In vitro fragmentation efficiency of holmium: yttrium-aluminum-garnet (YAG) laser lithotripsy - a comprehensive study encompassing different frequencies, pulse energies, total power levels and laser fibre diameters

被引:80
|
作者
Kronenberg, Peter [1 ]
Traxer, Olivier [2 ]
机构
[1] Hosp Prof Doutor Fernando Fonseca, Dept Urol, P-2720276 Amadora, Portugal
[2] Univ Paris 06, Hop Tenon, Paris, France
关键词
holmium:YAG laser lithotripsy; urinary calculi; high frequency lithotripsy; lithotripter settings; automated laser fragmentation testing system; CALCULUS RETROPULSION; STONE FRAGMENTATION; DEPENDENCE; VARIES;
D O I
10.1111/bju.12567
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective To assess the fragmentation (ablation) efficiency of laser lithotripsy along a wide range of pulse energies, frequencies, power settings and different laser fibres, in particular to compare high- with low-frequency lithotripsy using a dynamic and innovative testing procedure free from any human interaction bias. Materials and Methods An automated laser fragmentation testing system was developed. The unmoving laser fibres fired at the surface of an artificial stone while the stone was moved past at a constant velocity, thus creating a fissure. The lithotripter settings were 0.2-1.2 J pulse energies, 5-40 Hz frequencies, 4-20 W power levels, and 200 and 550 mu m core laser fibres. Fissure width, depth, and volume were analysed and comparisons between laser settings, fibres and ablation rates were made. Results Low frequency-high pulse energy (LoFr-HiPE) settings were (up to six times) more ablative than high frequency-low pulse energy (HiFr-LoPE) at the same power levels (P < 0.001), as they produced deeper (P < 0.01) and wider (P < 0.001) fissures. There were linear correlations between pulse energy and fragmentation volume, fissure width, and fissure depth (all P < 0.001). Total power did not correlate with fragmentation measurements. Laser fibre diameter did not affect fragmentation volume (P = 0.81), except at very low pulse energies (0.2 J), where the large fibre was less efficient (P = 0.015). Conclusions At the same total power level, LoFr-HiPE lithotripsy was most efficient. Pulse energy was the key variable that drove fragmentation efficiency. Attention must be paid to prevent the formation of time-consuming bulky debris and adapt the lithotripter settings to one's needs. As fibre diameter did not affect fragmentation efficiency, small fibres are preferable due to better scope irrigation and manoeuvrability.
引用
收藏
页码:261 / 267
页数:7
相关论文
共 12 条
  • [1] Holmium:yttrium-aluminum-garnet laser induced lithotripsy: in-vitro investigations on fragmentation, dusting, propulsion and fluorescence
    Eisel, Maximilian
    Stroebl, Stephan
    Pongratz, Thomas
    Strittmatter, Frank
    Sroka, Ronald
    BIOMEDICAL OPTICS EXPRESS, 2018, 9 (11): : 5115 - 5128
  • [2] Systematic evaluation of a holmium: yttrium-aluminum-garnet laser lithotripsy device with variable pulse peak power and pulse duration
    Netsch, Christopher
    Knipper, Sophie
    Tiburtius, Christian
    Gross, Andreas J.
    ASIAN JOURNAL OF UROLOGY, 2014, 1 (01) : 60 - 65
  • [3] Comparison of Different Pulse Modulation Modes for Holmium:Yttrium-Aluminum-Garnet Laser Lithotripsy Ablation in a Benchtop Model
    Terry, Russell S.
    Ho, Derek S.
    Scialabba, Dominick M.
    Whelan, Patrick S.
    Qi, Robert
    Ketterman, Brian T.
    Preminger, Glenn M.
    Zhong, Pei
    Lipkin, Michael E.
    JOURNAL OF ENDOUROLOGY, 2022, 36 (01) : 29 - 37
  • [4] Holmium: Yttrium-aluminum-garnet laser lithotripsy: Is there a difference in ablation rates between short and long pulse duration?
    Ntasiotis, Panteleimon
    Peteinaris, Angelis
    Lattarulo, Marco
    Tsaturyan, Arman
    Asutay, Mehmet Kazim
    Adamou, Constantinos
    Vagionis, Athanasios
    Pagonis, Konstantinos
    Koukiou, Georgia
    Al-Aown, Abdulrahman
    Liatsikos, Evangelos
    Kallidonis, Panagiotis
    UROLOGY ANNALS, 2023, 15 (02) : 202 - 206
  • [6] Effect of Pulse Energy, Frequency and Length on Holmium: Yttrium-Aluminum-Garnet Laser Fragmentation Efficiency in Non-Floating Artificial Urinary Calculi
    Wezel, Felix
    Haecker, Axel
    Gross, Andreas J.
    Michel, Maurice Stephan
    Bach, Thorsten
    JOURNAL OF ENDOUROLOGY, 2010, 24 (07) : 1135 - 1140
  • [7] Efficiency of treating asymptomatic ureteral stones: Shock wave lithotripsy versus ureteroscopy plus holmium: Yttrium-aluminum-garnet laser
    Liu, Yuqing
    Zhuang, Shenrong
    Xiao, Chunlei
    JOURNAL OF ENDOUROLOGY, 2007, 21 : A241 - A242
  • [8] Efficiency and cost of treating proximal ureteral stones:: Shock wave lithotripsy versus ureteroscopy plus holmium:yttrium-aluminum-garnet laser
    Segura, JW
    UROLOGY, 2004, 64 (06) : 1106 - 1106
  • [9] Second-look Flexible Nephroscopy Combined With Holmium: Yttrium-Aluminum-Garnet Laser Lithotripsy Under Local Anesthesia: A Prospective Study
    Giannakopoulos, Stilianos
    Giannopoulos, Stavros
    Gardikis, Stefanos
    Bantis, Athanasios
    Kalaitzis, Christos
    Zissimopoulos, Athanasios
    Touloupidis, Stavros
    UROLOGY, 2017, 99 : 27 - 32
  • [10] Prognostic outcome of cervical laser ablation using a holmium yttrium-aluminum-garnet (Ho:YAG) laser for the treatment of cervical intraepithelial neoplasia: A single-center retrospective study
    Suzuki, Wataru
    Ietani, Kana
    Makabe, Takeshi
    Oki, Shinya
    Ohno, Akiko
    Mikami, Yoshiko
    Yamashita, Hiroshi
    GYNECOLOGIC ONCOLOGY REPORTS, 2024, 53