Wireless Synchronization of a Multi-Pinhole Small Animal SPECT Collimation Device With a Clinical Scanner

被引:1
|
作者
DiFilippo, Frank P. [1 ]
Patel, Sagar [1 ]
机构
[1] Cleveland Clin, Dept Nucl Med, Cleveland, OH 44195 USA
基金
美国国家卫生研究院;
关键词
Single photon emission computed tomography; small animal; synchronization; wireless; MOTION; RESOLUTION;
D O I
10.1109/TNS.2009.2018843
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A multi-pinhole collimation device for small animal single photon emission computed tomography (SPECT) uses the gamma camera detectors of a standard clinical SPECT scanner. The collimator and animal bed move independently of the detectors, and therefore their motions must be synchronized. One approach is manual triggering of the SPECT acquisition simultaneously with a programmed motion sequence for the device. However, some data blurring and loss of image quality result, and true electronic synchronization is preferred. An off-the-shelf digital gyroscope with integrated Bluetooth interface provides a wireless solution to device synchronization. The sensor attaches to the SPECT gantry and reports its rotational speed to a notebook computer controlling the device. Software processes the rotation data in real-time, averaging the signal and issuing triggers while compensating for baseline drift. Motion commands are sent to the collimation device with minimal delay, within approximately 0.5 second of the start of SPECT gantry rotation. Test scans of a point source demonstrate an increase in true counts and a reduction in background counts compared to manual synchronization. The wireless rotation sensor provides robust synchronization of the collimation device with the clinical SPECT scanner and enhances image quality.
引用
收藏
页码:640 / 645
页数:6
相关论文
共 50 条
  • [21] A prototype low-energy, multi-pinhole SPECT system for small-animal imaging
    Peterson, TE
    Wilson, DW
    Barrett, HH
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2999 - 3002
  • [22] Design of a Multi-Pinhole Collimator in a Dual-Headed, Stationary, Small-Animal SPECT
    Shokouhi, S.
    Fritz, M. A.
    McDonald, B. S.
    Wilson, D. W.
    Metzler, S. D.
    Peterson, T. E.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2399 - 2402
  • [23] Compact Multi-Pinhole SPECT Imager with Sub-Millimeter Resolution for Small Animal Research
    Lackas, C.
    Hoppin, J. W.
    Schramm, N. U.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S254 - S254
  • [24] Evaluation of multiplexing in multi-pinhole SPECT
    Jorgensen, Aaron K.
    Zeng, Gengsheng L.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2213 - 2217
  • [25] Multi-pinhole spect: Recent results of an animal imaging system.
    Schramm, NU
    Behe, M
    Schurrat, T
    Engeland, U
    Ebel, G
    Behr, TM
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 9P - 10P
  • [26] Atherosclerotic plaque imaging on ApoE-/- mice using small animal multi-pinhole SPECT
    Mok, Seng Peng
    Yu, Jianhua
    Wang, Yuchuan
    Nimmagadda, Sridhar
    Pomper, Martin
    Tsui, Benjamin
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [27] A Sensitivity Model for Multi-Pinhole SPECT
    DiFilippo, Frank P.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 1842 - 1847
  • [28] Triple and dual PET quantification of a Small-Animal Multi-Pinhole PET/SPECT/CT System
    Echegoyen, P.
    Cuenca Bandin, T.
    Ecay, M.
    Prieto, E.
    Collantes, M.
    Pareja, F.
    Simon Martinez, J.
    Penuelas, I.
    Marti-Climent, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S144 - S144
  • [29] A scanning focus nuclear microscope with multi-pinhole collimation
    Nguyen, Minh Phuong
    Arif, Muhammad
    Oostenrijk, Bart
    Goorden, Marlies C.
    Beekman, Freek J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (07):
  • [30] Theoretical Analysis of Multi-Pinhole Brain SPECT
    Goorden, Marlies C.
    Rentmeester, Mart C. M.
    Beekman, Freek J.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3311 - 3313