a Recalled Total Hip Arthroplasty Implant" Reply
    Pivec, Robert
    Meneghini, R. Michael
    Hozack, William J.
    Westrich, Geoffrey H.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (02): : 450 - 451
  • [8] Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases
    Kop, Alan M.
    Swarts, Eric
    JOURNAL OF ARTHROPLASTY, 2009, 24 (07): : 1019 - 1023
  • [9] The morse taper junction in modular revision hip replacement - A biomechanical and retrieval analysis
    Schramm, M
    Wirtz, DC
    Holzwarth, U
    Pitto, RP
    BIOMEDIZINISCHE TECHNIK, 2000, 45 (04): : 105 - 109
  • [10] The Influence of Contamination and Cleaning on the Strength of Modular Head Taper Fixation in Total Hip Arthroplasty
    Krull, Annika
    Morlock, Michael M.
    Bishop, Nicholas E.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (10): : 3200 - 3205