Glucose pump test: a new method for blood flow measurements

被引:19
|
作者
Magnasco, A
Alloatti, S
Martinoli, C
Solari, P
机构
[1] Sestri Levante Hosp, Nephrol Dialysis Unit, I-16039 Genoa, Italy
[2] Univ Genoa, Dept Radiol, Aosta Hosp, Genoa, Italy
关键词
constant infusion; glucose; vascular access;
D O I
10.1093/ndt/17.12.2244
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
Background. A good test for monitoring blood flow (Q(a)) must be accurate, rapid and economical in order to allow frequent easy measurements. The glucose pump test (GPT) is based on a constant glucose infusion as a dilutional indicator of Q(a). Methods. GPT protocol requires a constant glucose infusion, by a syringe pump, into the arterial needle and two blood withdrawals from the venous needle, one basal before the infusion (C-a1), the other (C-a2) 11 s after the start of the infusion. At the bedside we measure glucose on C-a1 and C-a2. Knowing the infused glucose concentration (C) and the pump infusion rate (Q(i)) we can easily calculate Q(a) = Q(i) x (C-i-C-a2)/ (C-a2-C-a1). We verified the accuracy of this new method by comparing it with the in vitro results from a circuit reproducing vascular access circulation, and in vivo comparing GPT-Q(a) with Doppler ultrasound in pre-dialysis to the Transonic HD01-Q(a) during dialysis in 23 chronic haemodialysis patients. Results. GPT-Qa values were highly correlated with the in vitro Q(a)=1.01xGPT-Q(a)-16.6; r = 0.94. There was agreement between the mean flow values of GPT and Doppler (927.5 and 927.1 ml/min, respectively; P = NS) while the mean value of HD01 was significantly lower (HD01-Q(a) = 690 ml/min; P < 0.001 vs GPT-Q(a) and Doppler-Q(a)). The regression analysis showed a good correlation between GPT and Transonic results (r = 0.95; HD01-Q(a) = 0.86 x GPT-Q(a)-111.9), while there was a significant difference between the two measurements (mean Delta 235 +/- 117 ml/min; range from 15 to 451 ml/min). This difference could be caused by the large haemodynamic variations (different blood pressure, cardiac output, circulating effective volume, haematocrit) between pre-dialysis and intra-dialysis and in addition by the counter current flow during the reversal blood lines Transonic measurements. Conclusions. GPT offers the advantage of a simple bedside procedure easily performed before dialysis: it does not interfere with the dialysis treatment and it is less intrusive for the patient as it does not involve reversal of the blood lines. The preliminary data indicate that our method could be a useful, simple and cheap test for monitoring access flow in every dialysis unit.
引用
收藏
页码:2244 / 2248
页数:5
相关论文
共 50 条
  • [11] Performance of a New Blood Glucose Test Strip
    Bailey, Timothy
    Chu, Amy
    Parkes, Joan Lee
    Wallace, Jane
    Pardo, Scott
    Schachner, Holly C.
    Castro, Rita
    Simmons, David A.
    [J]. DIABETES, 2011, 60 : A246 - A246
  • [12] NEW BLOOD PUMPING METHOD - GRAVITY PUMP
    MOTTAGHY, K
    SCHUMACHER, E
    SCHMIDSCHONBEIN, H
    [J]. ARTIFICIAL ORGANS, 1977, 1 (02) : 127 - 127
  • [13] New near-infrared spectroscopy method for local measurements of cerebral blood flow
    Pham, T.
    Sassaroli, A.
    Blaney, G.
    Fantini, S.
    [J]. DIFFUSE OPTICAL SPECTROSCOPY AND IMAGING VII, 2019, 11074
  • [14] Reversed glucose pump test (R-GPT) compared to Transonic device (HD01) in access flow measurements.
    Magnasco, A
    Bacchini, G
    Lamilia, V
    Solari, P
    Locatelli, F
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2002, 13 : 233A - 233A
  • [15] A New Magnetic Sealless Coupling Axial Flow Blood Pump
    LIN Chang-yan
    [J]. Chinese Journal of Biomedical Engineering, 2003, (03) : 133 - 138
  • [16] A new method of measuring blood flow
    Gibbs, OS
    [J]. JOURNAL OF PHARMACOLOGY AND EXPERIMENTAL THERAPEUTICS, 1928, 34 (03): : 293 - 298
  • [17] Study on a new method of on-line test of screw pump
    Wang Zhaoxia
    Xi Jinlu
    Li Jing
    Xu Xiaoli
    [J]. PROCEEDINGS OF THE FIFTH INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1 AND 2, 2014, : 109 - 112
  • [18] NEW RADIOLOGICAL METHOD FOR BLOOD FLOW MEASUREMENTS THROUGH INDIVIDUAL VESSELS IN INTACT CONSCIOUS MAN
    RUTISHAU.W
    SIMON, HJ
    SCHAD, N
    STUCKY, JP
    WELLAUER, J
    [J]. CIRCULATION, 1967, 36 (4S2) : I228 - &
  • [19] PATCH TEST RESPONSES EVALUATED BY CUTANEOUS BLOOD-FLOW MEASUREMENTS
    STABERG, B
    KLEMP, P
    SERUP, J
    [J]. ARCHIVES OF DERMATOLOGY, 1984, 120 (06) : 741 - 743
  • [20] BLOOD-GLUCOSE MEASUREMENTS
    KING, AB
    MCCARTER, L
    DAVIS, J
    [J]. DIABETES CARE, 1982, 5 (02) : 144 - 144