Structure-dependent amplification for denoising and background correction in Fourier ptychographic microscopy

被引:9
|
作者
Claveau, Remy [1 ]
Manescu, Petru [1 ]
Fernandez-Reyes, Delmiro [1 ,2 ]
Shaw, Michael [1 ,3 ]
机构
[1] UCL, Fac Engn Sci, Dept Comp Sci, London WC1E 6BT, England
[2] Univ Ibadan, Dept Paediat, Coll Med, Ibadan, Nigeria
[3] Natl Phys Lab, Biometrol Grp, Teddington TW11 0LW, Middx, England
基金
英国惠康基金;
关键词
23;
D O I
10.1364/OE.403780
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Fourier Ptychographic Microscopy (FPM) allows high resolution imaging using iterative phase retrieval to recover an estimate of the complex object from a series of images captured under oblique illumination. FPM is particularly sensitive to noise and uncorrected background signals as it relies on combining information from brightfield and noisy darkfield (DF) images. In this article we consider the impact of different noise sources in FPM and show that inadequate removal of the DF background signal and associated noise are the predominant cause of artefacts in reconstructed images. We propose a simple solution to FPM background correction and denoising that outperforms existing methods in terms of image quality, speed and simplicity, whilst maintaining high spatial resolution and sharpness of the reconstructed image. Our method takes advantage of the data redundancy in real space within the acquired dataset to boost the signal-to-background ratio in the captured DF images, before optimally suppressing background signal. By incorporating differentially denoised images within the classic FPM iterative phase retrieval algorithm, we show that it is possible to achieve efficient removal of background artefacts without suppression of high frequency information. The method is tested using simulated data and experimental images of thin blood films, bone marrow and liver tissue sections. Our approach is non-parametric, requires no prior knowledge of the noise distribution and can be directly applied to other hardware platforms and reconstruction algorithms making it widely applicable in FPM. Published by The Optical Society under the terms of the Creative Commons Attribution 4.0 License.
引用
收藏
页码:35438 / 35453
页数:16
相关论文
共 38 条
  • [31] Structure-dependent behaviours of skin layers studied by atomic force microscopy
    Chang, A. C.
    Liu, B. H.
    Shao, P. L. .
    Liao, J. D.
    JOURNAL OF MICROSCOPY, 2017, 267 (03) : 265 - 271
  • [32] A positional misalignment correction method for Fourier ptychographic microscopy based on the quasi-Newton method with a global optimization module
    Zhang, Jinlei
    Tao, Xiao
    Sun, Peng
    Zheng, Zhenrong
    OPTICS COMMUNICATIONS, 2019, 452 : 296 - 305
  • [33] Multi-scale fractal Fourier Ptychographic microscopy to assess the dose-dependent impact of copper pollution on living diatoms
    Bianco, Vittorio
    Miccio, Lisa
    Pirone, Daniele
    Cavalletti, Elena
    Behal, Jaromir
    Memmolo, Pasquale
    Sardo, Angela
    Ferraro, Pietro
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [34] Wavelet-based denoising of the Fourier Metric in real-time wavefront correction for Single Molecule Localization microscopy
    Tehrani, Kayvan Forouhesh
    Mortensen, Luke J.
    Kner, Peter
    ADAPTIVE OPTICS AND WAVEFRONT CONTROL FOR BIOLOGICAL SYSTEMS II, 2016, 9717
  • [35] Diatoms as bio-sentinels to probe the dose-dependent impact of copper on aquatic environment: a multi-scale fractal analysis in Fourier Ptychographic Microscopy
    Bianco, Vittorio
    Pirone, Daniele
    Miccio, Lisa
    Cavalletti, Elena
    Behal, Jaromir
    Memmolo, Pasquale
    Sardo, Angela
    Ferraro, Pietro
    2023 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR THE SEA; LEARNING TO MEASURE SEA HEALTH PARAMETERS, METROSEA, 2023, : 243 - 248
  • [36] Author Correction: Discovery of toxoflavin, a potent IRE1α inhibitor acting through structure-dependent oxidative inhibition
    Kai-long Jiang
    Chang-mei Liu
    Li-tong Nie
    Hai-ni Jiang
    Lei Xu
    Kun-zhi Zhang
    Li-xia Fan
    An-hui Gao
    Lu-lin Lin
    Xiang-yu Wang
    Min-jia Tan
    Qi-qing Zhang
    Yu-bo Zhou
    Jia Li
    Acta Pharmacologica Sinica, 2023, 44 : 2550 - 2550
  • [37] Simultaneous fluorescence and high-resolution bright-field imaging with aberration correction over a wide field-of-view with Fourier ptychographic microscopy (FPM) (Conference Presentation)
    Chung, Jaebum
    Kim, Jinho
    Ou, Xiaoze
    Horstmeyer, Roarke
    Yang, Changhuei
    THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XXIII, 2016, 9713
  • [38] REPRODUCIBILITY OF PEAK FILLING RATE MEASUREMENTS EMPLOYING FRAME MODE ACQUISITION WITH CYCLE-DEPENDENT BACKGROUND CORRECTION AND FOURIER-ANALYSIS
    SCHWARTZ, RG
    BENHORIN, J
    HOPKINS, K
    WILSON, GA
    WEBER, DA
    OMARA, RE
    RICHESON, JF
    JOURNAL OF NUCLEAR MEDICINE, 1987, 28 (06) : 1070 - 1070