Design and Evaluation of a Low-Cost Smartphone Pulse Oximeter

被引:49
|
作者
Petersen, Christian L. [1 ]
Chen, Tso P. [1 ]
Ansermino, J. Mark [1 ]
Dumont, Guy A. [2 ]
机构
[1] Univ British Columbia, Dept Anesthesiol Pharmacol & Therapeut, Vancouver, BC V6T 1Z3, Canada
[2] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V6T 1Z4, Canada
来源
SENSORS | 2013年 / 13卷 / 12期
关键词
CHILDREN; ANESTHESIA; PNEUMONIA; HYPOXEMIA; MORTALITY; DEATHS;
D O I
10.3390/s131216882
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Infectious diseases such as pneumonia take the lives of millions of children in low- and middle-income countries every year. Many of these deaths could be prevented with the availability of robust and low-cost diagnostic tools using integrated sensor technology. Pulse oximetry in particular, offers a unique non-invasive and specific test for an increase in the severity of many infectious diseases such as pneumonia. If pulse oximetry could be delivered on widely available mobile phones, it could become a compelling solution to global health challenges. Many lives could be saved if this technology was disseminated effectively in the affected regions of the world to rescue patients from the fatal consequences of these infectious diseases. We describe the implementation of such an oximeter that interfaces a conventional clinical oximeter finger sensor with a smartphone through the headset jack audio interface, and present a simulator-based systematic verification system to be used for automated validation of the sensor interface on different smartphones and media players. An excellent agreement was found between the simulator and the audio oximeter for both oxygen saturation and heart rate over a wide range of optical transmission levels on 4th and 5th generations of the iPod Touch™ and iPhone™ devices. © 2013 by the authors; licensee MDPI, Basel, Switzerland.
引用
收藏
页码:16882 / 16893
页数:12
相关论文
共 50 条
  • [41] Why smartphone is a low-cost alternative in whole slide imaging?
    Mondal, Himel
    Mondal, Shaikat
    [J]. JOURNAL OF FAMILY MEDICINE AND PRIMARY CARE, 2020, 9 (10) : 5391 - 5391
  • [42] A Low-Cost Dual-Beam Smartphone Visible Spectrometer
    Jarujareet, Ungkarn
    Pichayawaytin, Grit
    Sripetch, Pongsakun
    Doljirapisit, Narusorn
    Sumriddetchkajorn, Sarun
    Prempree, Panintorn
    Chaitavon, Kosom
    Punpetch, Pongpun
    Amarit, Ratthasart
    [J]. JOURNAL OF CHEMICAL EDUCATION, 2023, : 546 - 553
  • [43] A Compact, Portable, and Low-cost Potentiostat with Smartphone for Electrochemical Sensors
    Permwong, Watcharin
    Thunkhamrak, Chidkamon
    Upan, Jantima
    Jakmunee, Jaroon
    Pencharee, Somkid
    [J]. CHIANG MAI JOURNAL OF SCIENCE, 2020, 47 (06): : 1183 - 1194
  • [44] Design and evaluation of a low-cost instrumented glove for hand function assessment
    Ninja P Oess
    Johann Wanek
    Armin Curt
    [J]. Journal of NeuroEngineering and Rehabilitation, 9
  • [45] Design and evaluation of a low-cost instrumented glove for hand function assessment
    Oess, Ninja P.
    Wanek, Johann
    Curt, Armin
    [J]. JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2012, 9
  • [46] A Low-Cost Entangling Drifter for Sargassum Tracking: Design and Experimental Evaluation
    Pixa, Chase
    Manganini, Kevin
    Myers, Alyson
    Michel, Anna P. M.
    [J]. OCEANS 2023 - LIMERICK, 2023,
  • [47] DESIGN, PROTOTYPE, AND EVALUATION OF A LOW-COST MULTIMODAL DEVICE FOR CARDIOVASCULAR MONITORING
    Ruckman, Sophia
    Bhatt, Jigar
    Cook, Jadyn
    Gamage, Peshala Thibbotuwawa
    Kakavand, Bahram
    Taebi, Amirtaha
    [J]. PROCEEDINGS OF ASME 2023 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, IMECE2023, VOL 5, 2023,
  • [48] Design and evaluation of a low-cost, high-speed signal amplifier
    Christian, NP
    Alexander, AW
    Reilly, JP
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01): : 243 - 246
  • [49] Design and validation of a pulse oximeter calibrator
    Hornberger, C
    Matz, H
    Konecny, E
    Frankenberger, H
    Bonk, R
    Avgerinos, J
    Benekos, K
    Valais, J
    Ikiades, A
    Gil-Rodriguez, J
    Wouters, P
    Meyfroidt, G
    Ponz, L
    Gehring, H
    [J]. ANESTHESIA AND ANALGESIA, 2002, 94 (01): : S8 - S12
  • [50] LOW-COST, COMMUNAL GRAZING DESIGN FOR PRELIMINARY EVALUATION OF GRAZING SYSTEMS
    MICHALK, DL
    MCFARLANE, JD
    [J]. JOURNAL OF THE BRITISH GRASSLAND SOCIETY, 1978, 33 (04): : 301 - &