Taper fretting behavior of PEEK artificial hip joint

被引:18
|
作者
Xu, Handong [1 ]
Zhang, Dekun [2 ]
Chen, Kai [2 ]
Zhang, Tao [1 ]
机构
[1] China Univ Min & Technol, Sch Mechatron Engn, Xuzhou 221116, Jiangsu, Peoples R China
[2] China Univ Min & Technol, Sch Mat Sci & Engn, Xuzhou 221116, Jiangsu, Peoples R China
关键词
PEEK; Ti6Al4V; Wear mechanism; Fretting behavior; FINITE-ELEMENT; WEAR; ALLOY; REPLACEMENTS; CORROSION; IMPLANTS; JUNCTION; DEBRIS;
D O I
10.1016/j.triboint.2019.04.027
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
To study the fretting tribological behavior of a PEEK hip prosthesis, Ti6Al4V was used as the stem taper material, PEEK was used as the head material, and CoCrMo was used as the contrast material for the PEEK. Fretting friction and wear experiments between the stem taper and head were carried out to demonstrate the feasibility of PEEK as a metal substitute. The results show that the wear values of the Ti6Al4V/PEEK and Ti6Al4V/CoCrMo pairs were all of a small order of magnitude. The Ti6Al4V/PEEK pairs had a better corrosion resistance than the Ti6Al4V/CoCrMo pairs. Meanwhile, the fretting characteristics and damage mechanism of the Ti6Al4V/PEEK pairs were obtained.
引用
收藏
页码:30 / 38
页数:9
相关论文
共 50 条
  • [1] Reducing Taper Fretting Corrosion in Artificial Hip Joints Using a PEEK Femoral Head
    Zhang, Dekun
    Zhang, Yueru
    Chen, Kai
    Zhang, Xinyue
    Xu, Handong
    JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2021, 30 (06) : 4619 - 4628
  • [2] Reducing Taper Fretting Corrosion in Artificial Hip Joints Using a PEEK Femoral Head
    Dekun Zhang
    Yueru Zhang
    Kai Chen
    Xinyue Zhang
    Handong Xu
    Journal of Materials Engineering and Performance, 2021, 30 : 4619 - 4628
  • [3] Ion release during fretting at taper joint interface of hip joint prosthesis
    Kalbarczyk, M.
    Liskiewicz, T.
    Yan, Y.
    Neville, A.
    TRIBOLOGY-MATERIALS SURFACES & INTERFACES, 2011, 5 (03) : 107 - 113
  • [4] A comparative study on fretting wear and frictional heating behavior of PEEK composites for artificial joint applications
    Shen, Fei
    Ke, Liao-Liang
    POLYMER TESTING, 2022, 109
  • [5] A comparative study on fretting wear and frictional heating behavior of PEEK composites for artificial joint applications
    Shen, Fei
    Ke, Liao-Liang
    Polymer Testing, 2022, 109
  • [6] Research on the torsional fretting behavior of the head-neck interface of artificial hip joint
    Chen, Kai
    Zhang, Dekun
    Zhang, Gaofeng
    Hao, Tianqing
    Ge, Shirong
    MATERIALS & DESIGN, 2014, 56 : 914 - 922
  • [7] Design of Artificial Hip Joint by Carbon/PEEK Composites
    Kim, Yun-Hae
    Yoon, Sung-Won
    Jung, Min-Kyo
    Ha, Jin-Cheol
    Murakami, Ri-ichi
    ADVANCED TECHNOLOGIES IN MANUFACTURING, ENGINEERING AND MATERIALS, PTS 1-3, 2013, 774-776 : 1336 - +
  • [8] Evaluation of the fretting corrosion behavior of the proximal pad taper of a modular hip design
    Chandrasekaran, V
    Sauer, WL
    Taylor, AM
    Hoeppner, DW
    WEAR, 1999, 231 (01) : 54 - 64
  • [9] Elastohydrodynamic lubrication analysis of polymer-on-polymer artificial hip joint of CF/PEEK composite
    Gu, Junfeng
    Sun, Xiujie
    Li, Zheng
    Ruan, Shilun
    Shen, Changyu
    JOURNAL OF APPLIED POLYMER SCIENCE, 2021, 138 (39)
  • [10] Influence of the taper on the fretting wear of the femoral stem-femoral head taper junction in total hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    INDUSTRIAL LUBRICATION AND TRIBOLOGY, 2022, 74 (04) : 385 - 391