MRI blind zone: A proposed novel index for spinal implants artifacts

被引:1
|
作者
Harel, Ran [1 ]
Kimchi, Gil [1 ]
Knoller, Nachshon [1 ]
机构
[1] Tel Aviv Univ, Sackler Med Sch, Sheba Med Ctr, Spine Surg Div,Dept Neurosurg, Ramat Gan, Israel
关键词
MRI blind zone; MRI artifacts; Metal implants; Metal artifacts; Implant index; METAL-INDUCED ARTIFACTS; TITANIUM;
D O I
10.1016/j.clinimag.2022.06.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To advocate the formulation of a new index termed MRI Blind Zone, designated to predict the dimensions of the magnetic resonance metallic artifact caused by specific spinal implants. The index may also specify the obscured organs of interest. Methods: A retrospective evaluation of post-operative MR images of patients operated for spinal instrumentation with various implants from different materials, in our institution, was performed. The MRI blind zone was described for each product, and the related obscured region of interest was discussed. A proposed 3D model was created as an example for suggested future reporting by the implants' industry. Results: Seven implant types are presented. The post-operative MR artifacts were detailed, and the clinical implications were discussed. Material type, processing methods and individual anatomical traits have a dramatic effect on the MRI blind zone and the obscured regions. Conclusion: MRI artifact is multifactorial and is influenced in part by the implant's shape, size, material and processing method. Individual products affect post-operative MR artifacts to different extents, and may carry clinical implications when post-operative imaging is required. A standardized index displaying the predicted post-operative artifact is warranted. For the manufacturers to accurately report the data to the surgeons, a parametric standardization should be performed. MRI blind zone index will allow surgeons to compare between different implants efficiently, and improve the informed decision-making process of implant selection.
引用
收藏
页码:78 / 83
页数:6
相关论文
共 50 条
  • [1] Artifacts Quantification of Metal Implants in MRI
    Vrachnis, I. N.
    Vlachopoulos, G. F.
    Maris, T. G.
    Costaridou, L. I.
    INTERNATIONAL CONFERENCE ON BIO-MEDICAL INSTRUMENTATION AND RELATED ENGINEERING AND PHYSICAL SCIENCES (BIOMEP 2017), 2017, 931
  • [2] Metallic Implants in MRI - Hazards and Imaging Artifacts
    Peschke, Eva
    Ulloa, Patricia
    Jansen, Olav
    Hoevener, Jan-Bernd
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2021, 193 (11): : 1285 - 1293
  • [3] MRI safety of cardiovascular implants: Evaluation of ferromagnetism, heating, and artifacts
    Shellock, FG
    Shellock, VJ
    RADIOLOGY, 2000, 214 (02) : 614 - 614
  • [4] Integral method for numerical simulation of MRI artifacts induced by metallic implants
    Balac, S
    Caloz, G
    Cathelineau, G
    Chauvel, B
    de Certaines, JD
    MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) : 724 - 727
  • [5] Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
    Hilgenfeld, Tim
    Prager, Marcel
    Schwindling, Franz S.
    Jende, Johann M. E.
    Rammelsberg, Peter
    Bendszus, Martin
    Heiland, Sabine
    Juerchott, Alexander
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2018, (135):
  • [6] Analysis of Artifacts on MRI from Electronic Identification Enablement in Silicone Gel Implants
    Munhoz, Alexandre Mendonca
    PLASTIC AND RECONSTRUCTIVE SURGERY-GLOBAL OPEN, 2022, 10 (06)
  • [7] MRI with a CI: Using the propeller technique reduces artifacts in children with cochlear implants
    An, Jingjiao
    Shu, Zhi
    Hao, Mengmeng
    Gao, Juan
    ACTA OTO-LARYNGOLOGICA CASE REPORTS, 2023, 8 (01): : 116 - 121
  • [8] A pneumatic phantom for mimicking respiration-induced artifacts in spinal MRI
    De Tillieux, Philippe
    Topfer, Ryan
    Foias, Alexandru
    Leroux, Iris
    El Maachi, Imanne
    Leblond, Hugues
    Stikov, Nikola
    Cohen-Adad, Julien
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (01) : 600 - 605
  • [9] A novel approach for reducing metal artifacts due to metallic dental implants
    Yazdi, Mehran
    Beaulieu, Luc
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2260 - 2263
  • [10] Quantitative Dosimetry Assessment of the Impact of Image Artifacts of Metal Implants in Spinal SABR Treatment
    Chen, T.
    Zhang, M.
    Hanft, S.
    Green, R.
    Yue, N.
    Goyal, S.