On an Aerodynamic Mechanism to Enhance Ion Transmission and Sensitivity of FAIMS for Nano-Electrospray Ionization-Mass Spectrometry

被引:45
|
作者
Prasad, Satendra [1 ]
Belford, Michael W. [1 ]
Dunyach, Jean-Jacques [1 ]
Purves, Randy W. [2 ]
机构
[1] Thermo Fisher Sci, San Jose, CA 95134 USA
[2] Univ Saskatchewan, Saskatoon, SK S7N 5A8, Canada
关键词
Sensitivity; FAIMS; Nano-ESI-FAIMS-MS; SIMION; CFD; PEPTIDE IDENTIFICATION; MS; FLOW; SELECTIVITY; SIMULATION; OPERATION; SPECTRA; MS/MS;
D O I
10.1007/s13361-014-0995-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Simulations show that significant ion losses occur within the commercial electrospray ionization-field asymmetric waveform ion mobility spectrometer (ESI-FAIMS) interface owing to an angular desolvation gas flow and because of the impact of the FAIMS carrier gas onto the inner rf (radio frequency) electrode. The angular desolvation gas flow diverts ions away from the entrance plate orifice while the carrier gas annihilates ions onto the inner rf electrode. A novel ESI-FAIMS interface is described that optimizes FAIMS gas flows resulting in large improvements in transmission. Simulations with the bromochloroacetate anion showed an improvement of similar to 9-fold to give similar to 70% overall transmission). Comparable transmission improvements were attained experimentally for six peptides (2+) in the range of m/z 404.2 to 653.4 at a chromatographic flow rate of 300 nL/min. Selected ion chromatograms (SIC) from nano-LC-FAIMS-MS analyses showed 71% (HLVDEPQNLIK, m/z 653.4, 2+) to 95% (LVNELTEFAK, m/z 582.3, 2+) of ion signal compared with ion signal in the SIC from LC-MS analysis. IGSEVYHNLK (580.3, 2+) showed 24% more ion signal compared with LC-MS and is explained by enhanced desolvation in FAIMS. A 3-10 times lower limits of quantitation (LOQ) (< 15% RSD) was achieved for chemical noise limited peaks with FAIMS. Peaks limited by ion statistics showed subtle improvement in RSD and yielded comparable LOQ to that attained with nano-LC-MS (without FAIMS). These improvements were obtained using a reduced FAIMS separation gap (from 2.5 to 1.5 mm) that results in a shorter residence time (13.2 ms +/- 3.9 ms) and enables the use of a helium free transport gas (100% nitrogen).
引用
收藏
页码:2143 / 2153
页数:11
相关论文
共 50 条
  • [1] Ionization and transmission efficiency in an electrospray ionization-mass spectrometry interface
    Page, Jason S.
    Kelly, Ryan T.
    Tang, Keqi
    Smith, Richard D.
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (09) : 1582 - 1590
  • [2] Biases in Ion Transmission Through an Electrospray Ionization-Mass Spectrometry Capillary Inlet
    Page, Jason S.
    Marginean, Ioan
    Baker, Erin S.
    Kelly, Ryan T.
    Tang, Keqi
    Smith, Richard D.
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2009, 20 (12) : 2265 - 2272
  • [3] Analysis of heparan sulfate oligosaccharides by nano-electrospray ionization mass spectrometry
    Pope, RM
    Raska, CS
    Thorp, SC
    Liu, JA
    [J]. GLYCOBIOLOGY, 2001, 11 (06) : 505 - 513
  • [4] Ion transmission in an electrospray ionization-mass spectrometry interface using an S-lens
    Yan, Yihui
    Schmitt, Lucas
    Khramchenkova, Anastasiya
    Lengyel, Jozef
    [J]. JOURNAL OF MASS SPECTROMETRY, 2023, 58 (07):
  • [5] Hyphenated sample preparation-electrospray and nano-electrospray ionization mass spectrometry for biofluid analysis
    Mikhail, Ibraam E.
    Tehranirokh, Masoomeh
    Gooley, Andrew A.
    Guijt, Rosanne M.
    Breadmore, Michael C.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2021, 1646
  • [6] Characterization of lead-phytochelatin complexes by nano-electrospray ionization mass spectrometry
    Scheidegger, Christian
    Suter, Marc J. -F.
    Behra, Renata
    Sigg, Laura
    [J]. FRONTIERS IN MICROBIOLOGY, 2012, 3
  • [7] Nano-electrospray ionization mass spectrometry:: addressing analytical problems beyond routine
    Karas, M
    Bahr, U
    Dülcks, T
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 366 (6-7): : 669 - 676
  • [8] Fast profiling of anthocyanins in wine by desorption nano-electrospray ionization mass spectrometry
    Hartmanova, Lucie
    Ranc, Vaclav
    Papouskova, Barbora
    Bednar, Petr
    Havlicek, Vladimir
    Lemr, Karel
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (25) : 4223 - 4228
  • [9] Nano-electrospray ionization mass spectrometry: addressing analytical problems beyond routine
    M. Karas
    U. Bahr
    T. Dülcks
    [J]. Fresenius' Journal of Analytical Chemistry, 2000, 366 : 669 - 676
  • [10] Dual Emitter Nano-Electrospray Ionization Coupled to Differential Ion Mobility Spectrometry-Mass Spectrometry for Shotgun Lipidomics
    Keating, James E.
    Glish, Gary L.
    [J]. ANALYTICAL CHEMISTRY, 2018, 90 (15) : 9117 - 9124